Veterinary Anesthesia Ventilators
The latest design in precision digital ventilation.
User Manual

Advertisement

Table of Contents
loading

Summary of Contents for Hallowell EMC 2024 PRO

  • Page 1 Hallowell EMC Model 2024 & Model 2024IE Veterinary Anesthesia Ventilators The latest design in precision digital ventilation. User Manual...
  • Page 2: Table Of Contents

    TABLE OF CONTENTS USER/OWNER RESPONSIBILITY ................3 WARNINGS AND CAUTIONS ................... 3 WARRANTY ......................4,5 INTRODUCTION ......................6 RECEIVING PROCEDURES ..................7 SET-UP PROCEDURE/ SET-UP DIAGRAM ............. 8,9,10 CONTROLS (FRONT & REAR PANEL) ............. 11,12 ALARMS & LEDS ....................13 VERIFICATION OF PROPER FUNCTION ............
  • Page 3: User/Owner Responsibility

    Hallowell EMC. Equipment that is not functioning correctly should not be used. This equipment and any of its accessories or component parts should not be modified.
  • Page 4: Warranty

    WARRANTY The Model 2024 /2024IE Veterinary Anesthesia Ventilator is covered under the warranty expressed on the warranty card attached to the unit at the time of sale to the end user, which reads as follows: LIMITED WARRANTY STATEMENT WHAT THIS WARRANTY COVERS: HEMC offers you a limited warranty that the enclosed subscriber unit and its enclosed accessories will be free from defects in material and workmanship, according to the following terms and conditions: 1.
  • Page 5 To obtain warranty service, you may contact us by telephone, fax, or email at: Tel. 1-413-445-4263, Fax. 1-413-496-9254 or info@hallowell.com. Visit www.hallowell.com Correspondence may also be mailed to: Hallowell EMC 35 Downing Industrial Park Pittsfield, MA 01201 Prices, terms, and product specifications are subject to change without notice.
  • Page 6: Introduction

    Optionally, Hallowell EMC provides mounting hardware for the Matrx VMC and Spartan, VMS and VML anesthesia machines. Mounting hardware is also available for anesthesia machines manufactured by A.M.
  • Page 7: Receiving Procedures

     External 12VDC Power Supply (PN 090A0280) w/ Country Kit (070A5174), medical safety approval: ANSI/AAMI ES60601-1/60601-1-11  Warranty Registration Card (DOCA5967B). Complete and return to Hallowell EMC.  Card Spin USB Flash Drive (w/ Operating Manual) (125A1845)  Pro Series Vent. Accessory Kit (000A6557): , DISS Inline Pressure Gauge.
  • Page 8: Set-Up Procedure/ Set-Up Diagram

    SET-UP PROCEDURE Inspect the control unit for debris from shipping. Inspect all three ports, the 50 psi SUPPLY GAS, EXHAUST, and DRIVING GAS ports on the back of the ventilator and remove any obstructions that may have become lodged inside during shipping and unpacking. Inspect the bellows assembly for debris from shipping.
  • Page 9 SET-UP PROCEDURE (Continued) Insert the Airway Pressure Sampling Tee into the breathing system Disconnect the patient breathing hose from the INHALE VALVE of the anesthesia machine. Connect the Airway Pressure Sampling Tee (PN 000A2420B) to the INHALE VALVE and reconnect the patient breathing hose to the Airway Pressure Sampling Tee.
  • Page 10 4/17/2024 DOCA6471 Operating Manual 2024PRO 10 of 24...
  • Page 11: Controls (Front & Rear Panel)

    CONTROLS (FRONT PANEL) Power Switch (PWR) - A maintained switch with built in green LED. Use to turn the ventilator ON and OFF. The green LED is illuminated to indicate the power is ON. (Pic.1 Front Panel, Page 12 – Ref. #1) Standby/Ventilate (STBY/VENTILATE) –...
  • Page 12 Panel Pic. 1 Front CONTROLS (REAR PANEL) Power Supply Inlet (12Vdc @ 2A) - a 2.1mm jack for external medical safety approved (ANSI/AAMI ES60601-1/60601-1-11) power supply (PN 090A0280). Fuse Holder (2A) – for fuse, 5x20mm 2A 250V Fast VDE (PN 053A1782). Features that can be installed by request are: MANUAL REMOTE INSPIRATORY HOLD.
  • Page 13: Alarms & Leds

