Low maintenance due to the Smart Sensor and solid state electronic design . The TD-120 contains five important components that it uses to make fluorescence measurements. • Light Source The TD-120 uses either near UV or deep UV light emitting diodes (LEDs), depending on the optics package chosen . The light sources emit photons over a wavelength range tailored to the application . •...
. AVERTISSEMENT: Un avertissement à faire preuve de prudence! Dans certains cas, des blessures ou des dommages à la TD-120 peut se produire si ces instructions ne sont pas suivies correctement . NOTE: An explanation or information of interest .
Design Information DESIGN INFORMATION This section includes TD-120 specifications and describes the monitor component parts. Figure 2.1 Monitor exterior The TD-120 monitor is not audible during normal operations . When the three-position ball valve is actuated, it briefly emits a sound of no more than 50 decibels. PHYSICAL SPECIFICATIONS Dimensions: 16 inches [406 .4 mm] H x 20 inches [508 mm] W x 5 .375 inches [136 .5 mm] D Weight: 24 lbs [10 .9 kg]...
Design Information TD-120 Oil in Water Monitor ELECTRICAL SPECIFICATIONS Power: 100–240 VAC 50 / 60 Hz, 1.3 A max 1 phase (inrush current not to exceed 40 A max) Four configurable dry-contact relays: 250 VAC, fused at 2 A Relays: or 28 VDC, fused at 1 .6 A NOTE: Benchmark International does not recommend placing heavy inductive loads on the relays .
TD-120 Oil in Water Monitor Installation The monitor displays the Home screen by default . Users can navigate to other function screens by using the keypad . The Home screen displays hydrocarbon concentration readings as well as any active alarms .
. MONITOR LOCATION The TD-120 monitor is intended to be used in a location that is sheltered from harsh ambient conditions such as direct sunlight, rainfall, and freezing temperatures. It must be affixed to a vertical wall or other rigid framework that is free from vibration . It is not intended for portable use . The monitor can be installed at altitudes up to 6,500 feet [2000 meters] .
. The best way to accomplish this is to sample from the center of the pipe . The sample outlet from the TD-120 should be piped to either an atmospheric drain with no back pressure or a pressurized system .
. DANGER: Make all connections at the TD-120 terminal blocks, as described in this section, before connecting to a power source . The following procedures assume that a power source for the TD-120 monitor has not been established .
Page 13
TD-120 Oil in Water Monitor Installation Figure 3.6.2 Enclosure Interior Service tablet Mounting nuts Backing panel USB cable To remove the service tablet Use the enclosure key to open the enclosure, as shown in Figure 3 .6 .2 Enclosure Interior prevent dust or dirt from entering the enclosure, do not allow the enclosure door to remain open longer than necessary .
Page 14
Installation TD-120 Oil in Water Monitor WARNING: Remove the service tablet carefully . Do not allow the tablet to hang from the USB cable . Make sure you grasp the tablet and its backing panel securely before pulling them out of the enclosure .
Page 15
TD-120 Oil in Water Monitor Installation Figure 3.6.5 Terminal blocks Power terminals (prewired) DC power in: 24 V- DC power in: 24 V+ Analog output 16) 4–20 mA out+ 17) 4–20 mA out- Relay terminals Relay 1 common Relay 1 normally opened Relay 1 normally closed...
Page 16
. To supply power to the TD-120 monitor After all wiring connections have been made at the TD-120 monitor, as described in the previous procedures, replace the service tablet with its backing panel in the enclosure . Tighten the four mounting nuts by turning them clockwise with an 8 mm nut driver. Do not exceed 6–8 ...
Page 17
TD-120 Oil in Water Monitor Installation Close and latch the enclosure door . Make connections at the power source . Power up the monitor with the switch or breaker you have supplied . IMPORTANT: After power has been supplied to the monitor, no further connections should be made or modified without using a voltage indicator or other appropriate device to verify that all power to the ...
Initial Start-Up and Settings TD-120 Oil in Water Monitor INITIAL START-UP AND SETTINGS Before start-up, the following items (not supplied with the TD-120) should be obtained . These items are widely available . • Process water (the water from your process, without the target hydrocarbon) •...
TD-120 Oil in Water Monitor Initial Start-Up and Settings TYPICAL INITIAL START-UP SEQUENCE Make sure a sample stream is connected and flowing through the Smart Sensor . See Section 2 Design Information for plumbing specifications. Sample outlet Sample inlet Flush line inlet When the monitor is first powered up, the Home screen appears, showing the firmware version, micro SD card status, and the current date and time . A ppm value is displayed in the center of the screen .
