WH501P High Purity Deionization System
WH501C High Capacity Deionization System
Thank you for purchasing WH Series Auto Pure Water Purifier of
Yamato Scientific.
To use this unit properly, read this "Instruction Manual" thoroughly
before using this unit. Keep this instruction manual around this unit
for referencing at any time.
WARNING: Carefully read and thoroughly understand the important
warning items described in this manual before using this unit.
AUTO PURE WATER PURIFIER
WH501 SERIES
MODEL
INSTRUCTION MANUAL
- FIRST EDITION –
Yamato Scientific America Inc.
Santa Clara, CA

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the WH501 Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Yamaha WH501 Series

  • Page 1 AUTO PURE WATER PURIFIER WH501 SERIES MODEL WH501P High Purity Deionization System WH501C High Capacity Deionization System INSTRUCTION MANUAL - FIRST EDITION – • Thank you for purchasing WH Series Auto Pure Water Purifier of Yamato Scientific. • To use this unit properly, read this "Instruction Manual" thoroughly before using this unit.
  • Page 2: Technical Specifications

    TECHNICAL SPECIFICATIONS WH501P WH501C Pressure range 10-100 psi 10-100 psi Temperature range 40 - 100 °F 40 - 100 °F Maximum flowrate 1.25 gallons per minute 1.25 gallons per minute Resistivity output 200 KOhm to 10 MegOhm 50 KOhm to 200 KOhm Resistivity DI light set point 200 KOhm 20 KOhm...
  • Page 3: Operational Procedure

    OPERATIONAL PROCEDURE INSTALLATION Install system on a wall that will support 55 lbs. within 5 feet of a water supply. Connection to water supply should include a shut off valve before the WH501 system. Use your choice of inlet fitting provided with the system to connect the 3/8” black tubing to the water supply valve.
  • Page 4: Cartridge Replacement

    OPERATIONAL PROCEDURE CARTRIDGE REPLACEMENT Turn off the inlet water feed supply. Depressurize the system by pushing the red pressure relief button on the top of the filter housing Place a container under the filter housing to collect and excess water. Unscrew the filter bowl housing.
  • Page 5: After Service And Warranty

    AFTER SERVICE AND WARRANTY WARRANTY POLICY Yamato Scientific America warrants, from the date of shipment from warehouse in Camden, New Jersey, U.S.A., for a period of one (1) year. All products, parts and materials shall be free of defects in material and workmanship under normal use consistent with the product instructions.
  • Page 6 AFTER SERVICE AND WARRANTY WARRANTY POLICY CONTINUED The purchaser is responsible for the shipping cost of return shipment. Insurance on the return shipment is required. Damage or loss of merchandise during shipping is the responsibility of the sender. Returned shipments that arrive damaged will be returned to the sender, and credit will not be rendered.