    - TABLE OF CONTENTS - Model 1020 Standard Electric Console Instruction Manual System Overview and Intended Use ..........1 Applicable Accessories and Part Numbers ........1 Symbol Definitions ..............2 - 4 Environmental Parameters ............... 4 Important Information ..............5 - 8 Warnings ..................
  • Page 4: System Overview And Intended Use

    1641 Electric SmartDriver or throttleless instruments using a foot pedal. The MicroAire 1020 is an instrument control console for driving small power instruments. It generates the signals necessary to run the motor in these instruments at the desired speed and direction as controlled by the operator.
  • Page 5: Symbol Definitions

    - CONSOLE ICONS - MARKINGS DEFINITION Power “OFF” Power “ON” Speed Indicator Speed Control Handpiece Reverse Indicator - CONSOLE MARKINGS - MARKINGS DEFINITION Attention, See Instructions for Use Product Catalog Number Authorized European Representative Product Serial Number Lot Number; Example (1010175891) Complies with BF Isolation requirements in accordance with UL 2601-1...
  • Page 6 - CONSOLE MARKINGS - MARKINGS DEFINITION Manufacturer Date of Manufacture YYYY-MM Expiration Date YYYY-MM European Conformity Mark Classified by Underwriter’s Laboratories, Inc. with respect to electric shock, fire and mechanical hazards only in accordance with UL 2601-1. Stepping Prohibited Sitting Prohibited DO NOT expose to stray magnetic fields DO NOT Lubricate DO NOT Immerse > Temperature Limitations < Humidity Limitations Atmospheric Limitations Dispose of as per WEEE Directive 2002/96-EEC Caution: Federal Law (U.S.A.) restricts this device to sale by or on the order of a physician (or properly licensed practitioner)
  • Page 7: Environmental Parameters

    INSTRUMENT MARKINGS [REFERENCE ONLY] NOTE: Markings not applicable to the PAL handpieces. Throttle safety lock SAFE “Safe” or “Lock” is engaged. Throttle safety lock “Run” or “Load” is NOT engaged. - ENVIRONMENTAL PARAMETERS - - OPERATING CONDITIONS - Devices have been tested and proven to operate within the following conditions: 80°F/ 27°C 50°F/ 10°C Temperature...
  • Page 8: Important Information

    - IMPORTANT INFORMATION - This manual has been written to help describe the procedures required to keep the MicroAire 1020 Standard Electric Console system operating properly. Throughout the manual, the following terms are used to identify tips and precautions that will help avoid accidental injury to patients or personnel, or prevent damage to the system.
  • Page 9: Warnings

    -1020 STANDARD ELECTRIC CONSOLE - NOTE: All personnel should become familiar with the power equipment before it is set up for use in any procedure. Personnel inserviced should include, but not limited to, central processing personnel, members of the surgical team, and the bioengineering department.
  • Page 10 CAUTION: Check all surgical accessories before attaching them to the handpiece. Make sure that blades, drills, burs, and/ or cannulae are not dull or bent. Check that accessories lock correctly into the handpiece before operating the power instrument. 7. Make sure the throttle lock on the handpiece is in the or SAFE position.
  • Page 11: Troubleshooting

    B.) Three (3) long “beeps” indicate an improper instrument type has been connected to the console. This signal is triggered if MicroAire 6000 series instruments or the 1641 SmartDriver is attached to the REF 1020. If a valid instrument is connected, remove and reinstall both ends of the cable.
  • Page 12: Routine Cleaning/Decontamination

    4. Handpiece will not start. A) Check that the console is “ON” (the main power switch is in the position, with the switch illuminated) and the front panel LEDs are illuminated. B) Make sure the throttle is in the or RUN position. C) Make sure the maximum speed display indicates a maximum speed, and the light over the cable receptacle is illuminated.
  • Page 13: Iec 60601-1-2 Compliance Summary

    IEC 60601-1-2 Compliance Summary Table 1 - IEC 60601-1-2 Table 201 Guidance and Manufacturer’s Declaration – Electromagnetic Emissions The MicroAire 1020 system is intended for use in the electromagnetic environment specified below. The customer or the user of the MicroAire 1020 system should assure that it is used in such an environment. Emissions test Compliance Electromagnetic Environment –...
  • Page 14 Table 2 - IEC 60601-1-2 Table 202 Guidance and Manufacturer’s Declaration – Electromagnetic Immunity The MicroAire 1020 system is intended for use in the electromagnetic environment specified below. The customer or the user of the MicroAire 1020 system should assure that it is used in such an environment. IEC 60601 Compliance Electromagnetic...
  • Page 15 Table 3 - IEC 60601-1-2 Table 204 Guidance and Manufacturer’s Declaration – Electromagnetic Immunity The MicroAire 1020 system is intended for use in the electromagnetic environment specified below. The customer or the user of the MicroAire 1020 system should assure that it is used in such an environment. IEC 60601 Compliance Electromagnetic...
  • Page 16 Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the MicroAire 1020 System The MicroAire 1020 system is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the MicroAire 1020 system can help prevent...
  • Page 17: Warranty, Service, Repair & Disposal

    We may be able to help solve the problem quickly without returning the item for service. DO NOT disassemble or attempt to service the equipment. It can only be serviced by MicroAire or an Authorized MicroAire Repair Facility. Unauthorized service will void the warranty.
  • Page 18 - 1020 CONSOLE WARRANTY - MicroAire Surgical Instruments LLC warrants its 1020 Standard Electric Console to be free from defects in material and workmanship in their manufacture for a period of 1 year from the original purchase date by the end customer.
  • Page 20 CHARLOTTESVILLE, VA 22911 11, RUE EMILE ZOLA BP 2332 F-38033 GRENOBLE CEDEX 2 FRANCE Phone: 800-722-0822 or 434-975-8000 International Phone: +001-434-975-8000 Order Fax: 800-648-4309 or 434-975-4131 www.microaire.com IM-1020 Rev. C ©2012 MicroAire Surgical Instruments LLC. All Rights Reserved Made in USA...