Summary of Contents for DeFelsko PosiTector DPM L+ Dew Point Meter Logger Plus
Page 1
Dew Point Meter Logger Plus Instruction Manual 한국 어 ﻲ ﺑ ر ﻋ 中文 English WiFi...
Page 2
Introduction Relative Humidity (measured) Air Temperature (measured) Surface Temperature (measured) Dew Point Temperature (calculated) Ts-Td Surface minus Dew Point Temperature (calculated) Wet Bulb Temperature (calculated) CAUTION: To ensure optimal performance of the PosiTector DPM L probe, do not obstruct the airflow near the air temperature and humidity sensors.
Page 3
Quick Start PosiTector DPM L+ probes are designed to be used with the PosiTector App. To power up the PosiTector DPM L+ probe unscrew the top cover (counter clockwise), then press and hold the power button until the LED indicator turns Solid White. Replace the top cover. The probe is now visible to the PosiTector App within 10 m (30 feet).
Page 4
The user can now Configure WiFi (pg. 4), Start Logging (pg. 6), View Live Data (pg. 6), and View Logged Datasets (pg. 8). More detailed instructions on using the PosiTector App are avail- able in the Help File, located at www.defelsko.com/app-help...
Page 5
Configuring WiFi using the PosiTector App The PosiTector DPM L+ features WiFi, allowing the probe to automatically sync datasets to PosiSoft.net: the free to use, cloud-based storage service. WiFi settings are configured using the PosiTector App, while connected to the probe via Bluetooth. WiFi Info View information on the current WiFi settings, including network name and the status of the most recent sync to PosiSoft.net.
Page 6
Things device and is ideal for users who want to stream live readings to a central application or database. More detailed instructions and further examples of configuring WiFi Streaming can be found at www.defelsko.com/developer- resources#wifi-streaming Reset WiFi Clears all saved WiFi settings from the PosiTector DPM L+, including Network SSID, and passcode.
Page 7
Logging Mode Start Logging In Logging Mode, the PosiTector DPM L+ automatically records datasets into memory at user-selected time intervals. Select Start Logging from the PosiTector App or menu, then enter the desired logging measurement interval (between 1 minute and 8 hours) and sync interval, then select OK.
Page 8
subsequent datasets at the user-specified measurement Interval. The PosiTector DPM L+ does not need to remain connected to the PosiTector App to continue logging datasets. A maximum of 10,000 datasets (Ta, Ts, Td, Ts-Td, RH, Tw) can be stored in PosiTector DPM L+ memory.
Page 9
Logged batches can also be viewed on PosiSoft.net or PosiSoft Desktop (pg. 9-10). NOTE: Additional datasets cannot be stored into a previously logged batch. Alarm Mode Alarms Enabling Alarm Mode on the PosiTector DPM L+ automatically alerts the user by email when current climatic conditions exceed pre-set values.
Page 10
Powerful desktop software (PC/Mac) for downloading, viewing, printing, and storing measurement data. Download and install PosiSoft Desktop from our website: www.defelsko.com/posisoft Measurement data that has been synchronized to PosiSoft.net, can also be synchronized to PosiSoft Desktop for local storage of readings, and for producing fully customizable reports.
Page 11
For organizations with re-certification requirements, the PosiTector DPM L+ may be returned at regular intervals for calibration. DeFelsko recommends that customers establish calibration intervals based upon their own experience and work environment. Based on product knowledge,...
Page 12
Troubleshooting Most PosiTector DPM L+ conditions can be corrected by installing a fresh set of AAA batteries. In the rare event the probe becomes unresponsive, with the probe powered ON, press and hold the power button for approximately 30 seconds. The LED indicator will first turn solid red, and then will begin flashing red rapidly, at which point the button can be released, and the probe will perform a factory reset.
Page 13
See: www.defelsko.com/service For issues related to the PosiTector App, please refer to the Help File located at: www.defelsko.com/app-help This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not...
Page 14
U.S. and in other countries. Other brand or product names are trademarks or registered trademarks of their respective holders. Every effort has been made to ensure that the information in this manual is accurate. DeFelsko is not responsible for printing or clerical errors.
Page 15
Dew Point Meter Logger Plus 说明书 한국 어 ﻲ ﺑ ر ﻋ 中文 English WiFi...
Page 27
Dew Point Meter Logger 사용 설명서 한국 어 ﻲ ﺑ ر ﻋ 中文 English WiFi...
Page 28
소개 소개 상대 습도(측정) 기온(측정) 표면(측정) 노점 온도(계산됨) Ts-Td 표면 온도 - 이슬점 온도(계산됨) 습구 온도(계산됨) 주의: PosiTector DPM L+ 의 최적 성능을 보장하려면 기온 및 습도 센서 근처의 공기 흐름을 방해하지 마십시오. 체온이 판독값에 영향을 미칠 수 있으므로 손가락을 센서에서 멀리 하십시오. 프로브를...
Page 29
빠른 시작 빠른 시작 PosiTector DPM L+ 프로브는 PosiTector 앱과 함께 사용할 수 있도록 설계되었습니다. PosiTector DPM L+ 프로브의 전원을 켜려면 상단 커버의 나사를 푼 다음 (시계 반대 방향) LED 표시등이 흰색 점등 상태로 바뀔 때까지 전원 버튼을 길게 누릅니다. 상단 커버를 교체합니다. PosiTector 앱에...
