Page 1
Issued on: 2020.12.10 Date of revision MODEL High-pressure syringe pump HSP-101 High withstand voltage Syringe Pump Instruction Manual ■ Please read this instruction manual thoroughly to ensure correct use. In particular, be sure to read "To use safely" before use and use it correctly.
Page 2
Preface This manual explains the handling of HSP-101. CAUTION (1) Copying the whole or part of this manual without notice is strictly forbidden. (2) The author reserves the exclusive right to make changes/modifications to this manual. (3) The contents of this document have been prepared with utmost care. However, if you have any doubt or error, Please contact us if you notice anything, such as omissions.
Page 3
Revision history The manual revision symbol is added to the tail of the figure number above the cover. Revised Figure No. 162040110 Revised Date of revision Revised parts and contents symbols 2020.12.10 First Edition Printing...
Table of Contents HSP-101 Instructions for Use SAFETY PRECAUTIONS ··································································· 1 [1] Before using the chapter 1-1. Check the main unit and attached items. ······································ 1-1 1-2. Description and Function of Each Part ················································ 1-1 [2] Chapter Operation 2-1. How to start up the power supply ·············································· 2-1 2-2.
SAFETY PRECAUTIONS (1)Labeling and Meaning for Safe Use In this manual, the following symbols are used to indicate the operation and handling of products that are hazardous in order to prevent harm to customers and other people and damage to property.
Page 6
SAFETY PRECAUTIONS Safety Instructions Warning Do not use them if you notice any abnormalities such as smoke, strange odors or noises. Otherwise an electric shock or fire could occur. Immediately turn off the power switch and disconnect the power plug from the outlet.
Page 7
SAFETY PRECAUTIONS Warning Do not block the ventilation slots on the device. Blocking the ventilation holes may cause heat inside, which may cause a fire. Space Do not place any objects on or near the ventilation slots. (special attention should be paid to flammable materials such as paper and plastics.) Do not place the product in a narrow room, in a box, or in any other place that is poorly ventilated.
Page 8
SAFETY PRECAUTIONS NOTE Do not place the product in an unstable place (e.g., on an unstable kitchen or on a slanted place). Doing so may cause the product to fall or fall, resulting in injury. Bruisi Do not place heavy objects on the device or ride on it. Doing so may cause the product to collapse or fall out of balance, causing injury.
Page 9
SAFETY PRECAUTIONS NOTE Do not place the product in a place subject to vibration, such as a place where vibration from another machine is transmitted. Failure to do so may result in an injury due to the product moving or dropping. Bruisi To prevent fire, locate the power outlet to which the power plug is connected near the equipment.
[1] Instructions for HSP-101 before use 〔1〕章 Before use This chapter explains what you should be aware of when using the device. Be sure to read this manual before use, and use the product safely and properly. 1-1. Check the main unit and attached items.
Page 11
[1] Instructions for HSP-101 before use Description of each part of the main unit 1.Syringe clamp Fix the outer cylinder of the syringe. 2.V-shaped holder Holder for holding the syringe. 3.Push block Push the pusher (plunger) of the syringe to deliver the solution.
Page 12
[1] Instructions for HSP-101 before use 1-3 Confidential...
[2] HSP-101 Operation Manual 〔2〕章 How to operate This equipment is grounded through a 3-conductor power cord with a ground line. Be sure to connect the power plug to an outlet with a protective grounding terminal. NOTE In addition, connect the protective earth terminal to the GND terminal of the high-voltage power supply or, in the case of our equipment, to the GND 2-1....
[2] HSP-101 Operation Manual 2-3. How to Set Various Parameters 1.Parameter changes are as follows: Press and hold the "SEL" key for approximately 1 second. RESE When the unit enters the parameter setting screen, the [] and [] lamps blink at the same time.
Page 15
[2] HSP-101 Operation Manual 2. Volume of solution delivered (FeedRate) change window FeedRate>>10.0[ml/h] The arrow indicates the >>> → FeedRate (the Diameter 10.0[mm] amount of solution delivered). ---Volume(Use/Set)--- The [] and [] lamps are blinking. 0.000 / 10.0[ml] Use the up and down [↑↓] key to set the dispensing volume of the solutions.
Page 16
[2] HSP-101 Operation Manual Use the [←→] buttons to move between the selected digits. Diameter>>10.0[ml/h] ↓ RESE Diameter>>10.0[ml/h] ↓ Diameter>>10.0[ml/h] When the setting is completed, press and hold SEL again for about 1 second to change the volume of solution (Volume Set) to be delivered.
[2] HSP-101 Operation Manual 2-4. How to start and stop 1.About Starting The cumulative amount of the delivered solution (Volume Use) is restored to zero by pressing RESET key for at least 1 second. Use this when you want to start the accumulated value from zero, such as when a new solution or syringe is replaced.
[3] How to Maintenance HSP-101 Instructions 〔3〕章 HOW TO MAINTENANCE Care should be taken regularly so that the device can always be used in good condition. 3-1. Cleaning (1) External ① Turn off the power and brush off the dust with a soft brush.
Page 19
[4] Chapter Trouble?If you think, HSP-101 Instructions for Use 〔4〕章 Don't you think it is a failure? In the event of an abnormal operation, first read this chapter and take appropriate action. If any of the above measures do not improve, or if an event occurs without a statement, record the following and contact our business or service.
[4] Chapter Trouble?If you think, HSP-101 Instructions for Use 4-1. Unusual events and addressing Symptoms Possible causes Corrective action Is the power plug not inserted or in Insert the power plug. Turning on the "POWER" an incomplete condition? switch on the rear panel...
Page 21
Product warranty regulations and services Provisions for product warranties The warranty period for this product is one year from the time it is shipped from the factory. Products that fail during this period are at our factories or at the business sites where this product was sought. Repair will be performed free of charge at the facility of the distributor.
Page 22
Only those with expert knowledge and able to perform safety management are included. We are not responsible for any accidents or damage caused by access or use by the general public. A spiral. MEC CORPORATION Head Office 〒838-0137 196-1 Fukudo, Ogori-shi, Fukuoka Tel:0942-72-7266 Fax:0942-73-3545 [Homepage URL] http://www.mecc.co.jp...