Directive is based upon EN 10524-1 standard. 2. INTENDED USE Comweld Healthcare Medical Regulators are designed for use with high- pressure medical gas cylinders equipped with a medical cylinder valve. They regulate pressure and flow of medical gases to the patient. They...
Before initial use the product should be kept in its original packaging. Comweld Healthcare recommends use of the original packaging (including internal sealing bag and caps) if the product is withdraw from operation (for transport, storage). Statutory laws, rules and regulations for medical gases, accident prevention and environmental protection must be observed.
EN ISO 60601-1 and 60601-1-2. C, E – Flow-metering device and flow outlet Comweld Healthcare regulators can be supplied with a flow-metering device - flow control head “C” or flowmeter “D”. This function is used to supply a gas flow (l/min) at atmospheric pressure directly to the patient through the flow outlet “E”, e.g.
Page 7
• Check that the total life time of the product and the gas cylinder has not been exceeded, (refer to Comweld Healthcare or owner’s date coding system). If life time has been exceeded, remove the product (or the gas cylinder) from service & suitably identify its status.
Page 8
Turn off the cylinder valve by turning the hand wheel in an clockwise direc- tion to stop position. Do not use excessive force. If any leakage is detected, use the procedure in chapter 6.3 and return the product to Comweld Healthcare for service. 6.1.4. FUNCTIONAL CHECKS BEFORE USE •...
Page 9
Regulator with threaded connector as pressure outlet shall be only an in- tegral part of medical equipment. Do not use it for other purposes! Pressure outlet II • Ensure the counterpart is compatible with the pressure outlet features. • Screw the counterpart. •...
6.2.5. USE OF PRODUCT THROUGH THE PRESSURE OUTLET. • Ensure that the flow control knob is on the ZERO position (if any). • Ensure the accessory IS NOT connected to the pressure outlet. • Slowly open the cylinder valve by turning the hand wheel in anticlockwise direction approx 1 to 1½...
The provider of the device shall prevent the reuse of the product and handle the product in compliance with local regulations. In accordance with Article 33 of REACH Comweld Healthcare, shall inform all customers if materials containing 0.1% or more of substances included in the list of Substance of Very High Concern (SVHC).
Page 12
• Indicator, • Quick coupler. The repairs shall be carried out by a Comweld Healthcare authorised person only. Any product sent back to a Comweld Healthcare authorised person for maintenance shall be properly packaged. The purpose of the maintenance has to be clearly specified (repair, overall maintenance). For product to be repaired a short description of fault and any reference to a claim number might be helpful.
9. GLOSSARY Consult instruction for Suitable for Home care Suitable for Hospital Caution care use Keep away from heat and Suitable for Emergency flammable material care use Keep away from oil and Serial number grease Humidity limit Catalogue number Temperature limit Batch code Keep dry Fragile, handle with care...
10. WARRANTY LIMITED WARRANTY: CIGWELD Pty Ltd, An ESAB Brand, hereafter, “CIG- WELD” warrants to customers of its authorized distributors hereafter “Pur- chaser” that its products will be free of defects in workmanship or material. Should any failure to conform to this warranty appear within the time period applicable to the CIGWELD products as stated below, CIGWELD shall, upon notification thereof and substantiation that the product has been stored, installed, operated, and maintained in accordance with CIGWELD’s speci-...
Page 15
Cigweld Customer Care 1300 654 674 for advice on your nearest Service Provider. CIGWELD reserves the right to request documented evidence of date of purchase. CIGWELD or our Accredited Distributor must be notified in writing of its claim within seven (7) days of becoming aware of the basis thereof, and at its own expense returning the goods which are the subject of the claim to CIGWELD or nominated Accredited Distributor/Accredited Service Provider.
Need help?
Do you have a question about the MEDISELECT II and is the answer not in the manual?
Questions and answers