Page 2
Evoqua assumes no respon- sibility for the completeness of this information. Users are responsible for making sure that the product is suitable for specific applications. Evoqua assumes no liability for specific or indirect damage or consequential damage arising from the sale, resale or misuse of its products.
Improper use ..........12 General safety instructions......12 Design ........... 26 Personnel qualification ........13 Rivo™ Backboard 4 (basic housing)....26 Spare parts/components ........15 HMI (housing cover)........27 Extensions and modifications......15 Electrical power..........15 IT security............15 Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
Page 4
Start-up ..........60 6.14.2 Without Acknowledge........42 Initial commissioning and putting back into 6.14.3 ACK with reset ..........42 operation ............60 6.14.4 Simple ACK............42 6.14.5 Safety functions..........43 Shut-down ..........62 6.15 Status LED states ..........43 W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
Page 5
CE mark ............111 17.2 UKCA mark ..........111 Troubleshooting ........97 LICENSE AGREEMENTS ....112 Maintenance ..........98 13.1 Maintenance intervals ........98 13.2 Replace battery ..........98 13.3 Replacing the fuses on the CPU board..99 Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
(m/f/d) are not used in parallel in this instruction manual. Nevertheless, all personal nouns and pronouns are understood to apply equally to all genders. We apologize for any inconvenience this may cause. W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
The instruction manual must be passed You can download the digital version of the instruction on to third parties with the device. manual from the official website of Evoqua Water Technol- ogies GmbH. Original version of the instruction manual •...
10 Electrical connection data: Supply voltage and supply output 11 Available inputs and outputs: Supply voltage 12 Article number Fig. 1 For example, type plate 1 Device name 2 Series code/model code (e.g. Mod. E10) 3 Sales Order/position/serial number W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
Access data (user/password) for the initial login in the If you lose the recovery key and the user administration Service-Center menu. access data, please contact your contractual partner. See chapter Operation Service-Center. Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
Failure to comply may lead to fatal or serious injury and 4 App Default to significant material damage. 5 Service Default 6 Code for the Important System Information 7 Recovery Key (example) W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
Failure to comply may lead to serious material damage 2 Consequence of action and impair functionality. 3 Further consequence of action => result/goal of the action List • List/list item • List/list sub-item Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
• Observe safety regulations and the prescribed protective clothing for handling chemicals. • All instructions in the product data sheet for the dosing medium must be complied with. W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
No substances may be used other than those described in this instruction manual (chemicals and prescribed calibration chemicals). • All inspection and maintenance work must be carried out at the specified intervals. Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
Work not described in this instruction manual must be per- formed only by authorized personnel. W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
The operating company must train all persons handling device that affect safety. the system, device, components, substances etc. and inform them of possible hazards. It is the responsibility of the operating company to monitor personnel qualification. Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
Page 16
When assessing spe- cialist training, a period of several years’ employment in the respective field may also be taken into consideration. W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
The electronic module may only be put into operation when the housing is closed, and must be connected to protective earth. Power cables must be connected in accordance with the wiring diagram. Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
Inform the responsible per- using high-pressure steam! son/authority immediately of any detected changes (including any changes in the operating perfor- mance)! W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
Only approved calibration chemicals may be used • The device must not be exposed to frost or explosive environments. • The prescribed maintenance work must be carried • Use of original spare parts Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
If necessary, contact your regional or local authorities for details of collection points and options for separating and collecting waste. W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
• Observe the specified temperatures for transport 3.4.2 Expansion board for installation on the HMI and storage (see Technical Data). Article number Designation W3T583003 Rivo™ Com-Board 485 Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
• Observe the specified temperatures for storage (see Technical Data). • Regularly check the general condition of the device and the packaging. W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
