Part list Multi backrest ..................................19 Exploded viwe backrest SitRite ................................. 20 Part list backrest SitRite ..................................21 Exploded view chassis Miniflex ................................ 22 Part list Chassis Miniflex ................................... 23 Part list chassis Miniflex ................................... 24 Control box ....................................... 24 Reconditioning ....................................
Dismantling the seat from the chassis • Turn off the wheelchair • Detach the bus cable A and if the chair is equipped with electric tilt / electric back angle, these cables are discon- nected from the actuator and anchors in the seat cross. •...
Replacing the batteries • Remove the cover, see above. • Loosen the straps that hold the batteries. • Turn off the main switch, on the circuit breaker. • Detach the plus and negative terminals from the batteries. • Remove the batteries. •...
Replacement of actuator module • Disconnect the Bus connector and the con- nectors for the seat actuators. • Loosen the screws holding the module, lift out the module. • Assemble the new module in reverse order. Replacing the drive belt •...
Castor Wheel Replacement Place a wooden block or similar under the control arm to relieve pressure on the wheel. 2. Loosen the wheel bolt with Allen key 5mm and fixed wrench 13mm. 3. Remove the screw and lock nut, loosen the wheel and replace or clean it.
Replacement drive motor • Dismantling the drive belt (see above) • Unplug your motor contact (M1, M2) from the electronics. • Cut off the cable ties on the battery box. • Loosen screw A on the right cable lock nipple and pull out the wire. •...
Mounting tilt actuators • Install plastic washer 1 on each axle pin. • Install tilt devices. • Install plastic washer 1 on each axle pin. • Install screw 3 M5 pre-treated with thread locker and washer 2. Same installation procedure applies to all models of tilt device.
Debugging Problem Measure The wheelchair won't start Circuit breaker tripped. Check the circuit breaker Batteries depleted. Charge the batteries, if not possible replace the batteries Gaps in cables/connectors. Locate gaps and fix, troubleshoot using Flashing codes or diagnostics via computer Batteries do not accept charging Circuit breaker tripped.
Part list Multi-backrest Position Article Name SW-01985 Back post lower back SW-01992 Back post upper part SW-02034 Back post Multi Back SW-01999 Pin bolt M8 274001–6005 Profile fastening 674001–6004 Profile back post ABC 675002-4400 Aluminum profile lock lug ABC SW-01993 Adjuster, back post SW-01997 Pole Adjuster back post...
Part list backrest SitRite Position Article nummer Designation 17a. 624525–3640 Backrest cushion SitRite 36*40cm 624525–4245 Backrest cushion SitRite 42*45cm 624525–5050 Backrest cushion SitRite 50*50cm 17b. 674001–6001 Back profile ABC high 280mm 17c. 7F6S-6*14 Screw MF6S-6*14 17d. SW-01502 Cover aluminum profile ABC SitRite SW-02034 Back post Multi Back 20f.
Test of the brakes and the brake release mechanism. Lubricated wires. Lifting columns / actuators Securely attaches and does not rotate. Greased with thin oil (Miniflex 500) Noise level low / no rattling noise (Miniflex 500) Lifts all the way up with the user in it.
Reconditioning Final inspection All optional equipment: fits in and is easy to adjust. Warning labels attached. CE marked with serial number. Test run (all functions are tested in the end positions and the wheelchair is tested with maximum load) The instructions for use are available and supplied with the product. Destruction Products that are permanently taken out of use must be dismantled and sorted at source.
Need help?
Do you have a question about the Miniflex and is the answer not in the manual?
Questions and answers