Do you have a question about the Permeater PD-1C and is the answer not in the manual?
Questions and answers
Summary of Contents for Gastec Permeater PD-1C
Page 1
Instruction Manual Calibration gas generator Permeater PD-1C Important Notices Read this manual thoroughly before use. Keep this manual with product in a safe place. 〒252-1195 8-8-6, Fukayanaka, Ayase City, Kanagawa Prefecture TEL 0467-79-3911 FAX 0467-79-3979 IM07PD1CE1...
Table of Contents 1. Introduction・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・3 2. For safe use・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・4 3. Contents of the package・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 8 4. Product overview・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・9 5. Names and functions of parts・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・10 6. Installation and setup・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・12 6.1 Location・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・13 6.2 Connection of the dilution gas・・・・・・・・・・・・・・・・・・・・・・・・・・・・・14 6.3 Adjustment of the dilution gas pressure・・・・・・・・・・・・・・・・・・・・・・・・・17 6.4 Connecting tubes at the outlets of the calibration gas・・・・・・・・・・・・・・・・・・・・18 6.5 Connecting the earth and Power Cord・・・・・・・・・・・・・・・・・・・・・・・・・・20 6.6 Filling water into the constant temperature water bath・・・・・・・・・・・・・・・・・・・・21...
Page 3
12.6 Service and maintenance・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・67 12.7 Lifetimes of the parts・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・68 12.8 Trouble shooting・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・69 12.9 Status Display List・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・70 12.10 Repair・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・73 13. Specifications・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・74 14. Flow Path Diagram・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・75 15. Block Diagram of Electronic Components・・・・・・・・・・・・・・・・・・・・・・・・・・76 Product Warranty and disclaimer・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・77 <Appendix> 1. Operation Manual for Permeation Tube 2.
Page 4
The instruction manual contains instructions for safe operation. Please read "2. Safety Precautions"carefully before use. Copyright The copyright of this publication is owned by GASTEC Corporation. No part of the contents may be reprinted or reproduced without permission.
When preparing calibration gases with this instrument, use permeation tubes (hereinafter referred to as P-tubes) and diffusion tubes (hereinafter referred to as D-tubes) of GASTEC products. GASTEC Corporation shall not be responsible for any accident that occurs in a situation that is contrary to the above items.
Page 6
⑤ This instrument is not explosion-proof. Do not allow calibration gas to come in contact with any part other than the calibration gas path. ⑥ P-tubes are sealed in intermediate packaging and storage(inner) packaging. To prevent inhalation of toxic gases, the intermediate packaging and storage(inner) packaging of the P-tube should be opened in a local exhaust ventilation system.
Page 7
or silica gel. ・Activated carbon and silica gel should be replaced in a timely manner as they become contaminated with long-term use. ㉑ To prevent electrical shock, use a power outlet with a protective ground connection. If the power cord does not connect to protective earth, connect the GND terminal on the backside of the main unit to protective earth.
Page 8
block the ventilation holes on the bottom. The water temperature may not reach the set value or the temperature accuracy may deteriorate. Also, the life of the internal electronic components may be shortened. ⑥ To maintain temperature uniformity in the TUBE HOLDER, check that the water level is within the H to L level of the WATER, and pour water when the water level is near the L level.
3. Contents of the package The package of this product contains the following accessories. Make sure you have all these together. Item description Rubber stopper Equipment for holding a standard thermometer inserted in a constant ① temperature water bath Power cord Cord for power supply* ②...
4. Product Overview Most environmental and industrial measuring instruments for measuring gas concentration use calibration gas to calibrate the scale and perform measurements. Therefore, reliable calibration gases are naturally required to obtain reliable measurement values. Calibration gases are also required for the measurement of odours by olfaction, research on gas analysis methods, tests on the effects of gases on animals and plants, and tests on the effects of specific gases on various materials.
Page 11
5. Names of parts and functions ① TUBE HOLDER P-tube and D-tube loading and unloading ports. P-tubes and D-tubes are loaded in using the provided tube holding cage. ② WATER INLET Water filling port for the constant temperature water bath. Use tap water (clean water). The capacity of the tank is approximately 1.5L.