    ALARMS and LEDs Maximum Working Pressure Limit Alarm (MWPL): The MWPL feature allows the operator to set an upper limit above which the airway pressure will not exceed. The ventilator will terminate the inspiratory phase of the breathing cycle and begin an expiratory phase when the pressure transducer senses a pressure above the MWPL setting.
  • Page 14: Verification Of Proper Function

    VERIFICATION OF PROPER FUNCTION A Note on Test Lungs The most readily available test lung will probably be the breathing bag you removed to connect the ventilator. A breathing bag is a very poor model of a lung. It can be used if one understands how poor a model of the lung it is and how to avoid using it in such a way that it creates problems that will not occur with a real patient.
  • Page 15 VERIFICATION OF PROPER FUNCTION (Continued)  Depress the INSP HOLD button. Yellow LED above will be on. Bellows stops moving. Hold it long enough, but no longer than 60 sec to verify that the breathing system is not leaking. The pressure should remain constant.
  • Page 16: Getting Started

    GETTING STARTED A TYPICAL USAGE SCENARIO: The Model 2024 /2024IE is a time cycled volume ventilator with an adjustable breathing pressure limit. The more you understand of the ventilator, how it works, and what it does in response to your settings, the more comfortable you will be with it.
  • Page 17 GETTING STARTED (Continued) This MV can be delivered in many ways from a few large TVs to a lot of small TVs. The most optimum combination is up to you to determine just as you would while bagging. Our ventilator, for those of you familiar with the term, is considered a MV divider. For those unfamiliar with the term, the ventilator delivers a consistent MV to the patient and that MV is divided into different size TVs by the RATE control.
  • Page 18: Operating Instructions

    OPERATING INSTRUCTIONS Output of the is adjusted by only two controls: a linearly Model 2024 /2024IE calibrated RATE control (breaths per minute), and a metering VOLUME control. Follow the setup and verification procedures and be certain the pop-off valve on the anesthesia machine is completely closed.
  • Page 19: Cleaning & Sterilization

    CLEANING & STERILIZATION A majority of the components do not come in contact with Model 2024 /2024IE breathing gas; consequently, they require cleaning with only a damp cloth. This includes the entire control unit as well as certain bellows assembly components. Only the bellows base interior and the inside of the bellows come in contact with the breathing system gases.
  • Page 20 CLEANING & STERILIZATION (Continued) After cleaning the pop-off valve seat, replace the valve taking care to ensure that the small orange O-ring (PN 180A1429) under the pop-off valve is securely in place. If this orange O-ring is dislodged or missing, the ventilator will not be able to function properly. Sterilizing the bellows base from the controller and bellows interior surface: The bellows and its interior surface do come in contact with the breathing gas and require periodic sterilization.
  • Page 21: Troubleshooting

    TROUBLESHOOTING Leaks in the circle (breathing) system are very common. Particularly, the reuse of "single use" circle systems frequently results in leaks. This reuse practice is not recommended. Circle system leaks are not as apparent when bagging a patient or when using an older style ventilator with a falling-during- exhalation bellows system.
  • Page 22   Inspiratory flow chugs intermittently. Pressure within the bellows Inflate collapsed bellows. housing is exceeding 70 cm H 2 O.  Unkink the drive gas tube.  Working pressure limit is  Unkink breathing system hose. shutting off the switch Inspiratory flow.
  • Page 23: Returning For Service

    Instructions Sheet.” This sheet will have the RA # at the top. Clearly mark the RA # on the outside of the box that you will be returning to Hallowell EMC. Please follow the procedures faxed to you carefully. It is based on our experience of how best to get your items back to us without damage and delay.
  • Page 24: Model 2024

    Model 2024 /2024IE by HALLOWELL EMC SPECIFICATIONS OPERATIONAL CHARACTERISTICS Rate……………………………………………….2-80, (6-40 and 0.5-60 by request) bpm Tidal Volume……………………………………. 0-3000 ml I:E Ratio…………………………………………. 1:2 (preset), 1:1 to 1:4 (IE version) Maximum Working Pressure Limit ……………. 10 - 60 cm H Gas Supply………………………………………. Oxygen or Clean Dry Air Flow &...