Instructions for use
ProLift Cacero
ProLift: Innovative Patient Hoist Solutions

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ProLift Cacero and is the answer not in the manual?

Questions and answers

Summary of Contents for NOVACARE ProLift Cacero

  • Page 1 Instructions for use ProLift Cacero ProLift: Innovative Patient Hoist Solutions...
  • Page 2: Table Of Contents

    In these instructions for use, people who need assistance are generally referred to as „patients“. The novacare ® ProLift Cacero is a patient lifter for everyday lifting and transfer functions in clinical, nursing and outpatient (homecare) areas. As a medical aid, the product is mainly given to medical experts or professional users.
  • Page 3: Product Information

    ® lifter that can easily handle most lifting situation. Ergonomics and design are equally considered in the ProLift Cacero. Many small details, such as the high-quality electronic components, complete the overall picture of the ProLift Cacero. The design of the base offers the patient more space for a more relaxed lifting situation and, thanks to its wide spread, enables access to large chairs or wheelchairs.
  • Page 4 Tension bolt with ball bearings Mounting bolt for tension bolt Rubber protection cap for base spreader Rubber protection cap for base rear Nut and bolt base Front double castors Rubber protection cap base front Power cord, Euro plug Base ProLift Cacero...
  • Page 5 Do not use the lift if the patient has a seizure and could fall. Technical Data Technical Specification Type patient hoist Model ProLift Cacero Hoist frame powdercoated steel, corrosion resistant, plastic, electronic components Turning circle base 1260 mm Castors front PU, 75 mm, rotatable 360°, double castors...
  • Page 6: Installation & Use

    795 to 1060 hPa 795 to 1060 hPa Always store the ProLift Cacero indoors. Take care that water cannot penetrate into electronic com- ponents (actuator, control box, handset). If the patient hoist is not used for a longer period of time, or during transport, the emergency STOP button should always be pressed.
  • Page 7 Always engage the parking brakes on the rear castors for all assembly work. From the rear of the hoist, take hold of the mast and pull upwards. Insert mounting bolt and tighten with hand grip. Screw the electronic box onto the mast. www.novacare.com...
  • Page 8 Attach actuator using mounting clips. Only install the actuator motor as shown; in other positions there is a risk of pinching during operation. Connect the actuator cable (socket M1) and the handset (socket HANDSET) to the electronic box. ProLift Cacero...
  • Page 9 Charger complète- ment la batterie avant premier usage ! Do not open the housing of the electronic box. The battery may only be changed by the ma- nufacturer or a person certified by the manufacturer. Risk of injury. www.novacare.com...
  • Page 10 The maximum duty cycle specified on the produkt label must be observed. Unless otherwise spe- cified, a maximum operating time of 2 minutes continuous operation followed by an 20-minute break applies. Exceeding the duty cycle can reduce the service life and damage the product. ProLift Cacero...
  • Page 11 Now hang the sling loops on the hooks of the sling bar in the following order: 1. Head loops 2. Leg loops (F) For the greatest possible safety, hang the leg loops diagonally so that one loop crosses the other. (G) Please note the colored markings. www.novacare.com...
  • Page 12 If you continue using the hoist with a low battery level, the diode flashes and an acoustic signal sounds continuously. At this point it is only possible to lower the lifter arm (no more lifting possible). Overload: if the maximum load is exceeded and the actuator is overloaded, the actuator stops and ProLift Cacero...
  • Page 13: Faults & Troubleshooting

    If these measures are unsuccessful, notify a service technician or your specialist dealer. Do not repair the lifter yourself. Risk of injury! www.novacare.com...
  • Page 14: Maintenance & Repair

    The last test date can be found on the seal on the device. All service, repair and testing measures may only be carried out by competent and trained personnel. Only original spare and consumable parts may be used. ProLift Cacero...
  • Page 15: Disposal & Environment

    Electrical devices from novacare gmbh are registered (see symbol): WEEE-Reg.-Nr.: DE-89403200 ® Lead acid batteries are used in all models of the ProLift series. The batteries in thes novacare gmbh product are registered (see symbol): Battery Return System ® Contract Number: 109101377 Disposal packaging No special types of packaging are used.
  • Page 16 Type BF applied part - degree of protection against electric shock IPXX Ingress Protection for electrical equipement maximum washing temperature 95 ° C do not bleach Clothes dryer can be used. Maximum outlet temperature 60 ° C. do not iron professional chemical cleaning (normal process) ProLift Cacero...
  • Page 17: Warranty

    • for damage and malfunctions caused by natural wear and tear or during transport. • for the replacement of wear parts and slings. • for products that have been modified, treated or changed without the consent of novacare® gmbh. • in the event of maintenance not been carried out, on time, improperly or by unauthorized per- sons, or damage resulting therefrom.
  • Page 18: Notes

    Notes ProLift Cacero...