Page 12
④ PRESSURE GAUGE Displays the pressure of the dilution gas supplied to the gas flow controller. Keep the pressure in the blue band range. ⑤ REGURATOR Control for the pressure of the dilution gas supplied to the gas flow controller ⑥...
6. Installation and set up Warning ① Since the calibration gas is flammable and toxic, place the Permeater in a local exhaust ventilation system or in a place with sufficient room ventilation. Poisoning or fire may occur. ② Provide a sink near this instrument. If you come in contact with toxic substances or if they get into your eyes, rinse them off immediately.
Note: ① Dilution gases should be supplied at pressures between 0.15 and 0.6MPa. Exceeding 0.6MPa may cause damage this instrument. ② Use nitrogen or air for the dilution gas. If other gases are used, the flow rate may be abnormal or internal corrosion may occur.
6.2 Connection of the dilution gas For the dilution gas, use nitrogen or air in a high-pressure gas container, or purified air compressed by an oilless compressor. The dilution gas should be regulated to 0.15 to 0.6 MPa with an external regulator and supplied to the DILUTION GAS (dilution gas inlet) of the Permeater.
Page 16
Connect the inlet of the dust filter (with a 200cm tube) to the outlet of the pressure regulator. The hose to be connected to the regulator is φ6 (outer diameter) × φ4 (inner diameter). Use a fitting that fits this size. Replace the dust filter when it is blocked.
Page 17
①Tighten the female nut on the short end of the tube to the male thread of the DILUTION GAS (dilution gas inlet) until it can be turned fully by hand. ②Use a 14mm spanner or the like to tighten the nut about 1/4 more.
0.15 - 0.6MPa Pressure Regulator Oil-free Permeater Compressor Activated Silica Dust Filter carbon DILUTION GAS *An example of a compressor connection 6.3 Adjustment of the dilution gas pressure Supply dilution gas, adjusting the PD-1C PRESSURE GAUGE indication to 0.15 to 0.3 MPa (in the range of the blue band).
After adjustment, push in and lock the lock ring. Note: When the regulator in PD-1C is pressurized to a pressure greater than 0.4 Mpa, the safety valve works to release the dilution gas into the main unit case. During the release of the dilution gas, a leakage sound of dilution gas is heard.
Page 20
③ Tighten the nut to hold the FEP tube. ④Check that the FEP tube does not come out easily by pulling gently. ⑤ When removing the FEP tube and connecting it again to CALIBRATION GAS (calibration gas outlet), cut approximately 10mm of the tube end and re-attach it. When discharging the calibration gas, remove the toxic substances by a pressure-free method before discharging them outdoors.
Warning Position the instrument and connect tubes so that the CALIBRATION GAS (calibration gas outlet) is away from the breathing zone when use. If only one CALIBRATION GAS outlet is used, close the other outlet. If it is left open, calibration gas will come out.
6.6 Filling water into the constant temperature water bath ① Check the water level sensor before filling the water bath. Press the POWER button to turn on the PD-1C, and STATUS lights up in red. (If the power does not turn on, check the leakage breaker on the rear panel.) ②After about 10 seconds, the buzzer sounds and the touch screen displays the start screen.
Page 23
⑤Check the WATER DRAIN (temperature water tank drain port) is closed. ⑥ Open WATER INLET (water filling port of the constant temperature water bath) and insert a funnel. ⑦ Make sure the inlet you are pouring water is WATER INLET. Watch the WATER gauge and pour water between levels H and L.
Page 24
Note: If water is added during operation, the temperature may drop and cause a "Heating/Cooling error". If this happens, turn the power OFF, and restart the main unit. When pouring water, reconfirm that the inlet is the WATER INLET. Pouring water into the TUBE HOLDER may cause it to malfunction.
7.Start Warning ① Since many of the calibration gases prepared by this system are toxic, skin corrosive, hazardous to health, or harmful to the environment, they should be used in a local exhaust ventilation system or in a place with sufficient indoor ventilation.
Page 26
not connect to protective earth, connect the GND terminal on the backside of the main unit to protective earth. To prevent electric shock and malfunctions, take care not to spill the water when pouring it into the ⑰ temperature water bath. Do not fill water exceeding H of WATER gauge. (During water injection, the water level gauge may not respond quickly to the water injection speed and may overflow.