Initial Start-Up and Settings TD-120 Oil in Water Monitor FUNCTIONALITY TEST As part of the initial start-up, perform a functionality test. The functionality test verifies that the monitor is reading the sample stream properly . Press MAINT on the monitor keypad . In the Maintenance Mode menu, scroll to Functionality Test, and then press ENTER . The three-position ball valve setting will change from "inlet"...
TD-120 Oil in Water Monitor Initial Start-Up and Settings The Performance OK screen confirms that the monitor Smart Sensor is operating as intended . If the functionality test fails, it is usually an indication that the flow cell is fouled . Flush or manually clean the flow cell and repeat the test .
Page 22
Use the enclosure key to open the monitor enclosure . Start the service tablet by pressing the power button and holding it for 3 seconds. The tablet will first start the Windows operating system and then start the TD-120 service software . Power button The tablet displays the Home tab of a login screen by default. In addition to the login field, this tab displays an ...
TD-120 Oil in Water Monitor Initial Start-Up and Settings 4.4.1 USER ACCESS LEVELS The onboard TD-120 service software allows different levels of access to monitor parameters and configuration, depending on which login password is entered. Access Level Access Code User user Super User super The Early and High Alarm set points and relay assignments are available through User access . The following table includes the configuration parameters available to Super Users.
Initial Start-Up and Settings TD-120 Oil in Water Monitor Once the monitor is in Service Mode, you can log into the service software to configure alarm settings. For more information on using the service tablet, see Section 4 .4 Using the Service Tablet and Service Mode .
TD-120 Oil in Water Monitor Initial Start-Up and Settings Once the monitor is in Service Mode, you can log into the service software to configure the analog output settings . For more information on using the service tablet, see Section 4 .4 Using the Service Tablet and Service Mode .
Initial Start-Up and Settings TD-120 Oil in Water Monitor NOTE: To latch the Early or High alarms, you must log in with the Super User access code . After you have logged in, tap the Alarm Setup tab and then tap the checkboxes next to the desired relay assignments .
TD-120 Oil in Water Monitor Initial Start-Up and Settings NOTE: To make changes to the measurement, datalog, and averaging interval settings, you must log into the service software with the Super User access code . After you have logged in with the Super User access code, tap the Advanced tab .
USING THE SECURE DIGITAL (SD) CARD The TD-120 monitor uses a micro SD card to save monitor data to memory . The card is inserted in a port at the bottom of the CPU circuit board, shown in this photo without its cover. The micro SD card is also used to install firmware updates.
TD-120 Oil in Water Monitor Operations OPERATIONS This section details features and procedures used in normal operation of the TD-120 monitor . REVIEWING THE HOME SCREEN The Home screen appears as soon as the monitor is powered on. It displays the monitor firmware version and the current date and time . The center of the screen displays the target hydrocarbon concentration in your sample stream .
Operations TD-120 Oil in Water Monitor REVIEWING THE DIAGNOSTICS SCREENS Press DIAG to view the Diagnostic screens . Diagnostic values are stored as text in the monitor’s micro SD card memory . These values are updated at least once every 20 minutes or with each new event, such as an alarm .
Section 4 .3 Functionality Test . CALIBRATING THE SMART SENSOR If the TD-120 Smart Sensor was not factory calibrated as part of an optional customer care package, a Calibration Recommended message appears on the display screen, which means a site-specific two- point calibration must be performed, as described in the following procedure .
Page 32
Maintenance TD-120 Oil in Water Monitor WARNING: Never attach the syringe to the Luer-Lok® injection fitting until you are prompted by the service software to make an injection. Failure to depressurize the flow cell prior to mounting the syringe on the Luer-Lok® fitting may result in personal injury or death. AVERTISSEMENT: Ne fixez jamais la seringue à l’injection Luer-Lok® raccord jusqu’à ce que vous êtes invité par le logiciel de service pour faire une injection . L’échec au cours de dépressuriser la cellule d’écoulement avant le montage de la seringue sur le raccord Luer-Lok® peut entraîner des blessures ou la mort .
Page 33
TD-120 Oil in Water Monitor Maintenance The monitor averages readings over a 20-second duration and displays the time remaining . When the reading is complete, you have the option to proceed or retry the injection . If you would like to proceed, use a clean syringe to slowly inject 60 mL of your prepared calibration solution via the Luer-Lok® Injection fitting. Allow the Smart Sensor reading ...
