Summary of Contents for advantest corporation R3267 Series
Page 1
Cover R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual MANUAL NUMBER FOE-8335241F00 Applicable models R3264 R3267 R3273 First printing Jun 20, 1999 1999 ADVANTEST CORPORATION All rights reserved. Printed in Japan...
Page 2
Safety Summary No. ESF00 Safety Summary To ensure thorough understanding of all functions and to ensure efficient use of this instrument, please read the manual carefully before using. Note that Advantest bears absolutely no responsibility for the result of operations caused due to incorrect or inappropriate use of this instrument.
Page 3
Safety Summary • When the product has ventilation outlets, do not stick or drop metal or easily flammable ob- jects into the ventilation outlets. • When using the product on a cart, fix it with belts to avoid its drop. •...
Page 4
Safety Summary Main Parts with Limited Life Part name Life Unit power supply 5 years Fan motor 5 years Electrolytic capacitor 5 years LCD display 6 years LCD backlight 2.5 years Floppy disk drive 5 years • Hard Disk Mounted Products The operational warnings are listed below.
Page 5
Environmental Conditions This instrument should be only be used in an area which satisfies the following conditions: • An area free from corrosive gas • An area away from direct sunlight • A dust-free area • An area free from vibrations Direct sunlight Dust Corrosive...
Page 6
PREFACE R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual PREFACE This manual provides the information necessary to check functionality, operate and program the R3267 Series Option 63, GSM/DECT measurement. 1. Organization of this manual This manual consists of the following chapters:...
Page 7
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual PREFACE 2. Typeface conventions used in this manual • Panel keys and soft keys are printed in a contrasting typeface to make them stand out from the text as fol- lows: Panel keys: Boldface type...
TABLE OF CONTENTS R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual TABLE OF CONTENTS INTRODUCTION ................... Product Overview ....................Accessories ......................Self Test Function ....................About Calibration ....................Explanation of the Connectors ................OPERATION ....................Measuring Multi-burst Signals (No External Triggers Available) ......
Page 9
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual Table of Contents 3.3.6.1 DC CAL ...................... 3-79 3.3.6.2 Channel Setting ................... 3-79 3.3.6.3 STD Setup ....................3-79 REMOTE CONTROL ..................GPIB Command Index ..................GPIB Command Codes ..................4-10 TECHNICAL INFORMATION ..............
Page 10
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual Table of Contents 6.2.2 Modulation Accuracy Measurement for EDGE ........... 6.2.3 Frequency Deviation Accuracy Measurement for DECT ......Performance Verification Test Record Sheet ............SPECIFICATIONS ..................APPENDIX ........................ A.1 Messages ......................ALPHABETICAL INDEX...
Page 11
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual LIST OF ILLUSTRATIONS Title Page Connections for GSM Measurements ................STD Measurement Parameter Set Dialog Box ..............Parameter Setup Dialog Box ..................... Phase Error Measurement Results ..................Displaying Phase Error vs Bit Graph ................
Page 12
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual List of Illustrations Title Page 3-36 Parameter Setup Dialog Box ..................... 3-77 A Signal Containing Multiple Bursts in a Frame ............. A Signal Containing Up Link/Down Link Bursts in a Frame ..........
Page 13
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual LIST OF TABLES Title Page Operating mode ......................... 4-10 ATT key (Attenuator) ....................... 4-10 COPY key (Hand copy) ....................4-10 COUPLE key (Couple function) ..................4-11 FREQ key (Frequency) ..................... 4-11 LEVEL key (Reference Level) ..................
The GSM/EDGE/DECT option (Option 63) is the software designed to measure and evaluate frequency er- rors in GSM/DECT as well as characteristics of burst waveform’s leading and trailing edges. This option is a factory option which is incorporated into the R3267 Series Spectrum Analyzer prior to ship- ment.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 1.5 Explanation of the Connectors Explanation of the Connectors Connectors used for this option are described as follows: EXT TRIG terminal Connector for inputting the external trigger signal. INPUT I channel terminal Connector for inputting the I channel signal (Baseband).
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 2.1 Measuring Multi-burst Signals (No External Triggers Available) OPERATION This chapter describes how to use this option using practical measurement examples. Measuring Multi-burst Signals (No External Triggers Available) This section describes how to measure the signals (from GSM mobile station or the base station) which con- tain multiple bursts in one frame.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 2.1 Measuring Multi-burst Signals (No External Triggers Available) Select 0 from TSC using the data knob, and press the data knob (or ENTR) to register the parameter. 10. Select WIDE from Filter Mode using the data knob, and press the data knob (or ENTR) to register the parameter.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 2.1 Measuring Multi-burst Signals (No External Triggers Available) 21. Select FREE RUN from Trigger Source using the data knob, and press the data knob (or ENTR) to register the parameter. 22. Press the arrow key↓ to shift the cursor to Search Level and enter -25 dB.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 2.1 Measuring Multi-burst Signals (No External Triggers Available) Figure 2-4 Phase Error Measurement Results Displaying the graphics Phase Error vs Bit graph is displayed. 29. Press Graphics. 30. Press Select Type to open Graphic Type of Analysis dialog box.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 2.1 Measuring Multi-burst Signals (No External Triggers Available) 33. Press SHIFT and MKR to delete the marker. 34. Press RETURN and RETURN (RETURN twice) to return to the measurement menu.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 2.2 Measuring the Multi-Burst Signals (Using an External Trigger) Measuring the Multi-Burst Signals (Using an External Trigger) This section describes how to measure a MULTI-BURST signal, which consist of multiple bursts in one frame, at mobile station or the base station that are used in the GSM system.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 2.2 Measuring the Multi-Burst Signals (Using an External Trigger) 10. Select WIDE from Filter Mode using the data knob, and press the data knob (or ENTR) to register the parameter. For the definition of the filter, refer to Section 5.1 “NARROW/WIDE in the GSM filter mode.”...