Warning Do not load the P-tube/D-tube into the TUBE HOLDER immediately after startup. Wait at least 30 minutes after the dilution gas has flowed before loading. If the P-tube/D-tube has not been removed after the last use, there is a possibility that highly concentrated gas may remain.
7.2 P-tube mode The status "STAND BY" is displayed in the upper part of the LCD touch screen. The operation mode "P-tube" is displayed below the status display. To use another operation mode, touch "Menu" in the lower right corner, then refer to section 7.9. The following parameters are set in P-tube mode.
Page 29
Reset all setting values to zero Back to the previous screen Touch the button (K, Flow, etc.) or the parameter value to be set to display the keypad. Enter the value to be set. Warning The temperature setting of the P-tube should be set below the maximum temperature setting on the label of the P-tube storage container.
Page 30
If a value out of the settable range is entered, or if the automatic calculation results in a concentration or required dilution gas flow rate out of the specification range, "Out of range" is displayed. Check that each parameter is set correctly before starting the run. Note Make sure Conc.
7.3 D-tube single gas mode The status "STAND BY" is displayed in the upper part of the LCD touch screen. The operation mode "D-tube single gas" is displayed below the status display. To use another operation mode, touch "Menu" in the lower right corner, then refer to section 7.9. The following parameters are set in D-tube single gas mode.
Page 32
The D-tube must be kept at a stable temperature for 30 minutes to 24 hours (depending on Dr, see instruction manual of diffusion tube for the details) for the Diffusion Rate to stabilize. At this time, the dilution gas must continue to run to prevent gas accumulation. If need to save dilution gas, minimize consumption by setting the flow rate to a minimum of 0.2L/min while waiting for the permeation rate to stabilize.
Page 33
NOTE The ambient pressure setpoint is changed only when the above settings are made. Even if the ambient pressure changes during operation, the set value does not change automatically. If a value out of the settable range is entered, or if the automatic calculation results in a concentration or required dilution gas flow rate out of the specification range, "Out of range"...
7.4 D-tube multi gas mode The status "STAND BY" is displayed in the upper part of the LCD touch screen. The operation mode "D-tube multi gas" is displayed below the status display. To use another operation mode, touch "Menu" in the lower right corner, then refer to section 7.9. The following parameters are set in D-tube single gas mode.
Page 35
The D-tube must be kept at a stable temperature for 30 minutes to 24 hours (depending on Dr, see instruction manual of diffusion tube for the details) for the Diffusion Rate to stabilize. At this time, the dilution gas must continue to run to prevent gas accumulation. If need to save dilution gas, minimize consumption by setting the flow rate to a minimum of 0.2L/min while waiting for the permeation rate to stabilize.
Page 36
NOTE The ambient pressure setpoint is changed only when the above settings are made. Even if the ambient pressure changes during operation, the set value does not change automatically. If a value out of the settable range is entered, or if the automatic calculation results in a concentration or required dilution gas flow rate out of the specification range, "Out of range"...
7.5 Manual mode The status "STAND BY" is displayed in the upper part of the LCD touch screen. The operation mode "Manual" is displayed below the status display. To use another operation mode, touch "Menu" in the lower right corner, then refer to section 7.9. In Manual mode, only temperature and flow rate are set for use.
Page 38
Note When using D-tube, the dilution gas flow rate should be in the range of 0.20 to 8.00 L/min. If the flow rate exceeds 8.00 L/min, the Diffusion Rate may become unstable. If a value out of the settable range is entered, or if the automatic calculation results in a concentration or required dilution gas flow rate out of the specification range, "Out of range"...
7.6 Setting ranges for parameters Parameter Setting range Values to be entered Increment/decrement Coefficient of conversion of 0.050 - 1.700 mass of the gas into volume 0.050 - 1.700 0.001 K (L/g) 1.00 - 9.99 0.01 10.00 - 99.99 0.01 100.0 - 999.9 Permeation rate 1.00 - 990000...
7.7 Start Warning Before opening the TUBE HOLDER, allow the dilution gas to flow for at least 30 minutes. If the P-tube or D-tube used last time remains in the TUBE HOLDER, highly concentrated gas may have accumulated inside the TUBE HOLDER. Before starting, make sure that each tube connection is fastened and that there are no leaks on the gas flow path, ensuring that the gas is led to an appropriate ventilation system or hazardous substance filtering system.
