Summary of Contents for Protec PROGNOST SH 7040-5 Series
Page 1
*The PROGNOST SH does not include any X-ray components (X-ray tube, collimator, X-ray generator) Version: 10.0 PROTEC GmbH & Co. KG Issued: 2024-04-03 In den Dorfwiesen 14, 71720 Oberstenfeld, Germany Subject to alterations Telephone: +49 (0) 7062/92 55-0 E-Mail: protec@protec-med.com...
Page 2
All sheets of this document contain proprietary and confidential information of PROTEC GmbH & Co. KG and is intended for exclusive use by current PROTEC GmbH & Co. KG customers. Copying, disclosure to others or other use is prohibited without the express written consent of PROTEC´s law department.
Cleaning ....................................18 5.3.2 Disinfection ................................... 18 Inspection and Maintenance ..............................19 5.4.1 Daily Monitoring before and during the Examination Operation............19 5.4.2 Regular Monitoring ................................19 PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 3 of 30...
Page 4
Description of Symbols, Labels and Abbreviations ............. 27 Symbols ....................................... 27 Type Label ......................................28 Labels ........................................28 Positions of the Signs and Labels............................29 Abbreviations ....................................30 PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 4 of 30...
PROTEC GmbH & Co. KG. Assembly personnel and other persons who are not employees of the technical service department of PROTEC GmbH & Co. KG are requested to contact the local branch of PROTEC GmbH & Co. KG before assembly or service work is started.
Hardware and Network System Requirements As a stand-alone product, the analogue version of the PROGNOST SH has no hardware or network connection and therefore no hardware or network requirements. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 7 of 30...
1.2.3 Installation NOTE The installation of the PROGNOST SH must be performed by PROTEC service department or a service company authorized by them. For more information, please see separate “Installation manual PROGNOST SH”. Contact information of persons qualified to perform installations are available upon request at: PROTEC GmbH &...
Settings and calibrations that are not described in these instructions for use must be carried out using the technical description of the device by PROTEC service department or a service company authorized by them.
X-ray system, the connected X-ray generator or the Power Box must be switched off. 2.1.2 Device Operation In case of a malfunction, do not use the PROGNOST SH anymore and notify PROTEC service department or a service company authorized by them. 2.1.2.1 Operating Type The PROGNOST SH is not intended for continuous operation.
The PROGNOST SH is intended for use in an environment in professional health care facilities (e.g., clinics, surgery centres, physiology practices ...). PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 12 of 30...
X-ray system tube support, floor stand 1 Indication of rope breakage 2 Operating unit with Touch-Display 3 Keypad 4 Operating unit with inclinometer 5 Handle 6 Foot pedal PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 13 of 30...
While comprising the handles, the electromagnetic locking of one or more movements can be released by pressing the buttons on the operating unit with the thumb, and the X-ray tube unit can be brought into the desired position. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 14 of 30...
While comprising the handles, the electromagnetic locking of one or more movements can be released by pressing the buttons on the operating unit with the thumb, and the X-ray tube unit can be brought into the desired position. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 15 of 30...
The detent in the new position centers itself. WARNING! There is an increased risk of injury if the X-ray floor stand is not engaged! PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 16 of 30...
Special feature for the Touch option: The system is only switched off if the Touch-Display is completely shut down. Do not restart the system until the Touch-Display is completely shut down. Otherwise, the Touch-Display does not start. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 17 of 30...
(e.g., torn cable, bent parts) or if the resulting image quality (e.g., artifacts in the image) is insufficient. In this case, please contact the PROTEC service department or a service company authorized by them immediately.
12 Months by the PROTEC service department or a PROTEC authorized service provider. In the case that the intended safety-related checks are not carried out, PROTEC GmbH & Co. KG accepts no liability for damage to the operator and third parties if and insofar as damage results from insufficient or non-performed checks.
The PROGNOST SH is designed for a service life of 10 years when used in accordance with the specifications and regular maintenance by PROTEC service department or a service company authorized by them. After the product has reached the end of its service life, further use is at your own risk.
Harmonic emissions Class A buildings that are used for residential purposes, EN 61000-3-2 provided the following warning is observed: Voltage fluctuations/ compliant PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 21 of 30...
Page 22
80 MHz to 2.7 GHz NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structure, objects and people. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 22 of 30...
Page 23
2450 2400 – 2570 b/g/n, 217 Hz RFID 2450, LTE Band 7 5240 Pulse modulation: 5500 5100 - 5800 WLAN, 802.11 a/n 217 Hz 5785 PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 23 of 30...
V10.0_2024-04-03 PROGNOST SH Instructions for use 5040-0-8002 Technical Data Dimensions *OP = Operator PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 24 of 30...
For wagon with rotation and telescopic arm (X-ray tube and collimator 27.5 kg): 398 kg NOTE The weight of the X-ray tube with collimator must not exceed 27.6 kg for the telescopic arm. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 25 of 30...
Environmental Conditions for Shipping and Storage Ambient Temperature - 10°C to + 70°C Relative humidity 10% to 95% (non-condensing) Atmospheric pressure 500 hPa to 1060hPa PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 26 of 30...
Caution: pinch-/crushing hazard for hands and fingers This symbol indicates the need to observe the instructions for use. This is provided in an electronic format (eIFU) on our website. www.protec-med.com/download PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 27 of 30...
Electrical release for rotating the X-ray tube stand by ± 180° (optional) Type Label Labels Labels on the X-ray system tube support Caution: pinch-/crushing hazard for hands and fingers Refer to Instructions for use PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 28 of 30...
V10.0_2024-04-03 PROGNOST SH Instructions for use 5040-0-8002 Caution: possibility of feet crushing Climbing forbidden Positions of the Signs and Labels PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 29 of 30...
Centimetre Kilogram °C Degree -Celsius Hectopascal German Industry Standard European Standard CE-Mark Hertz Duty cycle Ampere Serial number Volt (DC Voltage) Volt (AC Voltage) Watt PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 30 of 30...