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 2.2 Measuring the Multi-Burst Signals (Using an External Trigger) 22. Select + from Trigger Slope using the data knob, and press the data knob (or ENTR) to register the parameter. 23. Select 156.25 BIT from TDMA Structure using the data knob, and press the data knob (or ENTR) to register the parameter.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 2.2 Measuring the Multi-Burst Signals (Using an External Trigger) Figure 2-9 Phase Error measurement results Displaying the graphics Constellation graph is displayed. 31. Press Graphics. 32. Press Select Type to open Graphic Type of Analysis dailog box.
Page 25
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 2.2 Measuring the Multi-Burst Signals (Using an External Trigger) 34. Press MKR to display the marker. Turn the data knob to move the marker. 35. Press SHIFT and MKR to make the marker to go out.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3 REFERENCE REFERENCE This chapter describes the functions of the panel and soft keys for option 63 software. Menu Index This menu index is used to easily find the keys described in Chapter 3.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.2 Menu Map Menu Map This section shows the hierarchical menu configuration on a panel key basis. TRANSIENT T-Domain Power P3-8 ON/OFF Ratio F-Domain P3-9 Modulation Spurious P3-10 Due to Modulation P3-11...
Page 33
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.2 Menu Map TRANSIENT T-Domain Power Auto Level Set Trigger Setup Trigger Source: FREE RUN Window Setup Window ON/OFF VIDEO Template Set to STD Y Scale [dB/div] 10/5/2 Window Position Average Times ON/OFF...
Page 34
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.2 Menu Map TRANSIENT T-Domain ON/OFF Ratio Auto Level Set Trigger Setup Trigger Source: FREE RUN Window Setup VIDEO Window ON/OFF Y Scale [dB/div] 10/5/2 Set to STD Average Times ON/OFF ON Position...
Page 35
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.2 Menu Map TRANSIENT T-Domain Spurious Auto Level Set Trigger Setup Trigger Source: FREE RUN Table No. 1/2/3 Load Table Table Edit Slope: Table No. 1/2/3 Average Times ON/OFF Load Table Config...
Page 36
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.2 Menu Map TRANSIENT T-Domain Due to Modulation Auto Level Set Trigger Source: FREE RUN Trigger Setup Window Setup Window ON/OFF Table Edit Set to STD Average Times Slope: Window Position Config...
Page 37
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.2 Menu Map TRANSIENT F-Domain Power Auto Level Set Gate Setup Trigger Setup Trigger Source: FREE RUN Window Setup Gate Source VIDEO Trigger Y Scale [dB/div] 10/5/2 Gate Setup Ext Gate Average Times ON/OFF...
Page 38
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.2 Menu Map TRANSIENT F-Domain Auto Level Set OBW% Average Times ON/OFF Config Parameter Setup Detector: NORMAL Set to STD POSI NEGA SAMPLE Judgment: Upper Limit Lower Limit Average Mode: TRACE AVG...
Page 39
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.2 Menu Map TRANSIENT F-Domain Due to Transient Auto Level Set Template ON/OFF Template Copy from STD Shift X Marker Edit (When select to GSM/EDGE) Shift Y Average Times ON/OFF Insert Line Margin ∆X ON/OFF...
Page 40
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.2 Menu Map TRANSIENT F-Domain Due to Modulation Auto Level Set Gate Setup Trigger Setup Trigger Source: FREE RUN Template Gate Source Trigger VIDEO Marker Edit Gate Setup Ext Gate Average Times ON/OFF...
Page 41
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.2 Menu Map From the previous page Freq. Setting: START&STOP SPAN Detector: NORMAL POSI NEGA SAMPLE Result: MARKER RELATIVE ABS POWER Ref Power: REF MARKER MODULATION Display Unit: dBm dBµV Template Couple to Power:...
Page 42
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.2 Menu Map TRANSIENT F-Domain Inband Spurious Auto Level Set Template ON/OFF Template Shift X Marker Edit Shift Y Average Times ON/OFF ∆ Config Margin X ON/OFF Template Edit Copy from STD...
Page 43
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.2 Menu Map TRANSIENT F-Domain Outband Spurious Auto Level Set Table No. 1/2/3 Load Table Table Edit Copy from STD Average Times ON/OFF Table No. 1/2/3 Config Load Table Save Table Insert Line...
Page 44
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.2 Menu Map TRANSIENT (When select to GSM/EDGE) Modulation Phase Error Auto Level Set Graphics Select Type Constellation Parameter Setup Constellation(Line) Average Times ON/OFF Constellation(Dot) Constellation(Line & Dot) I EYE Diagram Q EYE Diagram...
Page 45
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.2 Menu Map TRANSIENT (When select to GSM/EDGE) Modulation Modulation Accuracy Auto Level Set Graphics Select Type Constellation Parameter Setup Constellation(Line) Average Times ON/OFF Constellation(Dot) Constellation(Line & Dot) I EYE Diagram Q EYE Diagram...
Page 46
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.2 Menu Map (When select to GSM/EDGE) TRANSIENT Modulation Tx Power Auto Level Set Parameter Setup Trigger Source: FREE RUN Average Times ON/OFF Burst Search: Search Level Trigger Slope: Trigger Level TDMA Structure: 156.25 BIT...