Page 41
Close the TUBE HOLDER firmly. Warning Always work inside a local exhaust ventilation system when loading liquid samples into the D-tube. When handling P-tube/D-tube, be sure to use appropriate protective equipment such as protective glasses, chemical protective clothing, protective gloves or footwear, etc. because of the toxicity and irritating odor.
7.8 After use After use, open the TUBE HOLDER and remove the tube holding cage while keeping the dilution gas flowing. Warning Remove P-tube/D-tube from PD-1C when not in use. High concentrations of gas will accumulate. TUBE HOLDER must be opened with dilution gas flowing. High concentrations of hazardous substances may remain in the TUBE HOLDER.
Page 43
After removing the tube holding cage, touch and hold the "CLEAN" for 3 seconds to clean up the inside of the flow path. (The flow rate is 1 L/min, the temperature controller is turned OFF, and the pump is turned ON.) The status display on the screen will change to "CLEANING" and the STATUS lamp will light up orange.
7.9 Menu screen Touching the "MENU" button on the main screen of each operation mode displays the menu screen. The following functions are available. Mode "P-tube", "D-tube single gas", "D-tube multi gas", "Manual": Switching the operation mode (7.2) (7.3) (7.4) (7.5) Memory "Load"...
Page 45
③ The save screen is displayed. The memory list of the previous operation mode is selected. Four settings can be saved for each operation mode. ④Here is an example of how to save the settings in P-tube 1. Touch the number or the letter part of P-tube 1 to display the keypad.
Page 46
Recalling saved settings ① Touch ''Load'' button on the MENU screen. Touch “Load” ② Touch the button in the operation mode that calls up the setting. ③ Touch the button in the operating mode that calls up the setting. Touch an applicable numerical Touch the operation mode button or setting name.
7.11 Action to be taken if the P-tube exceeds the "maximum temperature" on the P-tube storage container label Warning If the temperature of the P-tube becomes too high, the internal pressure will increase and there is a possibility of high gas concentration being released due to rupture/explosion. If the temperature inside the TUBE HOLDER exceeds the "maximum temperature"...
8. Calculation of conditions for preparation of calibration gas The PD-1C automatically calculates the preparation parameters. On the other hand, the setting temperature of the water bath and the diluent gas flow rate can also be calculated using equations described in this section.
Page 49
<Calculation Example> Obtain the dilution gas flow rate F for the preparation of Sulphur Dioxide 1.00 ppm and 0.20 ppm with the temperature set to 35 °C. The label on the P-tube storage container supplied with the SO P-tube indicates the following information.
When T is between 35°C and 50°C, T is 35°C. Warning The above Equations (5) and (6) are applicable only in the temperature range from the lower to the upper limit of the temperature indicated on the label of the storage container supplied with the P-tube.
Page 51
The calibration gas concentration can be calculated with the following formula: C: Calibration gas concentration (ppm) No atmospheric pressure compensation Dr: Diffusion rate (μg/min) F: Flow rate of dilution gas (L/min) K: Coefficient for converting a gas weight into a volume (L/g) When using D-tube, the dilution gas flow rate should be in the range of 0.2 to 8.0 L/min.
Page 52
dilution gas flow rate F to 1.94 L/min. ・The calibration gas concentration when using two D-tubes can be calculated with the following formula: : Diffusion rate of the first tube (µg/min) : Diffusion rate of the second tube (µg/min) ・Calculations to estimate Dr values not described in the catalogue of Permeator, the instruction manual for the Diffusion tube or the No.
Page 53
: Calibration gas concentration (mg/m T : Ambient temperature P : Ambient pressure (hPa) : Standard atmosphere ・Methods for measuring diffusion rate Dr values Dr values can be determined by actual measurement. For higher accuracy or for substances not listed in the tables, it is recommended to measure the actual values under the actual conditions of use.
Page 54
The diffusion rate Dr listed in the catalogue, the instruction manual for the diffusion tube or the No.3200 technical document for the multi-component diffusion tube is the value under the standard atmospheric pressure. In the PD-1C, the pressure around the D-tube is the same as the pressure at the calibration gas outlet.
