Page 1
EN – English Installation and operating instructions DL 200 Hand-held instrument DL 200 Seite 1 von 93...
Page 2
The device is solely intended for the described purpose. Riegler & Co. KG shall not assume any warranty with respect to the suitability for any particular purpose, and shall not assume any kind of liability for errors which are printed in this manual. Nor for consequential damages in connection with the delivery, performance or use of this device.
CABLE CROSS SECTION Sensor circuit points/Output signal: CONNECTION DIAGRAMS OF THE DIFFERENT SENSOR TYPES Connector pin assignment for all sensors at DL 200 Connection dew point sensors series FA 415/FA 300 Connection for dew point- and consumption sensors, series FA/VA 400...
Page 4
Check for new Software updates (USB) 7.3.2.1.4.5 Factory Reset 7.3.2.1.4.6 Calibrate touch-screen 7.3.2.1.5 Set backlight 7.3.2.1.6 Cleaning 7.3.2.1.7 System-Status 7.3.2.1.8 About DL 200 7.3.2.2 Chart 7.3.2.3 Chart / Real time values 7.3.2.4 Channels 7.3.2.4.1 Min/Max Function 7.3.2.5 Real time values 7.3.2.6...
Page 5
9.2.5 Definition of Operations 9.2.6 Definition of Unit 9.2.7 Value name, resolution of decimal places and recording of values ANALOG TOTAL (OPTIONAL) 10.1 Option „Analog Total“ activation 10.2 Selection of sensor type CLEANING BATTERY DL 200 Seite 5 von 93...
This instruction manual has to be available at any time at the operation site of the DL 200. Regional and national regulations respectively, have to be observed in addition to this instruction manual if necessary.
The plug of the power supply unit (charger) is used as a separator. This separator must be clearly recognisable and easily accessible by the user. A plug connector with a CEE7/7 system is necessary. NOTE! Only the supplied power supply may be used. DL 200 Seite 7 von 93...
Application Area Application Area The new DL 200 is an all-purpose hand-held measuring instrument for many applications in industry Like e.g.: ► Consumption/flow measurement ► Pressure/vacuum measurement ► Temperature measurement ► Moisture/dew point measurement The 3.5” graphic display with touch screen makes the operation very easy.
Inputsignals for Sensor Input signals ext. sensor DL 200 Input signals 0 – 20 mA / 4 – 20 mA Measuring range Current signal Resolution 0,0001 mA (0 – 20 mA / 4 – 20 mA) internal or external 0,03 mA 0,05 %...
Connection diagrams of the different sensor types Connection diagrams of the different sensor types Connector pin assignment for all sensors at DL 200 The interface connector to be used is a ODU Medi Snap 8 pin – Reference: K11M07-P08LFD0-6550 Available connection cables at CS-Instruments are: ODU with Open ends: Order no 0553 0501, cable length: 5 m.
FA 300 Sensor FA 415 Analog IN - I (500µA) The digital data transmission between DL 200 and the dew point +VB 24Vdc sensors FA 415 and FA 300 occur via the SDI bus line. blue -VB GND Connection for dew point- and consumption sensors, series FA/VA 400...
- RS485 R = 4K7 Analog IN + yellow Analog IN - grey Attention: I (500µA) The DL 200 is counting a +VB 24Vdc blue consumption unit, by switching -VB GND „power on“. + RS485 - RS485 Analog IN +...
Analogue two-, three-, and four-wire current signal Sensors with 4 - 20 mA-output in 2-wire technology + RS485 - RS485 ß 0 / 4 ...20mA Analog IN + Yellow DL 200 Analog IN - Sensor I (500µA) +VB 24Vdc blue -VB GND...
+ RS485 - RS485 ß 0...1 / 10 / 30 Vdc Sensor with voltage output in 4-wire technology Analog IN + Yellow Sensor Analog IN - grey I (500µA) +VB 24Vdc blue -VB GND DL 200 Seite 15 von 93...
