Please contact the manufacturer for replacement parts or service. Ultra-Low Temperature Model Numbers This owner’s manual is to be used for the following LACO Technologies Ascent Chillers: Table 1 - Chiller Model Numbers Type Model Number...
Only connect LACO Technologies Ascent Chillers to a properly grounded circuit. b. Never connect your LACO Technologies Ascent Chillers to voltage that is outside the acceptable voltage range. Connecting your chiller to any voltage outside of these ranges will damage the compressor.
Inlet/Outlet) AC0500-ECU 8. Use Only Approved Fluids - LACO Technologies Ascent Chillers designs every chiller for specific fluids. Only used approved fluids or consult with the manufacture prior to using an alternate fluid. Each chiller is tested prior to shipping and will have some residual fluid in the system.
WARNING: Adhere to all safety and warning labels. No not remove any of these labels. Removing or opening panels with this warning will expose live electrical circuits. All power should be disconnected from equipment prior to removing or opening these panels. Ascent Chillers Operations and Maintenance Manual LACO Technologies...
All power should be disconnected from equipment prior to removing or opening these panels. CAUTION: Earth (Ground) Protective conductor terminal. Technical Specifications Table 3 - Chiller Technical Specifications Model AC0500-ULT-240/3 AC1000-ULT-240/3 AC2000-ULT-480/3 AC5000-ULT-480/3 AC5000-ECU-480/3 Dry Weight 1110 lbs...
Place unit on a flat level surface in a well-ventilated area Leave adequate space around chiller for good ventilation and access to panels. See Chiller Spacing figure for minimum clearances. LACO Technologies Ascent Chillers Operations and Maintenance Manual...
Inlet/Outlet) AC0500-ECU WARNING: Do not position equipment in such a way that the main power disconnect is difficult to operate. WARNING: Only a licensed electrician should connect the chiller to electrical service. Ascent Chillers Operations and Maintenance Manual LACO Technologies...
30 minutes prior to turning on the refrigeration circuit. This is only required if the main disconnect has been turned off for 12 hours or more. LACO Technologies Ascent Chillers Operations and Maintenance Manual...
CAUTION: If the system is pressurized, while wearing safety goggles, slowly release the pressure from the system. 3. Open Fluid Expansion Bypass Valve. Figure 12 - Fluid Expansion Bypass Valve 4. Carefully fill the system with fluid. Ascent Chillers Operations and Maintenance Manual LACO Technologies...
Page 20
12. While the system is at ambient temperature, verify that the fluid level is between the minimum and maximum fill lines. 13. Adjust desired setpoint and setup is complete. LACO Technologies Ascent Chillers Operations and Maintenance Manual...
6 INSTALLATION AND ASSEMBLY WARNING: Extreme caution must be taken to prevent moisture from entering the system. The system shipped dry from LACO Technologies Ascent Chillers. It is the responsibility of the end user to make sure that there is ABSOLUTELY no moisture in their side of the system. If moisture gets into the process fluid, it must be removed immediately.
Enter username and password (Default Username: Operator (case sensitive), Default Password: NSC9876chi!! User Interface The main menu can be accessed by pressing the + Icon at the bottom left of the screen. LACO Technologies Ascent Chillers Operations and Maintenance Manual...
This slider will turn the system on/off. When on the pump will run and the chiller will run to maintain the target setpoint. 8.2.1.2. Setpoint Changing this value will adjust the temperature the chiller will maintain. 8.2.1.3. Temperatures Ascent Chillers Operations and Maintenance Manual LACO Technologies...
Page 24
“Reset Reservoir Alarm” button is pressed. When the chiller is first started there will be a Heater Overtemp alarm. The “Heater OT Reset” button will reset this alarm. This pane shows any active alarms. LACO Technologies Ascent Chillers Operations and Maintenance Manual...
Schedule is where you can set the system to auto turn on/off 8.2.5.1. Auto-Run Times Enter the on time for the day you want the chiller to auto-run on and the stop time. Ascent Chillers Operations and Maintenance Manual LACO Technologies...
Monday at 8:00 AM. You might want the chiller to not run on Memorial Day, so you would type in 05/29/23. Parameters (Only Adjustable by Admin) 8.2.6.1. Suction Accumulator Solenoid This should only be adjusted by qualified service personnel. LACO Technologies Ascent Chillers Operations and Maintenance Manual...
Page 27
R508b LP Pressure Cutin (psig) is the pressure that the chiller will turn back on at. This is preset by the factory. R404a Pres/R508b On (psig) controls when the R508b circuit comes on when the R404a circuit is ready to handle the load. Ascent Chillers Operations and Maintenance Manual LACO Technologies...
The main menu can be accessed by pressing the + Icon at the bottom left of the screen, then selecting Utility and then DateTime Config. Type in the time and time zone and press Save. LACO Technologies Ascent Chillers Operations and Maintenance Manual...
110° F. Verify that the Heater Over-temp alarm shuts down the heater and requires a manual reset at or below the 110°F. Reset the over temperature alarm to its default value of 150° F. Ascent Chillers Operations and Maintenance Manual LACO Technologies...
DO NOT use any decontamination or cleaning agents which could cause a hazard as the result of a reaction with parts of the equipment or material contained within. Consult the manufacturer if there is any doubt LACO Technologies Ascent Chillers Operations and Maintenance Manual...
Too much fluid in Each ULT is designed around a system maximum volume of Syltherm. • Add additional expansion Exceeding this volume can tank. result in excessively high pressures in the reservoir. Ascent Chillers Operations and Maintenance Manual LACO Technologies...
Page 36
Fluid outlet is not closing. and check suction pressure Temp temperature while trying to maintain -60C. had dropped below the When -60C is achieved the minimum cooling solenoid will start LACO Technologies Ascent Chillers Operations and Maintenance Manual...
LACO Technologies Ascent Chillers Product or for results of or damages caused to persons or property by the use of the LACO Technologies Ascent Chillers Product. Only the Manufacturer is authorized to make any warranty or representation and the customer may not rely on any other warranty or representation.
LACO Technologies Ascent Chillers product. a. When any LACO Technologies Ascent Chillers Product has been authorized to return for any inspection, or replacement, it must be returned as specified in the Return Merchandise Authorization Form.