9. Calibration for flow rate, temperature and atmospheric pressure This section describes the calibration of flow rate and temperature. For calibration of atmospheric pressure, please contact Gastec or your Gastec representative. (Refer to Section 12.10 "Repair".) 9.1 Flow rate calibration (recommended every year) ①...
Page 56
⑧ The flow rate of the "Built-in sensor readout" displayed on the screen changes. Wait until both the "Built-in sensor readout" value and the reading of the standard flowmeter stabilize near the calibration point. ⑨ Adjust the "Set to" value to the reading of the standard flowmeter touching the "up/down" buttons, then touch the "Enter"...
9.2 Temperature calibration (recommended every year) Prepare a standard thermometer with its sensor part of approximately 6 mm in outer diameter. ① Remove the P-tube/D-tube from the TUBE HOLDER. ② Remove the WATER INLET lid. ③ Insert the standard thermometer into the supplied rubber stopper until the tip of the thermometer (sensor part) is about 16.5 cm from...
Page 58
➄ Touch the "Cal" button on the menu screen. Touch the "Cal" button ➅ The calibration point selection screen appears. Select the temperature "30°C" to be calibrated. Touch the "30°C" button ➆ The temperature of the constant temperature water bath changes, and that of the "Built-in sensor readout"...
Page 59
➈ The confirmation screen appears. Check again that the displayed value agrees with the reading of the standard thermometer, and if it does, touch the "OK" button, saving the calibration result and returning to the calibration point selection screen. If it is different, touch the "Cancel" button to return to the previous screen and adjust the value to the reading of the standard thermometer again.
9.3 Atmospheric pressure calibration Please contact Gastec or your Gastec representative for the calibration. The calibration by the customer is not included in the operation of the PD-1C. The calibration point selection screen appears touching the "Cal" button on the menu screen.
10. Other settings 10.1 Screen Settings Screen brightness, timeout, and touch panel can be adjusted. ① Touch the "Screen" button on the menu screen. Touch the "Screen" button ② The setup screen appears. Back to the "Menu" screen ・"Brightness": Adjustment of screen brightness ・"Screen time out": Time until the screen goes off.
10.2 Auto-tuning of temperature controller with PID control Temperature fluctuations associated with temperature control can be reduced. This is done at the time of our inspection and is usually not required. ① Remove the P-tube/D-tube from the TUBE HOLDER. ② Touch the "PID" button on the menu screen. Touch the "PID"...
11. Uncertainty of prepared gas concentrations Uncertainty is one of the quantitative measures of the degree of variation in measured values. Uncertainty is expressed as the width of the interval in which the final measurement value would be included with approximately 95% probability in many cases where the measurement values follow a normal distribution.
Page 64
Dilution gas flow rate: F The flow rate variation of the mass flow controller at a flow rate of 6.52 L/min is ±3.7%**. Assuming that this standard uncertainty follows a rectangular distribution, the standard uncertainty (u(F)) and the relative standard uncertainty (u(F)/F) are as follows: **Cited from Section 13 (Calculated based on the repeatability and resolution of the built-in measuring instruments and the uncertainty of the external measuring instruments.) From the above, the combined relative standard uncertainty (u(C)/C) is...
Page 65
Factors of uncertainty Measurement of diffusion rate: Dr The uncertainty of the mass measurement of the diffusion tube is estimated referring to the uncertainty of the electronic balance in use, listed on its calibration certificate issued by the manufacturer. For example, assume that the uncertainty per weighing is estimated to be 0.05 mg when measuring an object weighing approximately 20 g using an electronic balance with a minimum display of 0.01 mg.
12. Maintenance and Inspection Warning Do not open the case of the main unit. There is a risk of electric shock due to the 100-240 V AC wiring inside the main unit. 12.1 Replacement/drainage of water in the constant temperature water bath Even when this instrument is not in use, water should be changed at least once every three months.Water should also be changed if the water level gauge is dirty or if the water visible through the window of the water level gauge is dirty.
⑤Close WATER INLET. Refer to "6. Installation and Preparation of Permeaters" for water injection. 12.2 Replacement of Activated Carbon and Silica Gel for Purification of Diluting Gas When compressor air is used as dilution gas, it must be purified with activated carbon and silica gel. Activated carbon and silica gel should be replaced in a timely manner, as they become contaminated after long-term use.
