  • Page 1 FARRAR ULTRA LOW TEMPERATURE FREEZER PRODUCT MANUAL FUF126...
  • Page 2 FARRAR products is expressly prohibited. Copyright and Trademark Copyright © 2024 TRANE TECHNOLOGIES LIFE SCIENCES LLC is a registered trademark of Trane in the United States of America. All other trademarks and registered trademarks are the property of their respective owners.
  • Page 3: Table Of Contents

    FARRAR® ULT Freezer Product Manual Contents 1. About this Manual Intended Audience Model Reference Safety Symbols and Precautions Avoiding Injury General Recommendations 2. Installation Location Placement and Leveling Rear Stand-Offs AC Power Cord Storage Shelves 3. FUF Model Operation Initial Start Up Operation Change Temperature Setpoint Set Alarm Parameters...
  • Page 4: About This Manual

    FARRAR® ULT Freezer Product Manual 1. ABOUT THIS MANUAL 1.1 Intended Audience This manual provides information on how to use FARRAR® ULT Freezers. It is intended for use by end users of the freezer and authorized service technicians. 1.2 Model Reference Models are indicated by a distinguishing model number that corresponds to the series, type, and capacity of the freezer.
  • Page 5: Avoiding Injury

    FARRAR® ULT Freezer Product Manual 1.4 Avoiding Injury Do not use mechanical devices or other means to accelerate the defrosting process, other than those recommended by the manufacturer. Do not damage the refrigerant circuit. Review safety instructions before installing, using, or maintaining the equipment. Before moving unit, remove contents from the chamber.
  • Page 6 FARRAR® ULT Freezer Product Manual General Use Allow freezer to come to room temperature before switching power on. During initial startup, the high temperature alarm may sound while the freezer reaches operating temperature. Note This refrigeration system is designed to maintain ultra-low temperatures in an ambient environment (15°C to 32°C), only when the freezer is used to store materials at temperature.
  • Page 7: Installation

    FARRAR® ULT Freezer Product Manual 2. INSTALLATION 2.1 Location Has a dedicated 15 A grounded circuit with dedicated single point receptacle meeting the electrical requirements listed on the product specification label. Is clear of direct sunlight, high temperature sources, and heating and air conditioning vents. Minimum 8”...
  • Page 8: Ac Power Cord

    FARRAR® ULT Freezer Product Manual Install Rear Stand-Offs Rear Stand-off Align the holes in the stand-offs with the corresponding threaded holes on the back of the freezer. Insert the 3/8” hex head cap screws through the holes in the stand-offs. Hand-thread the cap screws into the threaded holes.
  • Page 9: Fuf Model Operation

    FARRAR® ULT Freezer Product Manual 3. FUF MODEL OPERATION 3.1 Initial Start Up Plug the power cord into a 15 A grounded circuit with a dedicated single point receptacle. Switch battery ON/OFF switch ON. Switch AC ON/OFF switch ON. Note The Start screen is displayed when the Smart Access Panel is powered on.
  • Page 10: Operation

    FARRAR® ULT Freezer Product Manual 3.2 Operation Refer to the Smart Access Panel User Guide for Ultra-Low Freezers for complete information regarding the Smart Access Panel User Interface. The Smart Access Panel Home screen displays temperature and alarm information, and provides icons to gain access to other functions of the Smart Access Panel.
  • Page 11: Set Alarm Parameters

    FARRAR® ULT Freezer Product Manual 3.4 Set Alarm Parameters > > Enter the Settings password. Scroll down to select Alarm Settings. Touch minus (-) or plus (+) on spin box to set each alarm parameter. Settings screen Alarm Settings screens Alarm settings control the circumstances and timing of alarm condition indicators displayed on the Smart Access Panel Home screen.
  • Page 12: Active Alarms

    FARRAR® ULT Freezer Product Manual 3.5 Active Alarms Home screen with active alarm Active Alarms Alarm Description Primary Monitor Probe High Temp Primary monitor probe reading is above high temperature alarm setpoint Primary Monitor Probe Low Temp Primary monitor probe temperature reading is below low temperature alarm setpoint Probe Failure Probe not functioning properly Ambient Probe High Temp...
  • Page 13: Min/Max Temperature Monitoring

    FARRAR® ULT Freezer Product Manual 4. MIN/MAX TEMPERATURE MONITORING The Min/Max temperature display provides the highest and lowest Primary Monitor probe temperature reading since the last system reset (power-on event) or manually-initiated reset. Touch the Reset icon to the right of the display to manually reset. Min Max Display Reset Notes The Min/Max temperature display can be turned on or off through Display Settings.
  • Page 14: Access Control

    FARRAR® ULT Freezer Product Manual 5. ACCESS CONTROL Allows user-specific secure access to the freezer. Notes The Supervisor PIN must be used to set up user profiles. The Supervisor PIN does not allow access to the unit. At least one user ID must be set up to gain access to the unit.
  • Page 15: Open Freezer With Access Control

    FARRAR® ULT Freezer Product Manual 5.2 Open Freezer with Access Control Access Control Keypad Enter a valid PIN using the keypad.
  • Page 16: Product Specifications

    FARRAR® ULT Freezer Product Manual 6. PRODUCT SPECIFICATIONS 6.1 Operating Standards Indoor use only Altitude (maximum): 2000m Ambient temperature range: 15°C to 32°C Relative humidity (maximum for ambient temperature): 80% for temperatures up to 25°C, decreasing linearly to 53% at 32°C Temperature control range: -50°C to -86°C Overvoltage Category II Pollution Degree 2...
  • Page 17 FARRAR® ULT Freezer Product Manual Note Overall exterior dimensions include casters, handles, Smart Access Panel bezel, and door hinges. Ultra-Low Freezer Specifications Model Voltage Code Amps Capacity Dimensions Interior Net Wt. Dimensions W x H x D in. (mm) Cu. Ft (Liters) W x H x D in.
  • Page 18: Compliance

    FARRAR® ULT Freezer Product Manual 7. COMPLIANCE 7.1 Safety Compliance This product is certified to applicable UL and CSA standards by a NRTL. This product is IECEE CB Scheme certified and complies with national differences for safety certification beyond IEC 61010-1:2010, IEC 61010-1:2010/AMD1:2016, IEC 61010-2-011:2016 7.2 Regulatory Compliance This device complies with the requirements of directive 2015/863 Amending Annex II to Directive 2011/65/EU for RoHS.
  • Page 19: Appendix A Application Icons

    FARRAR® ULT Freezer Product Manual APPENDIX A Application Icons Icon Description Icon Description Icon Description Icon Description Contact Home Reset Settings FARRAR Zoom Temperature Display Event Log Information Graph Brightness Information Mute Zoom Out Icon Transfer Temperature Download Access Log Cancel Graph Forward/ Back Arrows...