I (500µA) Pink +VB 24Vdc -VB GND Connection with RS485 white + RS485 brown - RS485 Analog IN + Sensor Analog IN - Sensor with RS485 Interface I (500µA) +VB 24Vdc blue -VB GND DL 200 Seite 16 von 93...
Operation DL 200 Operation DL 200 The operation of the DP 500 7 DP 510 by means of a keypad and a touch panel Keypad 7.1.1 On- and Off-button On-or Off switching by long press buttons. 7.1.2 Brightness buttons With the button the display brightness can be changed.
Main menu Export Data Export Screenshots Change The selected date is always green, and the date numbers of the Sundays are red, like in the calendar. On days, where bitmaps were recorded, the date numbers are optical highlighted. DL 200 Seite 18 von 93...
Page 19
Operation DL 200l Main menu Export Data Export Screenshots Export The Screenshots of the selected period are exported to the USB-Stick. DL 200 Seite 19 von 93...
The menu navigation is generally in a green font! The table of contents and chapter references in blue font contain links to the respective chapter title. DL 200 Seite 20 von 93...
Main menu Main menu (Home) From the main menu, you can reach every available item 7.3.1 Initialization After switching on the DL 200 the channel is initialized and the menu „ „ Real time values appears. Attention: For the first initiation, there may be no...
Before the first sensor setting is made, the language and time should be set! Remark: Chapter 7.3.2.1.4.1 language Main Settings Device Settings Language) Chapter 7.3.2.1.4.2 ate & Time Main Settings Device Settings Date & Time) DL 200 Seite 22 von 93...
New password repeat in red font. If you can’t remember the password, please use Master password in order to enter a new password. Remark: The master password is supplied together with the instrument’s documentation. DL 200 Seite 23 von 93...
Main menu Settings Sensor settings The overview of the available channel appears after entering the password. 0.000 m³/h 648195 m³ Remark: 0.000 m/s Usually there is no preset for the external channel! DL 200 Seite 24 von 93...
(see next step). Main menu Settings Sensor settings C1 Type description field Digital Now the Type Digital is selected for the VA/FA 400 series and confirmed by pressing button. DL 200 Seite 25 von 93...
2345678 ltr 0.00 m/s Attention: Before the selected measurement data are recorded, the data logger must be activated after the settings (See chapter 7.3.2.1.3.2 Logger-Settings(Datalogger)) DL 200 Seite 26 von 93...
Main menu Settings Sensor settings C1 0.000 ltr/ min After alarm setting for Channel C1a. 2345678 ltr 0.00 m/s Remark: After confirm with OK, the font is black again and the values and settings are accepted DL 200 Seite 27 von 93...
More-Settings only for type CS-Digital 110.00 0.000 available! 200.000 The settings are completed after pressing the button! Remark: After confirming with OK, the font is black again and the values and settings are accepted. DL 200 Seite 28 von 93...
VA/FA 400 series and confirmed by pressing button. Main menu Settings Sensor settings C1 Dew Point The DL 200 detects, if the connected sensor is a flow or dew point sensor and set the Digital -9.2 °Ctd subtype automatically correct.
VA/FA 400 series and confirmed by pressing button. Main menu Settings Sensor settings C1 The DL 200 detects, if the connected sensor is a flow or dew point sensor of CS Instruments 0.000 ltr/ min and set the...
Page 31
By entering the white text fields the value could be added or changed content could be change Main menu Settings Sensor settings C1 description field Unit A preset selection of suitable Units. DL 200 Seite 31 von 93...
Page 32
Main menu Settings Sensor settings C1 arrow right (2.page Gas Constant description field A preset selection of suitable Constants. Remark: After confirming with OK, the font is black again and the values and settings are accepted. DL 200 Seite 32 von 93...
Input of a name, please enter the text field „Name“. It is possible to enter a name with max. 24 characters. Confirmation by pressing the OK-button. The connection with the FA 5xx sensor is “OK”. done after confirmation by pressing DL 200 Seite 33 von 93...
The value is entered in the corresponding text field. The unit can be freely selected, selection menu is opened by pressing the corresponding button units Confirm the settings by pressing the button. DL 200 Seite 34 von 93...
