Instructions for use
Power socket strip PSS300
WARNING
To properly use this product, read and
comply with these instructions for use.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PSS300 and is the answer not in the manual?

Questions and answers

Summary of Contents for Dräger PSS300

  • Page 1 Instructions for use Power socket strip PSS300 WARNING To properly use this product, read and comply with these instructions for use.
  • Page 2: Typographical Conventions

    Typographical conventions Consecutive numbers indicate steps of action, (A) Letters in parentheses refer to elements in the with the numbering restarting with "1" for each related illustration. new sequence of actions. A Letters in illustrations denote elements referred  Bullet points indicate individual actions or differ- to in the text.
  • Page 3: Target Groups

    Target groups Duties of the operating organization Description of target groups The tasks described in this document specify the The target groups may only perform the following requirements that have to be met by each respec- tasks if they meet the corresponding requirements. tive target group.
  • Page 4: For Your Safety And That Of Your Patients

    For your safety and that of your patients General safety information WARNING Risk of explosion or fire The following WARNING and CAUTION state- ments apply to general operation of the product. The power socket strip is not approved for use in areas where oxygen concentrations above WARNING and CAUTION statements specific to 25 Vol% or combustible or explosive gas mix-...
  • Page 5: Intended Use

    WARNING WARNING Risk of electric shock Risk due to modifications Connecting devices to the power socket strip Modifications to the product may lead to mal- may result in increased leakage current. If the functions and unforeseen risks. This may re- protective ground for one of these devices sult in injury to the patient or the user or in fails, the leakage current may exceed the per-...
  • Page 6 Overview The following images show different types of power socket strips (special-purpose relocatable power taps): Variant 1 Variant 2 Without clamps (only for anesthesia With clamps machine) A Power cable G Potential equalization connectors, 4 pieces B Strain relief for power cable of the power socket H Power sockets for auxiliary equipment, max.
  • Page 7 Symbols Connection for potential equalization Caution! Observe accompanying docu- ments Attention! (safety sign) Warning! Strictly follow these instructions for use MR unsafe Protective Earth (Protective ground) Manufacturer Date of manufacture XXXX Serial number Ordering number Storage temperature Rel. humidity Atmospheric pressure WEEE marking nominal weight: 3 kg Instructions for use Power socket strip...
  • Page 8: Assembly And Preparation

    Assembly and preparation Positioning the power socket strip Connecting the power socket strip WARNING Important information – The effects of the opening of a supplementary Risk of electrical hazard overcurrent protector, which may unintentional- The power socket strip is intended for perma- ly disconnect lifesaving medical equipment nent attachment to medical equipment and, from the mains power supply, needs to be con-...
  • Page 9  Establish a permanent connection between 3 Plug the protective ground connector into the the power socket strip and an isolation potential equalization socket in the wall. transformer. 4 Push the lower part towards the wall until it reaches the end position. WARNING 5 Fix the lower part with the screw.
  • Page 10 Immediate disconnection of all devices 1 Identify the power socket to which the power socket strip is connected. 2 Pull the power plug of the power socket strip.  The power plug of the power socket strip must not be fixed permanently to a power socket on the wall, e.g.
  • Page 11: Troubleshooting

    Troubleshooting Circuit breaker is tripped or the fuses are 4 If the circuit breaker has been tripped: – If necessary, wait at least 2 minutes for the blown circuit breaker to cool. – Press the push button of the circuit breaker 1 Disconnect the power socket strip from the back in.
  • Page 12: Safety Information

    Service Safety information WARNING WARNING Risk of patient injury Risk if service is not performed properly If maintenance activities are carried out Personal injury and property damage may oc- during operation, the patient will be put at risk. cur if service is not performed properly. Only carry out maintenance activities when Service must be performed by those target there is no patient connected to the device.
  • Page 13: Safety Checks

    Definition of service terminology Concept Definition Service All measures (inspection, maintenance, repair) intended to maintain or restore the functional integrity of a product Inspection Measures intended to determine and assess the current state of a product Maintenance Regular specified measures intended to maintain the functional integrity of a prod- Repair Measures intended to restore the functional integrity of a product after a failure Inspection...
  • Page 14: Disposal Of The Product

    Disposal of the product At the end of its useful life, dispose of the product in accordance with the applicable legal provisions. Technical data Environmental Conditions Operation conditions Temperature 0 to 40 °C (32 to 104 °F) Ambient pressure 570 to 1100 hPa (8.27 to 15.95 psi) Rel.
  • Page 15: Device Combinations

    Dimensions 390 x 90 x 60 mm (15.35 x 3.54 x 2.36 in) (W x H x D) Weight, approx. 3 kg (approx. 6.61 lb) without power cable Electrical safety Tested according to IEC 60601-1 or UL 60601-1, Class I Classification Class I in accordance with Directive 93/42/EEC Annex IX and...
  • Page 16 Manufacturer: Drägerwerk AG & Co. KGaA Moislinger Allee 53 – 55 D-23542 Lübeck Germany +49 451 8 82-0 FAX +49 451 8 82-20 80 http://www.draeger.com 9056277–GA en Drägerwerk AG & Co. KGaA Edition: 3 – 2019-07 (Edition: 1 – 2018-06) Dräger reserves the right to make modifications to the medical device without prior notice.