Page 47
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.2 Menu Map (When select to GSM/EDGE) TRANSIENT Modulation Power vs Time Auto Level Set Template Entry MS Template Meas Mode NORM/HIGH BTS Template1 Y [dB/div] 20/10/5 BTS Template2 Parameter Setup STD Template...
Page 48
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.2 Menu Map TRANSIENT (When select to GSM/EDGE) Modulation Wave Check Time & FFT Auto Level Set Time Time & FFT Select Type IQ Complex FFT Parameter Setup Ich Time & FFT Average Times ON/OFF Qch Time &...
Page 49
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.2 Menu Map TRANSIENT (When select to DECT) Modulation Freq Deviation Auto Level Set Graphics Start Bit Frequency vs Bit Parameter Setup Select Type Frequency EYE Average Times ON/OFF Demodulated Data Trigger Source: FREE RUN...
Page 50
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.2 Menu Map TRANSIENT (When select to DECT) Modulation Tx Power Auto Level Set Parameter Setup Trigger Source: FREE RUN Average Times ON/OFF Burst Search: Trigger Slope: Trigger Level Full Slot Number...
Page 51
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.2 Menu Map TRANSIENT (When select to DECT) Modulation Wave Check Time & FFT Auto Level Set Time Select Type Time & FFT Parameter Setup IQ Complex FFT Average Times ON/OFF Ich Time & FFT Auto Level Set Qch Time &...
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description Functional Description When modulation analysis hardware and software are installed, the following menus are assigned to the TRANSIENT key. ATT AUTO/MNL MinATT ON/OFF Printer Copy Config CONFIG File Trace Point 1001/501...
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description 3.3.1 Switching Communication Systems This section describes how to switch the communication systems.The analyzer must be set to the SPA mode to switch between the communication systems. NOTE: After the communication system has been switched, the parameters previously set for the former com- munication system will be cleared.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description 3.3.2.1 Power (T-Domain) This is a function to measure power in the time domain (zero span). There are two Pass/Fail judgment functions: a judgment function for the template and a judgment function for power.
Page 56
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description NOTE: When Delay Time is a negative value, signals before the trigger can be captured. Window Setup Sets the window used for power measurement. Window ON/OFF Displays a window showing the range for power measurement.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description Average Times ON/OFF Sets the averaging count. For the method of average processing, refer to "Average Mode" in the Config → Parameter Setup. Config Parameter Setup Sets the method of measurement, edits the template, and so forth.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description MAX HOLD: Displays the maximum value within the average counts of the swept waveforms. POWER AVG: Converts the measured data (Log data) to the linear data to take the root mean square value.
Page 59
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description Captures the signal in sync with the IF signal (the lead- ing edge of the burst). EXT: Captures the signal in sync with the external trigger sig- nal. Slope Selects the edge when triggering.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description Parameter Setup Sets the measurement parameters. Figure 3-5 Parameter Setup Dialog Box Detector NORMAL/POSI/NEGA/SAMPLE Selects the detector. Display Unit dBm/W/dBµV Selects the display unit of power. NOTE: The ON/OFF ratio is displayed in units of dB (fixed).
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description 3.3.2.3 Spurious (T-Domain) This is a function to measure power (or peak power) according to the frequency specified in the table by sweeping in the zero span mode. Auto Level Set Sets the internal reference level to an optimum value in accor- dance with the measurement signal.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description Load Table Loads the measurement table. Table Edit Edits the measurement table. Table No. 1/2/3 Selects the table to be edited. Load Table Loads the measurement table. Save Table Saves the measurement table.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description Judgment Sets ON/OFF of the Pass/Fail judgment for the limit value. Preselector Sets the preselector. NOTE: This menu is displayed on R3267 only. 1.6G: Used to measure harmonics of more than 1.6 GHz or spurious signals when the carrier frequency is lower than 1.6 GHz.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description Trigger Setup Sets a trigger. Figure 3-8 Trigger Setup Dialog Box Trigger Source Selects a trigger. FREE RUN: Captures data using the internal measurement timing. Captures the signal in sync with the IF signal (the lead- ing edge of the burst) .
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description Insert Line Inserts additional frequency data before the selected frequency number. Delete Line Delete the selected line. Table Init Initialize the table. Average Times ON/OFF Sets the averaging count.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description 1.6G: Used to measure harmonics of more than 1.6Ghz or spurious signal when the carrier frequency is lower than 1.6GHz. 3.6G: Used to set this parameter for cases other than that above.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description Gate Setup Sets the gated sweep. This setting is required when the input signal is a bursted signal and Sample Detector is used. Trigger Setup Sets a trigger. Figure 3-10 Trigger Setup Dialog Box...
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description Ext Gate Sets the gated sweep mode using the gate signal input from the EXT GATE terminal on the rear panel. Gate Setup Sets the gated sweep range when Trigger is selected for Gate Source.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description Parameter Setup Sets measurement conditions. Figure 3-12 Parameter Setup Dialog Box Detector NORMAL/POSI/NEGA/SAMPLE Selects the detector. Gated Sweep Sets the gated sweep to ON or OFF. Display Unit dBm/W/dBµV Selects the display unit.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description 3.3.3.2 OBW Measure an occupied bandwidth. Auto Level Set Sets the internal reference level to an optimum value in accor- dance with the measurement signal. The reference level is auto- matically adjusted when this key is pressed.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description MAX HOLD: Calculates OBW based on the waveform with the max- imum value within the average counts of measured da- POWER AVG: Calculates OBW based on the waveforms, which were...