à travers l’extrémité de la seringue . NOTES ON CALIBRATING Calibrate the TD-120 monitor when you have time to go through all the steps without interruption (about 10 minutes) . A calibration will not be saved to software memory until you tap Save and Use Calibration .
TD-120 Oil in Water Monitor Maintenance During calibration sequence, make sure to wait for blank value and calibration solution value readings to stabilize (within 10 percent of expected reading) before tapping the appropriate button on the service tablet . Do not allow calibration solutions to sit in the flow cell longer than necessary (approximately 2 ...
Page 36
Maintenance TD-120 Oil in Water Monitor Tools required (available in Calibration Kit, PN 100367) include: 10 mL graduated cylinder . 100-1000 µL micropipetter with disposable tips (blue) . 10–100 µL micropipetter with disposable tips (yellow) . 100 mL volumetric flask with stopper. Other tools you may need (not included with...
Page 37
TD-120 Oil in Water Monitor Maintenance 0–20 ppm calibration Mount a 200 µL disposable tip on the 10–100 µL micropipetter . 20–200 ppm calibration Mount a 1000 µL disposable tip on the 10–1000 µL micropipetter (blue tips, not as shown in picture) . 0–20 ppm calibration Adjust 10–100 µL micropipetter to draw 100 µL . 20–200 ppm calibration Adjust 100–1000 µL micropipetter to draw 1000 µL .
Page 38
Maintenance TD-120 Oil in Water Monitor 0–20 ppm calibration Transfer the 100 µL of hydrocarbon to the volumetric flask containing approximately 90 mL of 2-propanol . Dispense the hydrocarbon into the 2-propanol while maintaining the ultrasonic action (if necessary) . 20–200 ppm calibration Transfer the 1000 µL of hydrocarbon to the volumetric flask containing approximately 90 ...
Page 39
TD-120 Oil in Water Monitor Maintenance 0–20 ppm calibration To prepare 100 mL of Calibration Standard, dilute the following volumes of 1000 ppm stock solution into 100 mL of the site’s blank water using a second clean 100 mL volumetric flask. High Cal . 100 ppb 500 ppb 1 ppm 5 ppm 10 ppm 15 ppm 20 ppm Std .
Maintenance TD-120 Oil in Water Monitor Pour the prepared calibration solution into one of the calibration solution syringes while preventing liquid from exiting the other end, or suction the solution from a 250 mL beaker . Fill the second calibration solution syringe with the site’s blank water .
TD-120 Oil in Water Monitor Maintenance The warning screen that appears cautions that you are about to erase your existing calibration settings . Tap OK to proceed . Tap Confirm Reset to restore factory defaults, or tap Cancel to abort . Section 6 .1 Calibrating the Smart Sensorsor to recalibrate the Smart Sensor for a target hydrocarbon .
Page 42
Maintenance TD-120 Oil in Water Monitor 6.5.1 MANUAL FLUSH The flush line connection can be used to attach a clean water line or line from a pump or other cleaning system . Press MAINT on the monitor keypad . In the Maintenance Mode screen, scroll to Clean Water Flush and press ENTER .
Page 43
TD-120 Oil in Water Monitor Maintenance To clean the flow cell manually, press MAINT on the monitor keypad . In the Maintenance Mode screen, scroll to Manual Clean and press ENTER . The three-position ball valve setting will change from "inlet" to "flush" and then "closed." Inject clean water when prompted . Attach a clean syringe with 60 mL deionized water to the Luer-Lok®...
International représentant ou en appelant Benchmark International . Common Replacement Parts The TD-120 2-year spares kit, PN 106653, includes the following items . Note that part numbers are listed for reference only and are not sold separately . Part Number...
équipement annule toutes les garanties . TROUBLESHOOTING The TD-120 uses aspects of optics, chemistry, and electronics in its operation; therefore, it is important to collect all diagnostic data when troubleshooting . Diagnostic data is saved on the micro SD card as described in Section 4 .10 Using the Secure Digital (SD)
Page 46
Troubleshooting TD-120 Oil in Water Monitor Symptom Cause Solution Upsets in oil concentration do Monitor alarms have programmed Adjust delay to desired time . Note not immediately trigger alarm on delay times . that the release of alarms is not display screen .
Page 47
TD-120 Oil in Water Monitor Troubleshooting Symptom Cause Solution Water Temperature Alarm Excessive sample temperature . Reduce sample temperature to less than 50 °C . A/D Temperature Alarm Smart Sensor assembly Reduce temperature of Smart temperature is too high .
