Registered trade mark Magnetische Pruefanlagen GmbH Member of FOERSTER GROUP In Laisen 65, 72766 Reutlingen, Germany INSTITUT DR. FOERSTER GMBH & CO. KG Telefon: +49 7121 1099-0, Telefax +49 7121 470370 In Laisen 70, 72766 Reutlingen, Germany Email: info@mp-ndt.de, www.mp-ndt.de...
Only NiMH or NiCd rechargeable batteries may be installed in the ● DEFECTOMETER M 1.837. If an external power supply is connected to the device, the installed batteries are charged automatically. If the device is operated with non-rechargeable batteries, any...
1.1 Declaration of conformity DECLARATION OF CONFORMITY We declare that the following product Equipment: DEFECTOMETER M 1.837 Manufacturer: Magnetische Prüfanlagen GmbH complies with the requirements of the following European directives and corresponding standards: 89/336/EWG (Elektromagnetische Verträglichkeit) ● DIN EN 61326 – 1, Elektromagnetische Verträglichkeit □...
2. Introduction The DEFECTOMETER M 1.837 is designed for crack testing. It is characterised by a very high sensitivity to surface defects and by simple, intuitive operation. Compared to the predecessor model, the DEFECTOMETER 2.837, it has numerous new features.
Backlight enabled Battery charge indicator Red LED lights up when remaining operating time <20 min On / Off 3.2 Interfaces Socket for mains Probe socket charger/power supply Earphone socket USB interface - 7 - DEFECTOMETER M 1.837...
To skip the "Tilt Mat"function, briefly press and release the Zero button. The device will then select a standard default setting.,However, this setting will lead to greater fluctuations of the measured value when the probe is tilted. - 8 - DEFECTOMETER M 1.837...
- and therefore the sensitivity - of the device. A gain of 6 dB corresponds to doubling the signal amplitude. Setting the defect threshold with the Discriminator button The DEFECTOMETER M 1.837 allows the user to set a maximum permitted defect level (threshold). A signal above this preset threshold value is visually indicated by a red LED.
The current position of the zero point is displayed in the middle of the bottom line of the LCD display. For example, Z10 means that the zero point is situated at 10% on the scale. - 10 - DEFECTOMETER M 1.837...
When the battery charge drops below a minimum limit that allows continued operation for approx. 20 minutes, the red LED on the battery button starts to light up. After pressing the battery button during the charge process, the - 11 - DEFECTOMETER M 1.837...
Replacement of the batteries or charging of the batteries with an external charging device lead to battery calibration data and the calibration has to performed again. 3.12 Battery Compartment Screw of battery compartment - 12 - DEFECTOMETER M 1.837...
LED should be lit. Further calibration now consists in setting the correct gain. The gain is generally set in accordance with a test specification that lays down a - 13 - DEFECTOMETER M 1.837...
Page 14
In this case, it must be assumed that the crack depth displayed for the tested part does not exactly match the depth displayed for the calibration test piece. The extent of this deviation is proportional to the difference in conductivity and permeability of the respective materials. - 14 - DEFECTOMETER M 1.837...
The Software can be installed under Microsoft Windows XP ● Microsoft Windows 7 ● Microsoft Windows 10 ● For installation start the DefectometerM_1.4.x_setup.exe file from the CD. Follow the instructions of the Setup Wizard. - 15 - DEFECTOMETER M 1.837...
Here you can set the range of the complex plain. The X- and Y-ranges can be set separately. In the context menu of the waterfall chart you can set the range of the waterfall chart. The X- and Y-ranges can be set separately. - 16 - DEFECTOMETER M 1.837...
The buttons of the virtual DEFECTOMETER M can be used for remote control. The ON/OFF button is for connecting/disconnecting with the DEFECTOMETER. Additionally the Zero Offset, the gain and the Discriminator Threshold can be set via sliders. - 17 - DEFECTOMETER M 1.837...
“Print” in the main menu. Alternativly it can be created a PDF-file via “Print”, provided that a program to create pdf files is installed on the PC, such as Adobe Acrobat. - 18 - DEFECTOMETER M 1.837...
Under normal operating conditions, we recommend a calibration interval of one year for the device and two years for the calibration standards. Device calibration can be conducted by Institut Dr. Foerster or Magnetische Prüfanlagen GmbH or by one of our authorised service centres.
IP 54 (without connection sockets) Permissible operating -10 °C to + 55°C temperature Permissible storage/transport -20 °C to +70 °C temperature Permissible relative humidity 5 % to 95 % Weight (incl. batteries) 0.4 kg - 20 - DEFECTOMETER M 1.837...