Page 35
Main menu Settings Sensor settings C1 arrow right (2.page)Pressure Setting Text field Ref.Pressure Reference pressure is the pressure for that the dew point in relaxation will be back-calculated. Default- Value is 1013 mbar (Atm. Pressure). Confirm the settings by pressing the button. DL 200 Seite 35 von 93...
Page 36
Sensor error / System error Sensor error / System error 3.8mA – 20.5 mA) 4..20 Output according Namur ( < 4mA to 3.8 mA Measuring range under range >20mA to 20.5 mA Measuring range exceeding DL 200 Seite 36 von 93...
Input of a name, please enter the text field „Name“. It is possible to enter a name with max. 24 characters. Confirmation by pressing the OK-button. The connection with the VA 5xx sensor is “OK”. done after confirmation by pressing DL 200 Seite 37 von 93...
(Please, inquire at the manufacturer or measure by your own!) Main menu Settings Sensor settings C1 arrow right (2.page) diameter unit description field After pressing the Unit Text fields following units bare selectable. DL 200 Seite 38 von 93...
20 °C, 1000 hPa (according to ISO 1217 intake condition) 0 °C and 1013 hPa (= standard cubic meter) can also be entered as a reference. Do not enter the operation pressure or the operation temperature under reference conditions! DL 200 Seite 39 von 93...
7.3.2.1.2.9.5 Definition Unit of flow and velocity Main menu Settings Sensor settings C1 arrow right (2.page) Flow description Field Main menu Settings Sensor settings C1 arrow right (2.page) Velocity description Field DL 200 Seite 40 von 93...
Selection to confirm selection by pressing button. Important! When the counter reach 100000000 m³ the counter will be reset to zero. Remark: After confirmation with OK, the font is black again and the values and settings are accepted DL 200 Seite 41 von 93...
Output according Namur ( < 4mA to 3.8 mA Measuring range under range >20mA to 20.5 mA Measuring range exceeding “OK” Inputs / changes to be confirmed with button. Return “Back”. to main menu with DL 200 Seite 42 von 93...
High: Value over limit Low: Value under limit “OK” Inputs / changes to be confirmed with “Back”. button. Return to main menu with DL 200 Seite 43 von 93...
“CutOff” and insert the required value, here 10. With the Reset” button all entries could be set back to zero. “OK” Inputs / changes to be confirmed with “Back”. button. Return to main menu with DL 200 Seite 44 von 93...
It is possible to define a Offset-Value. With Set Value to-button (Offset) you enter it. The positive or negative difference of the Offset will be displayed. By pressing the Reset-button the Offset will be deleted DL 200 Seite 45 von 93...
Page 46
User unit be defined. Main menu Settings Sensor settings C1 Type description field 0/4 - 20 mA Here for example Type 4 - 20 10.55 bar DL 200 Seite 46 von 93...
°C are chosen, alternatively the sensor types PT1000 and KTY81, as well as the Unit °F can be selected. More setting options, see chapter 7.3.2.1.2.9 Type 0 - 1/10/30 Volt and 0/4 - 20 DL 200 Seite 47 von 93...
50 Hz m³/h Main menu Settings Sensor settings B1 arrow right (2.page) Unit Pulses Unit Pulse you can choose between a flow volume or a power consumption unit. DL 200 Seite 48 von 93...
Page 49
The available Units for the Unit Counter Type Pulse counter can be set any time to any value you need. More setting options, see chapter 7.3.2.1.2.9 Type 0 - 1/10/30 Volt and 0/4 - 20 DL 200 Seite 49 von 93...
Main menu Settings Sensor settings C1 Type description field No Sensor Is used to declare a not currently needed channel as No Sensor defined. If you go to Type No Sensor Back, the channel will appear as unused. DL 200 Seite 50 von 93...
In case that the DP 510 is the end of the RS485 bus system with activating Term- & Bias- button the required termination and biasing could be activated. Confirmation by pressing button. For resetting to the default values please press Set to Default. DL 200 Seite 51 von 93...