Page 30
연결하면 프로브의 배터리 수명이 줄어듭니다. 사용하지 않을 때는 PosiTector 앱을 종료합니다. 사용자는 이제 WiFi 연결(p.4), 기록 시작(p.6), 실시간 데이터 보기 (p.8), 기록된 데이터 세트 보기(p.6)를 사용할 수 있습니다. PosiTector 앱 사용에 대한 자세한 설명은 www.defelsko.com/app-help 에 있는 도움말 파일에서 확인할 수 있습니다.
Page 31
PosiTector 앱을 이용해 WiFi 연결하기 PosiTector 앱을 이용해 WiFi 연결하기 PosiTector DPM L+는 WiFi 기능을 갖추고 있어 무료로 사용할 수 있는 클라우드 기반 스토리지 서비스인 PosiSoft.net에 프로브가 데이터 세트를 자동으로 동기화할 수 있습니다. WiFi 설정은 블루투스를 통해 프로브에 연결된 상태에서 PosiTector 앱을 사용하여...
Page 32
있도록 구성할 수 있습니다. 이 솔루션은 프로브를 인터넷 장치로 전환하고 실시간 판독값을 중앙 애플리케이션이나 데이터베이스로 스트리밍하고 싶은 사용자에게 이상적입니다. WiFi 스트리밍에 대한 상세한 지침 및 추가적인 예는 www.defelsko.com/developer-resources#wifi-streaming에서 확인할 수 있습니다. WiFi 재설정 네트워크 SSID 및 암호를 포함하여 PosiTector DPM L+에서...
Page 33
단위를 변경하면 프로브의 메모리에서 기존 판독값이 모두 삭제됩니다. 기록 모드 기록 시작 기록 모드에서는 PosiTector DPM L+ 프로브는 사용자가 선택한 시간 간격으로 데이터 세트를 메모리에 자동으로 기록합니다. PosiTector 앱 또는 메뉴에서 기록 시작(Start Logging)을 선택한 다음 원하는 기록 측정 간격(1분에서 8시간 사이)과 동기화 간격을 입력한...
Page 34
측정 간격 설정 측정 간격은 데이터 세트가 메모리에 얼마나 자주 기록되는지를 결정합니다. 프로브는 메모리에 최대 10,000개의 데이터 세트를 저장할 수 있습니다. 동기화 간격 설정 동기화 간격은 메모리에 있는 데이터 세트가 WiFi를 통해 PosiSoft.net로 동기화되는 빈도를 결정합니다. 자동 배치 생성 간격 자동...
Page 35
기록된 데이터 세트 보기 현재 활성 상태의 기록된 배치와 이전에 다운로드된 기록된 배치는 PosiTector 앱에서 확인할 수 있습니다. 배치 현재 활성 상태의 기록된 배치를 보려면 PosiTector DPM L+ 데이터 세트 프로브(p.3)에 연결하기만 하면 됩니다. 배치가 자동으로 판독 다운로드되어 표시됩니다. 숫자 표시된...
Page 36
배치에 대한 PDF를 생성합니다. 여러 배치를 포함하는 보고서를 생성하려면, 메모리(Memory) 메뉴에서 배치 보고서(Batch Report) 를 선택합니다. 배치 보고서는 메모리(Memory) 메뉴에 있는 배치 보고서 구성 (Batch Report Configuration)에서 사용자가 원하는 대로 설정할 수 있습니다. 더 자세한 사항은 www.defelsko.com/app-help에 있는 도움말 파일에서 확인할 수 있습니다.
Page 37
웹에 연결된 모든 장치에서 PosiSoft.net을 접속하여 측정 데이터 .pdf 보고서를 생성하고 DPM 모니터(DPM Monitor)에 접속하여 실시간 측정 데이터 차트를 볼 수 있습니다. 더 자세한 사항은 www.defelsko.com/posisoftnethelp에 있는 PosiSoft.net 도움말 파일에서 확인할 수 있습니다. PosiSoft Desktop 측정 데이터 다운로드, 확인, 인쇄, 저장을 위한 강력한 데스크탑...
Page 38
교정 증명서와 함께 출하됩니다. 재인증 요건을 가진 기관들의 경우, PosiTector DPM L+ 를 정기적인 주기로 반송하여 교정받을 수 있습니다. DeFelsko는 사용자가 자신의 경험과 작업 환경을 기반으로 교정 주기를 설정할 것을 권장합니다. 제품 지식, 데이터, 고객 피드백을 기반으로 교정일, 구매일, 수령일로부터 1년의 교정 주기가...
Page 39
참고: 10 m (30 피트) 내에 있는 프로브만 보입니다. 표면 온도 프로브가 표면에 없는 경우, 기온(Ta) 과 표면 온도(Ts)가 다르게 나타납니다: 표면 프로브가 공기 중에 있는 정상적인 조건에서, Ta와 Ts는 각 센서의 결합 허용 오차 내에 있어야 합니다. (일반적으로 1°C/2° F) 표면 프로브가 최근에 공기와 온도가 다른 표면에 접촉되어 있었다면, 차이가...
Page 54
Corporation registered in the U.S. and in other countries. Other brand or product names are trademarks or registered trademarks of their respective holders. Every effort has been made to ensure that the information in this manual is accurate. DeFelsko is not responsible for printing or clerical errors. IDPMPLUS-Z.v.1.0-0624...
Need help?
Do you have a question about the PosiTector DPM L+ Dew Point Meter Logger Plus and is the answer not in the manual?
Questions and answers