B) Hold both ends of the fuse holder to release the latches and pull it out. C)Replace the two fuses. (The unbroken fuse is also likely to be damaged, so replace both fuses.) Fuse rating: Size 5×20mm 10A/250V AC time-lag type Use Littlefuse 0218010.MXP or equivalent.
Page 69
12.7 Lifetimes of the parts The parts listed in the table below will need to be replaced. The recommended replacement times are shown in the table below, but the time of failure may vary depending on the frequency of use and environment.
Page 70
12.8 Trouble shooting Try checking again, referring to the table below, before servicing. Troubles Possible causes / statues Actions to be taken PD-1C does not turn on The power cord is unplugged. ・Plug the power cord The earth leakage breaker is OFF. ・Check that there is no electric leakage, and turn the leakage breaker OFF once and then ON.
Page 71
12.9 Status Display List Normal operation Message on LCD Status Operating status To release the Automatic stop Start Start touchscreen error operation Cleaning STAND BY Waiting Possible Possible ✓ RUNNING In operation Possible ✓ CLEANING Cleaning ✓ possible Out of range An invalid value Enter the valid Not possible...
Page 72
Code Message on LCD Name Cause of the error Operating status when Action to Automatic stop Starting touchscreen the error occurs release the calibration error gas running Power supply unit Power supply voltage Abnormality Power ✓ ✓ ✓ ✓ ✓ ✓...
Page 73
・Warnings Code Message on LCD Name Cause of the Operating status when Action to release Automatic stop Starting touchscreen warning the warning occurs the warning calibration gas running Temp correction Temperature correction Abnormal Temperature Possible ✓ value load warning temperature calibration compensation value...
Page 74
12.10 Repair For repairs, please contact the distributor where you purchased the product or Gastec Corporation. Warning When transporting or carrying the PD-1C, remove the P-tube/D-tube from the TUBE HOLDER. If the P-tube/D-tube is left in the TUBE HOLDER with the dilution gas stopped, a high concentration of gas will accumulate.
Page 75
13. Specifications Calibration gas channel: 1 Gases to be prepared Permeation tube method S, NH , Cl , CH CHCl, CH SH, (CH S, C SH, CH , etc. The gas has a boiling point in the range of -60 to +30°C and is a stable high-purity substance. Diffusion tube method A liquid has a vapour pressure of 5-400 mmHg at 30-50°C and a stable, high-purity substance.
Page 76
UK:BS EN61326-1:2021 (For use in industrial locations) RoHS EU:EN IEC63000:2018 UK:BS EN IEC63000:2018 Warning Use a power cord that complies with the laws and regulations of the country/region where this instrument is used and meets the safety requirements of that country/region.
Page 77
15. Block Diagram of Electronic Components fuse power switch SELV 100~240V Inlet double pole double pole 5.0V 3.3V microcomputer and peripherals earth leakage breaker noise filter AC/DC power unit DC/DC down regulator mass flow controller 5.0V 5.0V normal operation / error LCD touchscreen water level sensor RS-485 driver...
Page 78
Gastec warrants that its products are free from defects in design, material, and workmanship and will comply with the specifications established by Gastec for a period of one year from the date of purchase. With respect to any defective product to which this Warranty applies, Gastec shall, at its sole option, either replace the defective product with a new product or refund the purchase price of the defective product.
Page 81
This instruction manual describes how to operate the permeation tube that is to be used together with Gastec calibration gas generator (PERMEATER, models PD-1B, PD-1B-2, and PD-1C). Please read this instruction manual and enclosed SDS carefully and fully understand the contents therein before using this unit.
Page 82
IM06PE8 WARNING ・ The product you have purchased contains an intermediate packaging, instruction manual, and SDS in the outer packaging. The P-tube is packaged in a storage container in the intermediate packaging with absorbent material. absorbent material intermediate packaging P-tube storage container ・...