0 ... 50 °C (32 ... 122 °F) Relative humidity < 80 %, non-condensing, moist environment Environment No direct sunlight, use indoors Atmospheric pressure 75 ... 106 kPa Max. working height (altitude) 2000 m Ambient noise <45 dB Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
Page 24
1x mA input for flow rate 0 ... 20 mA/4 ... 20 mA (electrically isolated) • 1x mA input for external setpoint or dosing factor (electrically isolated) • 0 ... 20 mA/4 ... 20 mA W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
Page 25
Typical use of the relay: Enabling contact for external alarm or activation of the dosing devices or dosing pumps. If the relay is used to switch inductive loads with direct voltage, the load must be equipped with a suppressor circuit against overvoltage (free-wheeling diode etc.). Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
Optocoupler output for connection of enabling inputs to dosing systems, dosing pumps, (only if Rivo™ Flex Mod electrolysis systems 2Rel-2DO is used) • Max. ext. power supply 24 V DC • max. current 20 mA W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
Galv. isolated up to 50 V relative to earth Rivo™ Com-Board 485 (optional) Rivo™ Com-Board 485 Article number W3T583003 Interface RS485 interface with Wallace & Tiernan protocol for connection to OPC server or control system for data visualization Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
15 Digital input 2 16 Digital input 1 17 Status LEDs of the digital inputs 18 24V DC connector 19 Status LED of relays K1/K2 20 Status LED 21 24V DC internal fuse W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
(ratio Controller outputs for positioners, dosing pumps, solenoid control) pumps, analog mA output are available. • quantity-proportional dosing of disinfectants with lin- earization of the actuator (with positioner) W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
1 Slot for Rivo™ Flex Mod - Mod 1 (not used) 2 Slot for Rivo™ Flex Mod - Mod 2 (not used) 3 Slot for Rivo™ Flex Mod 2Rel-2DO 4 Slot for Rivo™ Flex Mod 2AO-mA Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
Fig. 4 Ratio control • Positioner with feedback 1kOhm / 5kOhm / 0 bis 10 V / mA signal • Analog mA output (only with optional use of the Rivo™ Flex Mod 2AO-mA) W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
Page 33
If a dosing factor >100% is selected, the maximum dos- ing quantity is already reached at flow rates of less than 100%, i.e. it may not be possible to reach the desired dosing quantity with higher flow rates. Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
This keeps the dosing and therefore the current measurement stable. The function can be activated or deactivated once or permanently. W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
Page 35
This can either be the internally set the servomotor with feedback. dosing factor of the menu setting or the dosing factor as an external mA signal input or interface interface. Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
The control parameter Tp must always be adjusted to suit the pump employed: Dosing pump strokes/ Tp value up to 20 20 to 40 40 to 80 80 to 125 125 to 200 W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
With the help of a voltage-free enabling contact, e.g. external stop, it is possible to realize the various functions: • Alarm (freely configurable) • Ext. stop: Dosing is switched off. The positioner moves to 0 %. Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
See Chapter Interfaces . NOTICE For detailed information on the interfaces, see the sepa- rate installation manual “Rivo™ communication inter- faces”. You can request this installation manual from us or download it from our homepage. W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
TP. (pulse-frequency controller) Dosing pump with 2-point Analog output mA input Example: Cycle period TP = 100 s Output value Yout = 30% Duty cycle 30 s Off time 70 s Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
If an mA value of 3.4 mA is not reached in the setting 4 to 25... 30... 20 mA, this is detected as a line break and an error is gen- 10... 12... erated. 5... 6... W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
0% position and saves the corresponding values mea- sured for the feedback signal internally. The positioner runtime Ty is automatically determined during the calibra- tion process. During calibration, the positioner must not be unlocked for manual operation. Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
2Rel-2DO are not replaceable. If the internal power supply L1 and N/L2 is used for dos- ing pumps or other devices, the current consumption must not exceed the value of the selected back-up fuse. W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
The relay is only deactivated when the condition has • Latched alarms are displayed in red as messages. been eliminated and acknowledged. • The relay becomes inactive after acknowledgment if the condition is still pending. W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
No outputs active, or output based on Stop- Behavior. green illuminated RUN The module is in run mode. All outputs are active and are activated. This is the normal state during operation. Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
You can request this installation manual from is being carried out. us or download it from our homepage. • The firmware update can be downloaded free of charge from our homepage. W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