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description Delete Line Deletes the selected line. Sort Sorts the tables in order of frequency. Table Init Initializes the table. Marker Edit Sets the measurement frequency (frequency offset) and measure- ment band.
Page 73
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description Result Specifies how to display the result. MARKER: Displays the marker read value. The position of the marker is set by Marker Edit. RELATIVE: Displays the marker read value using a relative value.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description POWER AVG: Converts the measured data (Log data) to the linear data to take the root mean square value. NUMERIC: Converts the measured data (Log data) to the linear data to take the root mean square value.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description EXT: Captures the signal in sync with the external trigger sig- nal. Slope Selects the edge when triggering. Triggers at the leading edge. Triggers at the trailing edge. Trigger Level Sets the level to trigger.
Page 76
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description Template Sets and edits the template. For more information, refer to Section 5.8.2, "Template Setting in the F-Domain Measuring Mode." Template ON/OFF Sets the template display to ON or OFF.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description Parameter Setup Sets measurement conditions and so on. For more information, refer to Section 5.9.2, "Measurement re- sults Using Due to Modulation, Due to Transient and Inband Spu- rious Modes."...
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description Template Couple to Power Sets whether or not to raise or lower the template with the power set by Ref Power. Template Limit If the absolute value of the template is smaller than this vlaue when Template Couple to Power is set to ON, clip the template at this value.
Page 79
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description NOTE: The signal level must be constant while Auto Level Set is being carried out. Template Sets and edits the template. For more information, refer to Section 5.8.2, "Template Setting in the F-Domain Measuring Mode."...
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description Average Times ON/OFF Sets the averaging count. For more information, refer to Section 5.8.1, "Template Setting in the T-Domain Measuring Mode." Config Parameter Setup Sets measurement conditions and so on.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description Display Unit dBm/W/dBµV Selects the display unit. NOTE: When RELATIVE is selected for Result, the unit is dB. Template Couple to Power Sets whether or not to raise or lower the template with the power set by Ref Power.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description Table Edit Edits the table. Copy from STD Copies the template specified by the communication standard. Table No. 1/2/3 Selects the table number. Load Table Loads the table. Save Table Saves the table.
Page 83
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description 1.6G: Used to measure harmonics of more than 1.6 GHz or spurious signals when the carrier frequency is lower than 1.6 GHz. 3.6G: Used to set this parameter for cases other than that above.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description 3.3.4 Modulation (When select to GSM/EDGE) Modulation analysis by the DSP. 3.3.4.1 Phase Error This function is used to measure and phase errors. Only GSM signals can be analyzed.
Page 85
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description Captures the signal in sync with the IF signal (the lead- ing edge of the burst). EXT: Captures the signal in sync with the external trigger sig- nal. NOTE: The external trigger signal is input to the EXT TRIG connector on the rear panel.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description 3.3.4.2 Modulation Accuracy Performs the modulation analysis of 3PI/8 shift 8PSK. Auto Level Set Sets the internal reference level to an optimum value which de- pends on the measurement signal. Pressing this key executes the Auto Level Set function.
Page 87
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description Captures the signal in sync with the IF signal (the lead- ing edge of the burst). EXT: Captures the signal in sync with the external trigger sig- nal. NOTE: The external trigger signal is input to the EXT TRIG connector on the rear panel.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description 3.3.4.3 Tx Power Auto Level Set Sets the internal reference level to an optimum value in accor- dance with the measurement signal. The reference level is auto- matically adjusted when this key is pressed.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description TDMA Structure Selects the frame format. The value used with Trigger Delay is automatically calculated from the settings TDMA Structure and Slot Number. 156.25 BIT: Selects the format in which all slots consist of 156.25 bits.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description NOTE: HIGH mode cannot be selected when set to the multi burst mode. There is a possibility that measurements are not made in HIGH mode when a multi burst signal (which contains multiple bursts in a single frame) is input.
Page 91
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description Trigger Level Sets the trigger level. TDMA Structure Selects the frame format. The value used with Trigger Delay is automatically calculated from the settings TDMA Structure and Slot Number.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description 3.3.4.5 Wave Check Opens the menu used to display an IF or base-band signal waveform in time domain or FFT waveform. Time & FFT Displays an IF or base-band signal waveform in time domain or FFT waveform.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description Trigger Level Sets the trigger level. TDMA Structure Selects the frame format. The value used with Trigger Delay is automatically calculated from the settings TDMA Structure and Slot Number.
Page 94
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description FRAME: Displays the waveform of one frame. Trigger Source Sets the trigger signal. FREE RUN: Captures data using the internal measurement timing. Captures the signal in sync with the IF signal (the lead- ing edge of the burst).
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description 3.3.5 Modulation (When select to DECT) 3.3.5.1 Freq Deviation This function is used to measure frequency deviations. Only the signals compliant with DECT can be an- alyzed. Auto Level Set Sets the internal reference level to an optimum value in accor- dance with the measurement signal.
Page 96
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description NOTE: The external trigger signal is input to the EXT TRIG connector on the rear panel. Burst Search Toggles the burst search function (which uses the Software) on or off.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description 3.3.5.2 Timing Jitter Auto Level Set Sets the internal reference level to an optimum value in accor- dance with the measurement signal. The reference level is auto- matically adjusted when this key is pressed.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description Trigger Level Sets the trigger level. Full Slot Number Sets the full slot number of the signal to be measured Half Slot Number Sets the half slot number of the signal to be measured.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description Burst Search Toggles the burst search function (which uses the Software) on or off. Searches for a burst using the Software. OFF: Does not search for a burst using the Software.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description Parameter Setup Figure 3-33 Parameter Setup Dialog Box Trigger Source Sets the trigger signal. FREE RUN: Captures data using the internal measurement timing. Captures the signal in sync with the IF signal (the lead- ing edge of the burst).