Page 52
Modbus and read by the DL 200 This requires setting the desired register addresses in the DL 200 Entering the register / data address is here in decimal with 0-65535. Important: Required is the correct register-address.
Page 53
To move through the list please press the button Page. In case the unit is not available it is possible to create a user defined unit. Therefore, please select one of the User_X buttons. DL 200 Seite 53 von 93...
Page 54
By default or value = 0 no scaling is applied don’t scale and displayed in the field is Main menu Settings Sensor settings C1 OK By pressing the button, the inputs are confirmed and stored. DL 200 Seite 54 von 93...
2 seconds. In addition, if more than 25 measurement data are recorded at the same time, the smallest possible time interval of the data logger is 5 seconds. DL 200 Seite 55 von 93...
Page 56
Remark: If the start time is activated, it will automatically be set at the current time plus a minute. DL 200 Seite 56 von 93...
Page 57
Main menu Settings Logger settings timed Start button/timed Stop button Date/Time description field After pushing the date/time description field a window will appear where the yellow marked area of the time or date can always be set and changed. DL 200 Seite 57 von 93...
Page 58
The settings cannot be changed, if the data logger runs. Important: If a new recording file should be created, the force new record file button must be activated. Otherwise, the last applied recording file is used. DL 200 Seite 58 von 93...
Main menu Settings Device settings Overview of Device settings 7.3.2.1.4.1 Language Main menu Settings Device settings Set language Here you can select one of 10 languages for the DL 200 DL 200 Seite 59 von 93...
By pushing the Time Zone description field and enter the correct UTC, you can set the correct time all over the world. The summer and wintertime switchover is realized by pushing the Daylight Saving button. DL 200 Seite 60 von 93...
SD-Card will be blocked for use in DS 400. Nevertheless all data are still stored and available for external use only. By pressing Erase SdCard all Data on the SD-Card will be deleted. DL 200 Seite 61 von 93...
If after pushing the Check USB Stick for new Software updates button the following messages in the window appears, then DL 200 is not connected properly with the USB stick or no files are available. DL 200 Seite 62 von 93...
Page 63
Geräte-Einstellung / Datum & Uhrzeit Main menu Settings Device settings Update System Update selections If the DL 200 is correctly connected to USB, and new version available it will displayed. Right aside it shows the current (old) and...
System / Factory Reset 7.3.2.1.4.5 Factory Reset Main menu Settings Device settings Factory Reset Reset to Defaults If necessary or required, by pressing the Reboot System-button the DL 200 could be rebooted. DL 200 Seite 64 von 93...
Backlight is committed automatically on the last set value before dimming. Remark: At the first touch, the Backlight in our example is reset to 50%, after that a "normal" function operation is possible. Important: DL 200 Seite 65 von 93...
Page 66
Analog Total If the Backlight dimming after button is not activated, then the Backlight stays permanently on, in the currently set brightness. DL 200 Seite 66 von 93...
7.83 V supply unit. 5.05 V By the Runtime, you always know how long the DL 200 was in total in operation 5.01 7.3.2.1.8 About DL 200 Main menu Settings About DP 510 Brief description of the Hardware...
The smallest possible range is represented, depending on the time interval of the recording. Additional zooming and scrolling options in Chart Chart/Real time values bigger period of time Time later time earlier Smaller period of time DL 200 Seite 68 von 93...
Page 69
(min, max, step) and several channels (Plots) and a Colour. The y-axis left is already enabled, you can choose a Colour for it. Remark: Grid setting is already possible at this point, but later when a record is selected it is more reasonable! DL 200 Seite 69 von 93...
Page 70
Now, the grid can be set with min, max, and m³/h step. By pushing the A.Scale-button a calculated auto-scaling will be defined. In the same way the remaining y-axes can be labelled! m³/h Two different grid settings with various Units and Colours. DL 200 Seite 70 von 93...
Page 71
Chart Main menu Chart DL 200 Seite 71 von 93...
For each channel, you can select a value to be represented in the Chart and one to display values). In addition, it can be set, like in Main Chart, a colour and the grid (min, max, step) of the y-axis. DL 200 Seite 72 von 93...