You can read off the currently installed firmware version of (optional). the CPU in the system menu under system informa- tion - Software information. NOTICE For information on installing a firmware update, see Chapter Firmware update via USB interface. Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
Connection to a network is possible using a patch cable (1:1) or a crossover cable (crossover network cable). The LEDs are fitted in the RJ45 socket. They display the inter- face statuses. W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
Page 49
LED (yellow) LED (green) Fig. 6 Ethernet connection The connection runs in Auto negotiation mode. The data transfer rate and full or half duplex are defined automati- cally with the connected switch/HUB. Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
The device operates with liquids. For this rea- son, DIN EN IEC 62368/60950 must be The assembly accessories included in the scope of observed when connecting the devices. delivery must be used for installation. W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
• Do not install the device outdoors. • Do not expose the device to strong vibration or impact, magnetic fields or electromagnetic radiation. • Provide sufficient aeration and ventilation. Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
These mounting fixtures are not 3 Fasten the electronics module to the solid wall at the included in the scope of delivery! bottom by the brackets using dowels and screws. W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
7.2.3 Installation drawing, Electronics Module ( Montageschiene / mounting rail ) Brackets for hooking onto tallow-drop screws. Alternative: Installation with top-hat rail BZ-1588_001 Mounting brackets for fix- ing the housing in position with tallow-drop screws Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
(4x5 mm, 2x6 mm) are included with the accessories. 3 Plug the Ethernet cable into the HMI as shown in the wiring diagram. Fig. 10 HMI (housing cover) 1 Battery insulator W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
When connecting system components (e.g. devices, motors, pumps) as well as when entering operating data, the system components must be switched off in order to prevent uncontrolled activation or incor- rect operation. Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
Page 58
Ambient temperature <30°C: Cables heat-resistant up to at least 60°C Ambient temperature <40°C: Cables heat-resistant up to at least 70°C Ambient temperature >40°C: Cables heat-resistant up to at least 80°C W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
If the module configuration is changed, a scan must be carried out. To do this, open Setup - System settings - General and press the ‘Scan’ (hardware) button. Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
1 Disconnect the Electronics Module from the power supply and check that it is de-energized. 2 Remove the housing cover of the Electronics Module. 3 Release the four screws on the metal cover of the HMI (Pos. 1). W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
Page 61
11 Fit the housing cover again and tighten by hand (to a maximum torque of 0.7 Nm ± 0.15 Nm). 12 Reconnect the power supply. 13 Switch the Electronics Module on or reconnect the power supply. 14 Configure the interface. Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
Risk of injury or death! lithium battery button cell on the HMI. Possible consequence: fatal or serious injury. Refit the housing cover. • The electronic module must not be operated with flammable liquids. W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
Page 63
9 Where applicable, configure signal input for external dosing factor. 10 Where applicable, perform initial positioner calibration. 11 Configure digital inputs as required. 12 Configure alarms as required. 13 Parametrize interfaces. Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
Possible consequence: fatal or serious injury. • Observe safety regulations and the prescribed protective clothing for handling chemicals. • All instructions in the product data sheet for the dosing medium must be complied with. W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
Bar graph display with limit value indication Control output/actuator feedback 10 Operation mode (example: Automatic mode) 11 Home key (main menu) 12 Back key 13 Login 14 Alarms - ACK 15 System menu Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
Icon blue = “Operating mode“ inactive Log out Icon green = “Operating mode“ active Log out Reset button Log in Alarm unwiderruflich quittieren Apply Alarm information Cancel Calibration Back button Modul information Home button Trend Service-Center W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
- App Default (user/password). Sample water stop Setting Users log in with the corresponding rights (user levels) in the “Login” menu. Three user levels (roles) with different Button Reset rights are available. Service-Center Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
Page 68
All settings 3 Click the symbol (system menu). • Setup 4 Click user administration. • Operation modes 5 Click the symbol (create user) and create further • Calibration user. Click to save. W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
NOTICE 4 Click to log in. Resetting access data If you forget your access data, you can reset the system at any time using the Recovery Key. See label “Important System Information“. Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
