Do you have a question about the ACCUSTAT P2 and is the answer not in the manual?
Questions and answers
Summary of Contents for Biological Controls ACCUSTAT P2
Page 1
ACCUSTAT™ P2/DL Portable Differential Pressure Monitor User Manual Read and Save these Instructions Manufactured by 749 Hope Road, Suite A- Eatontown, NJ 07724 USA Tel: 800-224-9768, 732-389-8922 Fax: 732-389-8821 www.biologicalcontrols.com 12/14/2009...
Page 3
CONDITIONS THAT MAY CAUSE PERSONAL INJURY OR PROPERTY DAMAGE. WEAR SAFETY GLASSES AND WORK GLOVES AND FOLLOW ALL SAFETY LOCAL BUILDING AND ELECTRICAL CODES. CONSULT A QUALIFIED INSTALLER, SERVICE AGENCY OR YOUR SUPPLIER FOR INFORMATION OR ASSISTANCE Copyright 2009 Biological Controls 12/14/2009...
Page 4
Standard P2 Data Logger P2/DL Copyright 2009 Biological Controls 12/14/2009...
48+ hours of use. When the charge is very low a “BAT” symbol appears in the LCD display. When this occurs the unit must be connected to the power-module to recharge the battery to avoid halting an active data logging session. Copyright 2009 Biological Controls 12/14/2009...
Set the two (2) Range Switches before performing the Meter Zero Adjustments. Meter ZERO Adjustment Note: This calibration also sets the Zero-point for the data log graph data. • Power the ACCUSTAT P2™. • Allow 5-15 minutes warm-up to stabilize the circuits.
LED will flash red (and ultimately set off the audible alarm after 1 minute) when the pressure is more negative than the set-point. 3.6. Audible Mute The Pressure Alert Mode Switch can be moved to the Mute position to turn-off the audible sound without affecting the LED functions. Copyright 2009 Biological Controls 12/14/2009...
Any other battery will void the warranty. The battery-pack assembly may be purchased from Biological Controls. Never use Lithium-Ion type batteries as replacements. The fire- hazards associated with Lithium-Ion technology is not safe for this instrument, the user or the environments where the instrument will be operated or stored.
Accustat™ P2/DL. Simply insert the CD into the computer and follow the automatic icon-driven instructions. The latest software and User Manual can be downloaded from the Biological Controls web site if needed. The Accustat™ power switch must remain” on” to record data unless the log Suspend-Mode is activated before turning the main power switch off.
The optional Data logging function should be ordered at the time of purchase. It is possible to retrofit the logging feature, but the unit must be returned to the factory. (Always obtain the required RMA number from Biological Controls before sending the unit for retrofit). Copyright 2009 Biological Controls...
The sole responsibility of Biological Controls shall be to repair or replace the product within the terms stated above. Biological Controls shall not be liable for any loss or damage of any kind, including any incidental or consequential damages resulting, directly or indirectly, from any breach of warranty, expressed or implied, or any other failure of this product.
Accustat™ to the Suspend-Mode the battery status check reveals that there is only a 60% charge. Not having time to charge the battery the user puts the unit in the Suspend-Mode, turns the power switch off, and leaves for home. The green Log Status LED continues to Blink. Copyright 2009 Biological Controls 12/14/2009...
The user may save the data in .txt format for Note Pad and Word programs or the data may be exported directly to Mircosoft ® Excel by using the green “Export” tab. Control Screen User chooses one of the 3 Icons or checks the Options or Help tab Copyright 2009 Biological Controls 12/14/2009...
Page 16
Set-Up Screen Result of selecting the Green Set-Up Icon on the Control Screen User chooses Units, Range and Sampling Rate and starts logging session Status Screen Selected under the Options tab Copyright 2009 Biological Controls 12/14/2009...
Page 17
Stopped Condition Screen Result of selecting the Red Set-Up Icon on the Control Screen Graph-View Screen Result of selecting the 3 Graph-Icon on the Control Screen Copyright 2009 Biological Controls 12/14/2009...
1-Wall Power Module eq/w international adaptors Accuracy: 1% of set point 1-Accessory storage bag User manual Adjust ± the meter range 1- Calibration adjustment tool Alert LED (select Green or blinking Red) 1- Software CD &1-6’ USA/B cable (P2/DL model) Copyright 2009 Biological Controls 12/14/2009...
Need help?
Do you have a question about the ACCUSTAT P2 and is the answer not in the manual?
Questions and answers