Page 73
If there is no grid entered in the setup, will be 0, 100 and step In the same way the remaining setups can be set! DL 200 Seite 73 von 93...
Start of recording is immediately after setting of the sensor, but there is always the possibility to reset the Min and Max values. Main menu Channels I1 ↑ ↓ = Max-Wert = Min-Wert DL 200 Seite 74 von 93...
Page 75
For resetting the single value the Reset –Button for all Min/Max-Values the Value –Button has to be pressed. Reset All Values DL 200 Seite 75 von 93...
You can choose between 6 different layouts showing 1-5 measurements. see below. The values to be displayed could be selected in the Val.1 to Val.5 description fields. Different variants: DL 200 Seite 76 von 93...
Export Logger data Change The selected date is always green, and the date numbers of the Sundays are red, like in the calendar. On days, where measurement data were recorded, the date numbers are optical highlighted. DL 200 Seite 78 von 93...
Page 79
The measurement data of the selected period are exported to a USB stick. Main menu Export data Export system settings By using Export system settings, all existing sensor settings can be exported to a USB stick. DL 200 Seite 79 von 93...
New File storage: Select the location for storing by pressing the button SDCard. choosing button file menu inserting/defining the filename appears. The file name length is limited to 8 chars. OK OK File save/confirm with: DL 200 Seite 80 von 93...
After importing of the new settings a reboot is required too. For the complete takeover of the new sensor settings, they have to be activated for channel C1. Main menu Settings Sensor Settings Channel DL 200 Seite 81 von 93...
Option „Virtual Channels“ activation After purchasing of the option „Virtual Channels“ the functionality have to be activated first. Main menu Settings About DL 200 Please push the button Buy for „Virtual Channels“ and you will requested to insert the...
If still no sensor has been configured, the Type No Sensor appears. By pushing the button Generic the virtual channel is selected. Pushing the button No Sensor will reset the virtual channel. Confirmation of selection is done by pressing the button DL 200 Seite 83 von 93...
Main menu Settings Sensor Settings Virtual Channels V1 arrow right (2.page) 1stOperand By accessing the text field 1st Operand The list with all channels (HW and virtual channels) and const. Value appears. DL 200 Seite 84 von 93...
Page 85
With the buttons it is possible to correct the input. ß Button deletes the last figure Button clears the whole field This approach is analogous to the other operands. (1st Operand, 2nd Operand and 3rd Operand). DL 200 Seite 85 von 93...
To move through the list please press the button Page. In case the unit is not available, it is possible to create a user defined unit. Therefore please select one of the User_X buttons. DL 200 Seite 86 von 93...
Page 87
Each calculation allows you the use of maximum 3 operands and 2 operations. The calculation is then based on following formula: Example: V1a = (1st Operand 1st operation 2nd Operand) 2nd operation 3rd Operand V1a = (A1c – A2a) * 4.6 DL 200 Seite 87 von 93...
Attention: Before the selected measurement data are recorded, the data logger must be activated after the settings (See chapter 7.3.2.1.3 Logger-Settings (Data logger)). See also chapter 7.3.2.1.2.2 Name the measurement 7.3.2.1.2.3 Recording measurement data DL 200 Seite 88 von 93...
0-1/10/30V and 0/4 – 20mA. 10.1 Option „Analog Total“ activation After purchasing of the option „Analog Total“ the functionality has to be activated first. Main menu Settings about DL 200 for „Analog Total“ Please push the button...
The text field „Unit-Consumption“ is only editable in case of measurement values (Units) with volume per time unit and thus also the consumption calculation. For labelling and setting of the description fields, see also chapter 7.3.2.1.2.7 label and setting the description field DL 200 Seite 90 von 93...
Chapterl 7.3.2.1.6. Cleaning of the DL 200 must be undertaken using a slightly damp (not wet) cotton cloth or one-way wipe, and mild, commercially available cleaner/soap. For decontamination, spray the cleaner on an unused cotton cloth or one-way wipe, and wipe the component comprehensively.