1 Call up the Home menu (if applicable, click the Home key). 2 Click “Application setting” and make the correspond- ing settings. Input or selection options Application settings Application name Input: individual W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
Home button. operation mode set under “next operation Press the arrow key to return to the previous level. mode”. • With man. dosing time = 0, the running time man. dosing is unlimited. Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
Input or selection options User administration Create a new user. Name Input: User name Password Input: Password Description Input: z.B. Name, Position Roles Select: Operator - Administrator - Maintainer admin => Edit the created users W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
Page 73
If ON is selected, the date and time are taken from the network. Timezone Select: Europe – Asia – America – Africa – Antarctica – Pacific – Australia – Atlantic – Indian Date Display: Year – Month – Day Time Display: Hours – Minutes Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
Page 74
Display (automatic) Display of the physical hardware address of the device in the Ethernet. Global unit settings Global Unit Setting Units of Measurement: Metric – Imperial Temperature Unit Select: °C – °F W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
Page 75
Select: Direct – Inverse Debouncing Input: max 1000 ms Controller function Input: No function – Stop – Max. Dosing – Constant – 2xYout – Ignore external setpoint / dosing factor Controller Input: Flow Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
Page 76
Select: 0 mA – 4 mA – 10 mA – 12 mA – 20 mA – 22 mA – Hold last value Measurement Select: No select – Dosing factor – Flow sehr lang W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
Page 77
Select: #0 – #0.0 – #0.00 – 0.000 Limits Dosing factor - Limit values Limit value I Input – delete Input – delete Hysteresis Input Limit value II Input – delete Input – delete Hysteresis Input Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
Page 78
Select: None – Simple ACK – ACK with reset Assignment Select: Digital In 2 – Digital In 2 Relay Assignment Select: Rel 3 BB00 S04 C01 – Rel 4 BB00 S04 C02 W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
Page 79
Operation Rivo™ I Controller Input or selection options Factory setings Reset system settings Reset Reset the device. Reset application settings Reset Reset the applications. Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
Page 80
Yaml name Display of current version Yaml version Display of current version Yaml hash Display of current version Serial number HMI Display of current version Serial number Device Display of current version W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
Output load error (AO 1), backboard-addr. 00, slot 05, componente: 01 10.7.2 User Log in/Log out NOTICE The users currently logged in are displayed in this menu. Login Log out User Input: User Password Input: Password Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
Interactive menu navigation. password. The password must be saved in a safe place. If it is lost, it can be reset in the Service Center with the Recovery Key (see Chapter Resetting a password) W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
Page 83
2CH Analyser: 2 Channel measuring system without control function 2CH Control PC: 2 Channel measuring system with control function Rivo I Control SC: Ratio control without measured value support Import/Export Menu for importing or exporting the device configuration of an identical device.
“Recovery” button appears in the login window. 1 Enter Recovery Key. 2 Click the "Reset device" button. => Device is reset. All passwords and settings are now deleted, and the device is reset to the factory settings. W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
To change, it is necessary to log in at Function buttons (selection option via symbols) the appropriate user level. Display of measured values - measured value - sensor signal - percentage (display only) Dosing (display only) Operation mode (display only) Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
Page 87
=> Cancel/Confirm => Flow - Calibration archive (Date, In1, Out1, In2, Out2) Lower Value Automatic display Switch off the sample water supply Upper value Automatic display Switch off the sample water supply Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
Page 88
Select: Diagram view controller output Sample water stop Select: 2h – 6h – 12h – 24h – 7d Back/forward button for diagram display Advance button for diagram display Select: Flow - Yout - Digital In 1 W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
Page 89
Unit Input: % Lower range Input: % Uper range Input: % Format Select: #0 – #0.0 – #0.00 – #0.000 Upper calibration point input Input: mA Upper calibration point output Input: % Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
Page 90
Select: Limit value 1 - Limit value 2 Limit value 1 Input - delete Input - delete Hysteresis Input: % Limit value 2 Input - delete Input - delete Hysteresis Input: % W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
Page 91
=> Display: Dosing quantity, Average dosage, Max. dosing, Min. dosing Dosing Statistics Reset Label Input: individual Unit dosing performance Input: g Time base dosing performance Select: Min. – h – Day Max. rate Input: g/h Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
Page 92