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description 3.3.5.5 Wave Check Opens the menu used to display an IF or base-band signal waveform in time domain or FFT waveform. Time & FFT An IF or base-band signal waveform in time domain and FFT waveform are displayed.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description Trigger Slope Selects the polarity (leading or trailing edge) of a trigger signal. Trigger Level Sets the trigger level. Full Slot Number Sets the full slot number. Half Slot Number Sets the half slot number.
Page 103
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description FRAME: Displays a waveform corresponding to one frame. Trigger Source Sets the trigger signal. FREE RUN: Captures data using the internal measurement timing. Captures the signal in sync with the IF signal (the lead- ing edge of the burst).
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description 3.3.6 Sets parameters used for measurement and relationship between the channel number and frequency. 3.3.6.1 DC CAL Compensates for direct current components inside the circuit. 3.3.6.2 Channel Setting Sets the relationship between the channel number and frequency.
Page 105
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description Meas Mode Selects the measurement mode. BURST: Selected when there is one burst in one frame. MULTI-BURST: Selected when there are multiple bursts in one frame. CONTINUOUS: Measures a continuous wave. Although no...
Page 106
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description NOTE: Select WIDE, when measuring a signal with a large frequency error; and select NARROW when measuring one carrier selected from multiple carriers. WIDE: Selects the filter with a wide bandwidth.
Page 107
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description NOTE: When RF is selected for Input, Cont Auto Level Set takes effect only for Phase Error, Power vs Time, and Tx Power. Use the soft key Auto Level Set when adjusting the reference level.
Page 108
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description NOTE: To detect the position of a burst compliant with the standard, measure the burst in sync with the sync word. SYNC WORD: Measures the signal in sync with the sync word.
Page 109
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 3.3 Functional Description The RF input route is set. BASEBAND (I&Q): The IQ input route is set. The input signal magnitude is from 0.25V to 0.9Vp-p (±0.47V or less). Baseband Input Selects the coupling of signals.
Page 110
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4 REMOTE CONTROL REMOTE CONTROL GPIB Command Index This GPIB command index can be used as the index for Chapter 4. GPIB Command Pages GPIB Command Pages *CLS ............4-71 CHTBL1 ENBL........4-17, 4-19 *ESE ............
Page 129
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (7 of 56) Talker Request Function Listener Code Remarks Code Output Format BaseBand Input Setup BBINPUT AC BBINPUT? 0: AC (DECT) BBINPUT DC 1: DC...
Page 130
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (8 of 56) Talker Request Function Listener Code Remarks Code Output Format T-Domain Auto Level Set AUTOWFL Power TDPAUTOLVL Trigger Setup Trigger Source FREERUN...
Page 131
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (9 of 56) Talker Request Function Listener Code Remarks Code Output Format T-Domain Window Setup Power Window TDPWDO ON TDPWDO? 0: OFF TWDO ON...
Page 132
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (10 of 56) Talker Request Function Listener Code Remarks Code Output Format T-Domain Template Power Template TDPTMPL ON TDPTMPL? 0: OFF TDPTMPL OFF 1: ON...
Page 133
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (11 of 56) Talker Request Function Listener Code Remarks Code Output Format T-Domain Template Limit TDPTMPLBTM * TDPTMPLBTM? Level Power (dBm/W/dBµV) Judgment TDPJDG ON...
Page 134
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (12 of 56) Talker Request Function Listener Code Remarks Code Output Format ON/OFF Auto Level Set OORAUTOLVL Ratio Trigger Setup Trigger Source FREERUN OORTRGSRC FREE...
Page 135
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (13 of 56) Talker Request Function Listener Code Remarks Code Output Format ON/OFF Average Mode Ratio TRACE AVG OORAVGMD TRACE OORAVGMD? 0: Trace Avg...
Page 136
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (14 of 56) Talker Request Function Listener Code Remarks Code Output Format T-Domain Auto Level Set TDSAUTOLVL Spurious Trigger Setup Trigger Source FREERUN TDSTRGSRC FREE...
Page 137
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (15 of 56) Talker Request Function Listener Code Remarks Code Output Format T-Domain Save Table TDSSV Spurious SVSTBL * Integer (1 to 3) Init Table...
Page 138
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (16 of 56) Talker Request Function Listener Code Remarks Code Output Format Judgment T-Domain Spurious TDSJDG ON TDSJDG? 0: OFF TDSJDG OFF 1: ON...
Page 139
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (17 of 56) Talker Request Function Listener Code Remarks Code Output Format T-Domain Auto Level Set TDMAUTOLVL Due to Trigger Setup Modula- tion Trigger Source...
Page 140
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (18 of 56) Talker Request Function Listener Code Remarks Code Output Format T-Domain Average Times TDMAVGCNT * TDMAVGCNT? Integer Due to (1:OFF, 2 to 999)
Page 141
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (19 of 56) Talker Request Function Listener Code Remarks Code Output Format T-Domain Starts measurement Due to Due to Modulation TDMMEAS TDMMEAS? n<CR+LF>,f1,lp1,up1, Modula- j1<CR+LF>,..fn,lpn,...
Page 142
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (20 of 56) Talker Request Function Listener Code Remarks Code Output Format F-Domain Detector Power Normal TGTDET NRM TGTDET? 0: Normal Posi TGTDET POS...
Page 143
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (21 of 56) Talker Request Function Listener Code Remarks Code Output Format Parameter Setup F-Domain Power Detector Normal FDPDET NRM FDPDET? 0: Normal Posi...
