FIA Meter Plus
Quick Operation Instruction for Use
Standard Test Procedure
1. Connect the instrument with provided power adapter and make sure
stays in "MAC". After swithing on, the instrument will boot up,
the
PC/MAC
and automatically pull out the Test Device holder.
2. Once the operation interface appears, choose the "Standard Test" by
clicking the "Test Mode", and input patient's information. Prepare the
test kit and insert its matched ID chip "
" into the ID Chip Port.
3. Draw the accurate sample volume which is in accordance with the
instructions of test kits to the buffer to mix thoroughly.
4. Take out the Test Device. After adding indicated volume of
sample-buffer mixture into the sample well of Test Device, immediately
insert the Test Device into the Test Device holder of Finecare™ FIA
Meter Plus and click "StartTest".
5. When the Test Device is tested by Finecare™ FIA Meter Plus, the
instrument will identify and match the lot numbers of Test Device and ID
chip. (The instrument pops up a prompt, reading "insert a corresponding
ID chip", if the lot numbers fail to match.) Subsequently the instrument
starts to count down the incubation time and shows the loading progress.
6. The instrument will automatically read and show the result as soon as
the incubation time ends.
7. Click "Print" to print the result.