Linearization point 30% => Delete linearisation point Linearization point 40% Linearization point 50% Linearization point 60% Linearization point 70% Input Linearization point 80% => Delete linearisation point Linearization point 90% Linearization point 100% W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
Page 93
Select: Off – Once – Permanent Feed delay time Select: Hours – Minutes – Seconds Safety MAN. Mpde Activate/deactivate Source dosing factor Select: Internal – Dosing factor Dosing factor Input: % Zero threshold Input: % Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
Please contact service. cated module. E-IOC-104 Error Unknown hardware. Hardware could not Please contact service. be identified. E-IOC-105 Error Communication error in the indicated Check the Hardware. module. Please contact service. E-IOC-106 Error W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
Page 95
Hardware or electronic error in the Indi- Please contact service. cated module. E-RIV-118 Info Communication error in the indicated Please contact service. module. E-RIV-119 Error Input signal outside of measurable range. Check input signal and wiring. Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
Page 96
Please contact service. module. E-RIV-130 Error Actuator was unlocked. E-RIV-131 Info Communication error in the indicated Please contact service. module. E-RIV-132 Error Input signal outside of measurable range. Check the input signal and wiring. W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
Page 97
Please contact service. cated module. E-RIV-151 Info Note for a successful upgrade. Data was Message can be deleted by ACK. successfully migrated. E-RIV-152 Info E-RIV-158 Info Calibration data are outside the valid range. Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
Page 98
System messages Rivo™ I Controller E-RIV-159 Info Calibration data are outside the valid range. E-SYS-101 E-SYS-102 E-SYS-103 E-SYS-104 E-SYS-105 W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
Replace battery of the electronics module after 5 years. 7 Fit housing cover. 8 Turn the power supply on. 9 Set the date and time. No further settings are Article number Designation required. W2T555401 Battery, electronics module W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
Clean the touchscreen with a non-linting microfiber cloth. To remove heavy soiling, moisten a corner of the cloth and wipe the touchscreen. Then dry immediately using the dry side of the cloth. Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
Disposal must be effected in compliance with local, regional, national and international regulations. W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
The statutory requirements of the country in which the product is put into use apply. If necessary, contact your regional or local authorities for details of collection points and options for separating and collecting waste. Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
Optional accessories Article number Designation W3T557914 Rivo™ Flex Mod 2Rel-2DO Relay module for activation of dosing outputs and alarms W3T557912 Rivo™ Flex Mod 2AO-mA (mA signal output module, 2-channel) W3T583003 Rivo™ Com-Board 485 W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
L1 and N/L2 L2/N are for connection of externals loads. (Note prefuse!) 100...240V AC 24V DC 50/60Hz 0,4...0,3A Pre fuse: Pre fuse: max. C20A max. 4A AC << alternatively >> DC power consumption: 15W Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
Page 106
0-2/10V See chapter Relay outputs and technical data. Attention External voltage must 1) Shield grounded not be connected on one side to the digital inputs ! 5) contact open = DI activ W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
Page 108
L N PE switch is set to OFF ! Pulse triggered Note dosing pump Observe permissible contact load! See chapter Relay outputs ext. enable L N PE and technical data. input Dosing pump W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
Page 109
Wiring Diagram Rivo™ I Controller page 1 of 2 Connection Rivo Flex Module (optionally) Rivo Flex Mod 2 AO W3T557869 OUT1 OUT2 Load max. 500Ohm Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
Page 110
DO 3 DO 4 max. max. alarm signals 40V DC 40V DC 40mA 40mA Note (Without function) Observe permissible contact load! See chapter Relay outputs and technical data. F*) Fuse not replaceable W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
Page 111
Wiring Diagram Rivo™ I Controller Ethernet USB 1 Backboard ETH 1 RJ45 Backboard Danger! Do not unplug or plug in interface connectors while they are in operation! Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
Page 112
Terminals 5 – 7 (no potential reference) for connecting the shielding establish. An additional PE connection at the start or end of the bus. Danger! Do not unplug or plug in interface connectors while they are in operation! W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...
The following standards were applied: The following harmonized European standards were applied: • BS EN 61010-1:2010 • BS EN 61326-1:2021 • EMV/EMC: 2014/30/EU • BS EN 63000:2018 • NRL/LVD: 2014/35/EU • RoHS: 2011/65/EU and 2015/863/EU Evoqua Water Technologies GmbH | W3T597263 Issue 01-0624...
Rivo™ I Controller LICENSE AGREEMENTS The Electronics Module (Mod. E10) contains copy-right- protected software components covered by various Open Source licenses. Detailed information can be obtained from the Service-Center on the device. W3T597263 Issue 01-0624 | Evoqua Water Technologies GmbH...