Page 144
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (22 of 56) Talker Request Function Listener Code Remarks Code Output Format Auto Level Set OBWAUTOLVL OBW% OBWPER * OBWPER? Real Number (0.5 to 99.5)
Page 145
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (23 of 56) Talker Request Function Listener Code Remarks Code Output Format Measurement results OBWMEAS? f1, f2, f3, j1 f1: OBW Frequency f2: Lower side Fre-...
Page 146
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (24 of 56) Talker Request Function Listener Code Remarks Code Output Format Due to Average Times DTSAVGCNT * DTSAVGCNT? Integer Transient (1: OFF, 2 to 999)
Page 147
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (25 of 56) Talker Request Function Listener Code Remarks Code Output Format Result Type Due to Transient DTSRES ABS DTSRES? 0: Absolute DTSRES REL...
Page 148
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (26 of 56) Talker Request Function Listener Code Remarks Code Output Format Due to Auto Level Set DTMAUTOLVL Modulation Gate Setup TGTSETUP ON TGTSETUP?
Page 149
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (27 of 56) Talker Request Function Listener Code Remarks Code Output Format Due to Template Modulation Template DTMTMPL ON DTMTMPL? 0: OFF DTMTMPL OFF...
Page 150
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (28 of 56) Talker Request Function Listener Code Remarks Code Output Format Due to Average Times DTMAVGCNT * DTMAVGCNT? Integer Modulation (1: OFF, 2 to 999)
Page 151
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (29 of 56) Talker Request Function Listener Code Remarks Code Output Format Due to Result Type Modulation DTMRES ABS DTMRES? 0: Absolute DTMRES REL...
Page 152
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (30 of 56) Talker Request Function Listener Code Remarks Code Output Format Inband Auto Level Set SPRAUTOLVL Spurious Template Template SPRTMPL ON SPRTMPL? 0: OFF...
Page 153
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (31 of 56) Talker Request Function Listener Code Remarks Code Output Format Parameter Setup Inband Spurious Detector Normal SPRDET NRM SPRDET? 0: Normal Posi...
Page 154
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (32 of 56) Talker Request Function Listener Code Remarks Code Output Format Starts measurement Inband Spurious Inband Spurious SPRMEAS Starts measurement in the same mode...
Page 155
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (33 of 56) Talker Request Function Listener Code Remarks Code Output Format Outband Average Mode Spurious TRACE AVG FDSAVGMD TRACE 0: Trace Avg FDSAVGMD?
Page 157
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (35 of 56) Talker Request Function Listener Code Remarks Code Output Format Tx Power Average Times TXAVG * TXAVG? Integer (GSM) (1: OFF, 2 to 32)
Page 158
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (36 of 56) Talker Request Function Listener Code Remarks Code Output Format Power vs. Auto Level Set AUTOLVL Time Trigger Setup (GSM) Trigger Mode...
Page 159
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (37 of 56) Talker Request Function Listener Code Remarks Code Output Format Power vs. Average Times GPTAVG * GPTAVG? Integer Time PTAVG * PTAVG?
Page 160
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (38 of 56) Talker Request Function Listener Code Remarks Code Output Format Power vs. OffLev Unit Edit Time #0 dBm GPTUNIT0 DBM GPTUNIT0? 0: dBm...
Page 161
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (39 of 56) Talker Request Function Listener Code Remarks Code Output Format Power vs. Entering a set of times Time (GSM) GPTTMENT t0 to t4: Time...
Page 162
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (40 of 56) Talker Request Function Listener Code Remarks Code Output Format Phase Error Auto Level Set AUTOLVL (GSM) Trigger Setup Trigger Mode FREERUN...
Page 163
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (41 of 56) Talker Request Function Listener Code Remarks Code Output Format Phase Starts measurement Error Phase Error PHACC (GSM) Starts measurement in the...
Page 164
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (42 of 56) Talker Request Function Listener Code Remarks Code Output Format Modulatin Auto Level Set AUTOLVL Accuracy Trigger Setup (GSM) Trigger Mode FREERUN...
Page 165
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (43 of 56) Talker Request Function Listener Code Remarks Code Output Format Modulation Average Times TAVGMOD * TAVGMOD? Integer(1:OFF,2 to 200) Accuracy Starts measurement...
Page 166
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (44 of 56) Talker Request Function Listener Code Remarks Code Output Format Graphics Constellation GGPTYP INP GGPTYP? 0: Constellation Selection Constellation(Line) GGPTYP LIN 1: Constellation(Line)
Page 167
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (45 of 56) Talker Request Function Listener Code Remarks Code Output Format Demodulated Data Demodulated data output DEMOD? n<CR+LF>+d1$<CR+LF> +..+dn$<CR+LF> n: Number of output char-...
Page 168
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (46 of 56) Talker Request Function Listener Code Remarks Code Output Format Graphics Constellation EGPTYP INP EGPTYP? 0: Constellation Selection Constellation(Line) EGPTYP LIN 1: Constellation(Line)
Page 169
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (47 of 56) Talker Request Function Listener Code Remarks Code Output Format Data Output Constellation I-channel data output EGPHI? n<CR+LF>+d1<CR+LF>+ ..+dn<CR+LF> Constellation n: Number of output...
Page 170
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (48 of 56) Talker Request Function Listener Code Remarks Code Output Format Freq Auto Level Set AUTOLVL Deviation Trigger Setup (DECT) Trigger Mode FREERUN...
