1.4 OPERATING ENVIRONMENT The BTL-08 Spiro Pro is intended to be used primarily in hospitals, but it can also be used in clinics, medical centers or wherever the spirometric examination is performed. The BTL-08 Spiro Pro is not intended for home-use.
Intensive exercise / movement activity 30 minutes or less before the examination Overeating or the consumption of a big meal two hours or less before the examination Tight-fitting clothing that can restrict the maximum breath [2] BTL-08 SPIRO – USER’S MANUAL I PAGE 4 OF 20...
USB to mains voltage fuse connect PC or switch external printer Manufacturing and type plate and clamp for device shielding are located on the bottom cover of the device. BTL-08 SPIRO – USER’S MANUAL I PAGE 5 OF 20...
Page 6
2.2 DESCRIPTION OF PNEUMOTACHOGRAPH plastic mouthpiece spirometric sensor (with a metal paper measuring mesh mouthpiece inside) silicon seal bacterial filter button to pneumotachograph release body spirometric sensor communication cable BTL-08 SPIRO – USER’S MANUAL I PAGE 6 OF 20...
We recommend that you keep the device packaging for possible further transport of the device. The BTL-08 Spiro Pro device is cooled by natural and forced air circulation. The cooling vents are located on the bottom side of the device and around the printer. These vents must not be covered. Do not put the device on a soft surface as this could block or limit the air circulation to through bottom cooling vents.
Page 8
Pressing the accept button confirms the acceptance of the recorded manoeuvre and that it is saved. If the manoeuvre is unsatisfactory, press the start button again to record a new manoeuvre. BTL-08 SPIRO – USER’S MANUAL I PAGE 8 OF 20...
Page 9
Note: Normally, the predictive norm for predicted values is a part of the profile. The selection remains valid only if you start the examination immediately afterward, so it is necessary to select the patient before using this option. contrast BTL-08 SPIRO – USER’S MANUAL I PAGE 9 OF 20...
Page 10
Note: If the device does not respond to any buttons (including on / off) and no time-consuming operation is running (such as print, export, etc.), reset the device by pressing the reset button (see Chapter Description of Unit). BTL-08 SPIRO – USER’S MANUAL I PAGE 10 OF 20...
5 CALIBRATION The accuracy of the air flow through the BTL-08 Spiro Pro is the basic precondition for reliable and correct results of your spirometric examinations. The BTL-08 Spiro Pro has three mechanisms for the ensuring of the maximum measuring accuracy possible: calibration, calibration check and adjusting zero flow. Zero flow adjusting is automatic process, whereas calibration and calibration check has to be performed by user.
Connect the pneumotachograph to the BTL-08 Spiro Pro unit and switch it on. Firmly connect the pneumotachograph to the output of the calibration syringe and make sure that the connection is tight.
Parameters. 6.1 RECORD A SPIROMETRIC EXAMINATION Connect the pneumotachograph to the BTL-08 Spiro and turn on the device. Check the cleanness of the spirometric sensor and the measuring mesh (see Chapter Maintenance). Make sure that particular spirometric sensor is calibrated. If not, please run the calibration (see Chapter Calibration).
Page 14
The following screen enables the operator to print out the final report on printer. BTL-08 SPIRO – USER’S MANUAL I PAGE 14 OF 20...
Page 15
15. Pressing the results button will display the table with the results of the entire examination and the comparison to the predictive values. To return to the previous screen with the table of all manoeuvres, press the back button. BTL-08 SPIRO – USER’S MANUAL I PAGE 15 OF 20...
Press the back button to return to the examination results screen where you can see comparison of PRE and POST stage. Record a bronchoconstrictor test Select the profile with the required protocol for examination. Record the PRE stage in the standard way (see Chapter Record a Spirometric Examination). BTL-08 SPIRO – USER’S MANUAL I PAGE 16 OF 20...
Wu 1961, Ip 2006, Polgar 1979, Wang Yang 2013, Perez Padilla 2003, Platino 2006, PDPI 2004, Thai 2000 Stored records User defined profiles unlimited Automatic interpretation ATS, BTS, Enright, ATS + GOLD BTL-08 SPIRO – USER’S MANUAL I PAGE 17 OF 20...
Page 18
Internal chemical supplies Battery lithium battery CR2430 Lead-acid backup battery 1x 12 V / 1.2 Ah, 48 x 97 x 52 mm, maintenance - free Backup battery capacity up to 30 prints BTL-08 SPIRO – USER’S MANUAL I PAGE 18 OF 20...