Page 83
IM06PE8 ・ P-tubes should be sealed and stored in the storage container provided with the P-tube. The maximum storage temperature is indicated on the P-tube storage container as "-5°C or lower" or "25°C or lower". The minimum storage temperature is -25°C. If P-tubes are stored at temperatures lower than -25°C, there is a possibility of temporary leakage of filling gas due to the difference in expansion coefficients of the tubes and crimps when they are returned to room temperature.
Page 84
Therefore, a calibration gas can be obtained by blending with a constant flow of a dilution gas. Calibration gas concentration as well as dilution gas flow calculation can be made from the www.gastec.co.jp website, go to technical information, then preparation conditions calculation for simple instructions.
Page 85
IM06PE8 value of the permeation rate at each temperature is indicated on the label of the storage container supplied with the P-tube for each bottle; use this Pr value for the calculation. The effective length of a P-tube is the length of the part through which liquefied gas permeates; the label on the storage container supplied with the P-tube indicates the effective length for each P-tube, so use this L value for calculations.
Page 86
IM06PE8 Calculations to estimate Pr values at temperatures not listed on the label. (only when PD-1C is used) Within the temperature ranges listed on the label, the Pr at temperatures between the Pr values can be estimated in 1°C increments. Larger temperature differences between Pr values tend to increase the error in the estimated Pr.
Page 87
IM06PE8 : Calibration gas concentration (mg/m T : Ambient temperature P : Ambient pressure : Standard atmosphere Methods for measuring permeation rate Pr values Pr values can be determined by actual measurement. For higher accuracy, it is recommended to measure the actual values under the actual conditions of use. When using the PD-1C, using the Manual mode will display only the temperature and flow rate on the main screen, making it ideal for measuring the diffusion rate Dr value.
Page 88
IM06PE8 3) Time necessary for a P-tube to stabilization It takes about 24 hours for the P-tube to reach the specified permeation rate after it is placed in the TUBE HOLDER. After this time has elapsed, prepare the calibration gas. 4) Procedures after use After use, open the lid of the TUBE HOLDER with dilution gas flowing, and remove the tube holding cage.
Page 89
IM06PE8 is stored or used in the presence of moisture. If the crimps are corroded and deteriorates, the Pr value will be higher than the specified value. During storage, pay attention to the color of the absorbent (green gel) in the storage container and replace the green gel as necessary if the absorption capacity decreases.
Page 90
IM06PE8 caustic when breathed. Perform the above procedures under a local exhaust ventilation system. When disposing of the adsorbent material, dispose it as industrial waste or incinerate it. Some may emit hazardous gas by incineration. In such cases, use a chemical incinerator which has flue gas treatment equipment.
Page 91
IM973100E1 Operation Manual for Diffusion Tube Foreword This operation manual describes how to operate the diffusion tube, model No. 3100 that is to be used together with the calibration gas generation unit (PERMEATER, models PD-1B, PD-1B-2, and PD-1C). Please read this operation manual carefully and fully understand the contents therein before using this unit. We also recommend keeping this manual on hand while you use this unit.
Page 92
The diffusion rates listed in Tables 1 and 2 at the end of this document are the values actually measured by Gastec. For higher accuracy or for substances not listed in the table, it is recommended that actual measurements are carried out under the actual conditions of use. When using D-tube, the dilution gas flow rate should be in the range of 0.2 to 8.0 L/min.
Page 93
IM973100E1 The calculation for the use of a D-10 type D-tube is as follows To prepare C=5.0 ppm when using PD-1B series, set the water bath temperature to 30 °C, use D-10 type D-tube, and set the dilution gas flow rate F to 2.0 L/min. When using PD-1C, set the dilution gas flow rate F to 1.94 L/min.
Page 94
IM973100E1 : Calibration gas concentration (mg/m T : Ambient temperature P : Ambient pressure : Standard atmosphere Methods for measuring diffusion rate Dr values Dr values can be determined by actual measurement. For higher accuracy or for substances not listed in the table 1 and 2, it is recommended to measure the actual values under the actual conditions of use.
Page 95
IM973100E1 a large-diameter activated carbon bed that does not cause backpressure. Then vent to the outdoors using a suitable ventilation system. 1) When injecting the sample liquid into the D-tube, ensure that the liquid does not adhere to the inner walls.
Need help?
Do you have a question about the Permeater PD-1C and is the answer not in the manual?
Questions and answers