Page 171
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (49 of 56) Talker Request Function Listener Code Remarks Code Output Format Freq Measurement results Deviation Freq Deviation (DECT) When Bit DFMDEV? Pk1, Pk2, Fr...
Page 172
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (50 of 56) Talker Request Function Listener Code Remarks Code Output Format Timing Auto Level Set AUTOLVL Jitter RFP→PP (DECT) DTJLNK ON DTJLNK? 0: OFF...
Page 173
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (51 of 56) Talker Request Function Listener Code Remarks Code Output Format Timing Starts measurement Jitter Timing Jitter DTJIT (DECT) Starts measurement in the same mode...
Page 174
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (52 of 56) Talker Request Function Listener Code Remarks Code Output Format Tx Power Auto Level Set AUTOLVL (DECT) Trigger Setup Trigger Mode FREERUN...
Page 175
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (53 of 56) Talker Request Function Listener Code Remarks Code Output Format Tx Power Measurement results (DECT) Tx Power TXPWR? d1, d2 DTXPOW? d1: Tx Power(dBm)
Page 176
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (54 of 56) Talker Request Function Listener Code Remarks Code Output Format Power vs Measurement Mode Time Normal PTMOD NORM PTMOD? 0: NORM (DECT)
Page 177
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (55 of 56) Talker Request Function Listener Code Remarks Code Output Format Power vs OffLev Unit Edit Time #0 W DPTUNIT0 W DPTUNIT0? 0: W...
Page 178
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 4.2 GPIB Command Codes Table 4-12 TRANSIENT Key (56 of 56) Talker Request Function Listener Code Remarks Code Output Format Graphics Frequency vs Bit DFMGTYP BIT DFMGTYP? 0: Frequency vs Bit Selection...
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 5.1 NARROW/WIDE in the GSM filter mode TECHNICAL INFORMATION NARROW/WIDE in the GSM filter mode The characteristics of a filter used for signal reception (referred to as a receiving filter hereafter) are not ex- plicitly specified in the GSM standard.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 5.4 Measurement of a Signal Containing Multiple Bursts in a Frame (IN GSM) Measurement of a Signal Containing Multiple Bursts in a Frame (IN GSM) Figure 5-1 A Signal Containing Multiple Bursts in a Frame In GSM, the targeted slot of a signal, which contains multiple bursts in a frame as shown in Figure 5-1, can be measured by setting Meas Mode (in STD Setup) to MULTI-BURST.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 5.5 Measurement of a Signal Containing UP LINK/DOWN LINK Bursts in a Frame (IN DECT) Measurement of a Signal Containing UP LINK/DOWN LINK Bursts in a Frame (IN DECT) Figure 5-2 A Signal Containing Up Link/Down Link Bursts in a Frame In DECT, a signal, which contains uplink/downlink signals in a frame as shown in Figure 5-2, can be mea- sured.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 5.6 T-DOMAIN Power T-DOMAIN Power 5.6.1 Default values of Time Template The templates, which are used with T-Domain power measurements of the signals containing raising/fall- ing edges, are set as follows depending on communication systems and their power control levels: 5.6.1.1 GSM Template...
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 5.7 Phase Error Measurements Compliant With GSM Phase Error Measurements Compliant With GSM Graphic Display Ranges The GSM standard recommends that only the useful part should be measured in phase error measurements.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 5.8 Template Edit Function Template Edit Function In TRANSIENT mode, the user can change template. It is necessary to pay attention when entering template, because the data can be interpreted as a relative or absolute value, depending on the setting of Template Cou- ple to Power ON/OFF in the Config menu.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 5.8 Template Edit Function Set the template using the relative value to the average power. Figure 5-5 Template settings When you shift the template to the direction of Y axis using Shift X/Y function while the Template Couple to Power is set to ON, the relative value to the average power is: Relative value (set on the template) + Shifted data on Y axis.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 5.8 Template Edit Function 5.8.2 Template Setting in the F-Domain Measuring Mode In F-Domain measurement mode, the carrier frequencies depend on the channel numbers. As a result, use the offset frequency from the carrier frequency for template’s X axis data.
Page 192
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 5.8 Template Edit Function When Template Couple to Power is set to OFF, template (Y axis data) is interpreted as an absolute value. As a result, the template is made up of the data you entered.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 5.9 Measurement Parameter Settings in Due to Transient, Due to Modulation and Inband Spurious Measurement Parameter Settings in Due to Transient, Due to Modulation and Inband Spurious In TRANSIENT mode, any parameters are compliant with the communication standard when you specify the communication standard.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 5.9 Measurement Parameter Settings in Due to Transient, Due to Modulation and Inband Spurious Figure 5-10 Marker Edit Setting Peak marker is set using the Peak Marker Y Delta soft key in the Config menu.
Page 195
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 5.9 Measurement Parameter Settings in Due to Transient, Due to Modulation and Inband Spurious This method is used when the absolute value of the adjacent channel power cannot be measured. The ratio of the adjacent channel power to the carrier power can be measured only when Detector is set to Posi.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 5.10 Phase Error 5.10 Phase Error Phase Error is defined as shown in Figure 5-12, and the value is calculated using the following formula. value value Magnitude Error Figure 5-12 Mag Error, Phase Error, E.V.M.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 5.11 8 PSK Graph 5.11 8 PSK Graph When 3 pai/8 Shift 8PSK is selected to analyze EDGE signal modulation, 8PSK Constellation or 8PSK I(Q) Eye can be selected from the Modulation Accuracy menu.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 5.12 Tx Power Peak Factor 5.12 Tx Power Peak Factor When 3 PI/8 Shift 8PSK is selected to analyze EDGE signal modulation, measuring the Tx power displays the peak factor, which is calculated using the equation shown below.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 5.13 Block Diagram 5.13 Block Diagram This section shows the block diagram for the modulation analysis hardware. The Figure 5-15 shows the modulation analysis part. Therefore the spectrum analyzer part is simplified. The area inside the double lines is the block diagram for the spectrum analyzer, and the part outside that area rep- resents the modulation analysis hardware.