Page 48
Troubleshooting TD-120 Oil in Water Monitor Symptom Cause Solution Analog output not working or 4 mA and 20 mA values not Set monitor to Service Mode not producing correct value on properly set in the monitor . and access the service software .
TROUBLESHOOTING WORKSHEET All diagnostic data is saved to the micro SD card . NOTE: Use only the micro SD card supplied with the TD-120 . Other storage devices will not function properly . NOTE: Utilisez uniquement la carte SD fournie avec le TD-120 . D’autres dispositifs de stockage ne fonctionneront pas correctement .
General Arrangement Drawings TD-120 Oil in Water Monitor GENERAL ARRANGEMENT DRAWINGS Figure 8.1 Monitor front view Figure 8.2 Monitor bottom view PN 106407 TD-120 Oil in Water Monitor Operation Manual REV D...
Page 51
TD-120 Oil in Water Monitor General Arrangement Drawings Figure 8.3 Monitor left view Figure 8.4 Monitor right view Figure 8.5 Monitor back view (mounting bracket) PN 106407 TD-120 Oil in Water Monitor REV D Operation Manual...
Page 52
General Arrangement Drawings TD-120 Oil in Water Monitor Figure 8.6 Monitor connections TD-120 Oil in Water Monitor PN 106407 Operation Manual REV D...
Page 53
TD-120 Oil in Water Monitor General Arrangement Drawings Figure 8.7 Terminal blocks legend PN 106407 TD-120 Oil in Water Monitor REV D Operation Manual...
TD-120 Oil in Water Monitor INSTALLATION CHECKLIST This checklist outlines work that is required prior to and during installation of the TD-120 . If any of these requirements cannot be met, in some cases alternative installation procedures will provide reliable results;...
TD-120 Oil in Water Monitor Installation Checklist SAMPLE SUPPLY □ Monitor sample supply flow must be between 0.03 gal/min [0.1 L/min] and 0.79 gal/min [3.0 L/min] (0.26 gal/min [1.0 L/min]–0.52 gal/min [2.0 L/min] recommended). □ Monitor sample supply pressure must be between 10 and 100 psig [69 and 690 kPa(g)] . If the sample pressure is below 10 psig [69 kPa(g)], a sample supply pump must be specified. □ Monitor sample supply temperature must be greater than 32 °F [0 °C] and less than 122 °F [50 °C] .
Glossary TD-120 Oil in Water Monitor GLOSSARY AC Relays (1-4): The relays are part of the terminals block and provide connections for remote monitoring . Some alarms are pre-assigned to a relay and cannot be changed . Other alarms and functions can be reassigned, as described in Section 4.7 Configuring Relays .
Page 57
LED (light emitting diode): A semiconductor diode that emits light when a voltage is applied to it . Your TD-120 uses either near UV or deep UV lamps, depending on the optics package you have chosen . Light source: The origin for the light to which the sample is exposed .
Benchmark International . 11.2 TRADEMARKS TD-120, TD-5100 ECA, NexTD, E09 TD-4100XD, E09 TD-4100XDC, TD-4100XD, TD-4100XDC, TD- 4100C, TD-3100, TD-500, TD-500D, and TD-107 are trademarks of Benchmark International . Other brand and product names are trademarks and/or registered trademarks of their respective holders .
TD-120 Oil in Water Monitor Legal & Warranty and clearance fees, inland freight and handling fees from third parties to and from the factory are not covered by this warranty . All costs for removal, re-installation and for recalibration of the monitor are not covered under warranty and are the responsibility of the purchaser or customer .
Legal & Warranty TD-120 Oil in Water Monitor NOTE: Under no conditions should the monitor or accessory be returned without notice . Prior correspondence is needed for the following reasons . ▪ Many problems may be rectified on site. ▪ The nature of the problem must be defined in detail to speed the repair of the monitor. ▪ A RMA number must accompany the monitor. This number is used to track the repair cycle. NOTE: En aucun cas, le moniteur ou accessoire pourraient être renvoyés sans préavis . La correspondance préalable est nécessaire pour les raisons suivantes .
TD-120 Oil in Water Monitor Legal & Warranty Proceed exactly as for Warranty Service . Verbal or written support from Benchmark International's representative is free . Repair Service will be billed for time, materials, and custom clearance (if any) . Shipment to Benchmark...
Need help?
Do you have a question about the TD-120 and is the answer not in the manual?
Questions and answers