In the table, PV is abbreviation of performance verification. NOTE: 1. The R3267 Series with OPT63 to be tested should be warm up for at least 30 minutes before starting test. 2. Make sure that the test equipment used meets its own published specifications.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 6.1 General 6.1.4 Performance Verification Test Record Sheet The performance verification test record sheets are provided at the end of this chapter. The test record lists test specification and acceptable limits. Recommend that make a copy of this table, record the complete test results on the copy, and keep the copy for calibration test record.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 6.2 Performance Verification Test Procedure Performance Verification Test Procedure This section provides performance verification test procedure for R3267 Series OPT 63. Built-in calibration signal is used for performance verification. 6.2.1 Phase Error Measurement Accuracy for GSM...
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 6.2 Performance Verification Test Procedure On the R3267 Series, set the STD parameter as shown Figure 6-2. Figure 6-2 Setting of STD Parameter for Phase Error Accuracy Measurement (GSM) On the R3267 Series, press DC CAL, Phase Error and AUTO LEVEL.
On the R3267 Series, after preset, set control as follow: Center Frequency :30.050781 MHz On the R3267 Series, set the measurement parameter as shown in Figure 6-4. Figure 6-4 Setting of STD Parameter for Modulation Accuracy Measurement (EDGE) On the R3267 Series, press DC CAL to perform dc calibration.
On the R3267 Series, press keys as follows to perform auto level. Parameter Setup, AUTO LEVEL On the R3267 Series, press SINGLE for a single sweep. On the R3267 Series, after the single sweep has completed, record the result on the performance verification test record sheet. 6.2.3...
On the R3267 Series, after preset, set control as follow: Center Frequency : 30.228 MHz On the R3267 Series, set the measurement parameter as shown in Figure 6-7. Figure 6-7 Setting of Measurement parameter for Frequency Deviation Accuracy Measurement (DECT) On the R3267 Series, press DC CAL, Freq Deviation and AUTO LEVEL.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual 6.3 Performance Verification Test Record Sheet Performance Verification Test Record Sheet Model : OPT3264/67/73+63 S/N : Phase Error Measurement Accuracy for GSM Specification Result Items Measured Pass/ Min. Max. Value Fail Phase Error Accuracy (rms) Phase Error Accuracy (peak) 5°...
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual A.1 Messages APPENDIX A.1 Messages In this section, the messages that are displayed while the analyzer is being used are described. Code Messages Description System Error. Fatal Error occurred. Data area for the calculation is Cannot allocate the required mem- insufficient on the memory.
Page 211
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual A.1 Messages Code Messages Description Modulation Gain CAL error!(#200) Check 30 MHz CAL signal for connection. Modulation Gain CAL error!(#300) Check 30 MHz CAL signal for connection. Modulation Gain CAL error!(#110) Check 30 MHz CAL signal for connection.
Page 212
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual A.1 Messages Code Messages Description Cannot measure 3PI/8 Shift 8PSK This measurement function cannot be used to ana- signal. lyze 3PI/8 Shift 8PSK signals. Set Modulation Type to GMSK. To enable this measurement function, set Modula- tion Type to GMSK.
R3267 Series OPT63 GSM/DECT Measurement Option Operation Manual Alphabetical Index Y [dB/div] 20/10/5 ........3-22, 3-25, 3-65, 3-74 Y Scale [dB/div] 10/5/2 ......3-8, 3-9, 3-12, 3-31, 3-34, 3-43 Z-Field ............ 3-27, 3-82...
Page 220
IMPORTANT INFORMATION FOR ADVANTEST SOFTWARE PLEASE READ CAREFULLY: This is an important notice for the software defined herein. Computer programs including any additions, modifications and updates thereof, operation manuals, and related materials provided by ADVANTEST (hereafter referred to as "SOFTWARE"), included in or used with hardware produced by ADVANTEST (hereafter referred to as "PRODUCTS").
Page 221
LIMITED WARRANTY 1. Unless otherwise specifically agreed by Seller and Purchaser in writing, ADVANTEST will warrant to the Purchaser that during the Warranty Period this Product (other than consumables included in the Product) will be free from defects in material and workmanship and shall conform to the specifications set forth in this Operation Manual.
Page 222
CUSTOMER SERVICE DESCRIPTION In order to maintain safe and trouble-free operation of the Product and to prevent the incurrence of unnecessary costs and expenses, ADVANTEST recommends a regular preventive maintenance program under its maintenance agreement. ADVANTEST s maintenance agreement provides the Purchaser on-site and off-site maintenance, parts, maintenance machinery, regular inspections, and telephone support and will last a maximum of ten years from the date the delivery of the Product.
Page 223
SALES & SUPPORT OFFICES Advantest America Measuring Solutions, Inc. (North America) New Jersey Office 258 Fernwood Avenue, Edison, NJ 08837 Phone: (1) (732) 346-2600 Facsimile: (1) (732) 346-2610 Santa Clara Office 3201 Scott Blvd., Santa Clara, CA 95054 Phone: (1) (408) 988-7700 Facsimile: (1) (408) 987-0688 ROHDE &...
Need help?
Do you have a question about the R3267 Series and is the answer not in the manual?
Questions and answers