  • Page 1 Distribution Holder CF-501 Series Instruction Manual CODE: M000793-3200305802-GZ0000214589...
  • Page 2 Advanced Techno, Co., Ltd., any malfunctioned or damaged Product attributable to responsibility of HORIBA Advanced Techno, Co., Ltd. for a period of one (1) year from the delivery unless otherwise agreed with a written agreement. In any one of the following cases, none of the warranties set forth herein shall be extended;...
  • Page 3 For Your Safety Hazard Classification and Warning Symbols Warning messages are described in the following manner. Read the messages and follow the instructions carefully. Hazard classification This indicates an imminently hazardous situation which, if not avoided, will result in Danger death or serious injury.
  • Page 4 Safety Precautions This section provides precautions to enable you to use the product safely and correctly and to prevent injury and damage. The terms of “DANGER” and “WARNING” and “CAUTION” indicate the degree of imminency and hazardous situation. Read the precautions carefully as it contains important safety messages.
  • Page 5: Table Of Contents

    Contents Overview ....................1 Unpacking....................2 Names of Parts ..................3 Distribution Holder ....................3 Installation ..................... 4 Piping ........................5 Measurement ..................6 Before Measurement....................6 Maintenance ..................7 Replenishing the Internal Liquid (3.3 mol/L KCl) ............. 7 About Calibration by Standard Liquid ..............7 Cleaning an Electrode.....................
  • Page 7: Overview

    Overview Overview This Distribution Holder can be used only for a system under which a measured liquid line is opened to atmospheric pressure. If an internal pressure exists in the measured liquid line, please consult with HORIBA Advanced Techno.
  • Page 8: Unpacking

    Breakdown of Complete Set (standard) Distribution Holder Accessories O-ring (P44) ....1 Instruction Manual Distribution pH/ORP Holder CF-501 Series Instruction Manual After unpacking, store the accessories carefully not to lose them. Options ● Electrode ● Internal liquid (3.3 mol/L KCl)
  • Page 9: Names Of Parts

    Names of Parts Names of Parts Distribution Holder Tightening Nut Sensor Adapter Plastic Body Sensor Tightening Nut Lock Plate O-ring (P44) Fixing Screw Distribution Holder Distribution [Ordinary State] [With the Sensor and the Adapter Being Detached] Caution Caution Against Blowout If a clamping nut is loosened, the liquid in a Distribution Holder may blow out.
  • Page 10: Installation

    Installation Installation Be sure to following the following instructions for setup. Install the Distribution Holder at a location where maintenance work can be easily performed. Leave a maintenance space of 15 cm or more at the top of the CF-501. Give room to an electrode cable to detach it for maintenance, etc.
  • Page 11: Piping

    Installation Piping For installation of the Distribution Holder, provide a sampling line from the main line so that the measured liquid flows into the bottom side of the Distribution Holder and flows out of the lateral side of the Distribution Holder. Make the length of an outlet pipe of the Distribution Holder as short as possible.
  • Page 12: Measurement

    Measurement Measurement Before Measurement A sensor is not factory-incorporated in this Holder. Read the Instruction Manual for the Sensor carefully prior to use. Incorporating a Sensor Note Tightening Nut An electrode is made of glassware. An electrode may be broken if any shock or strong force is applied. 1.
  • Page 13: Maintenance

    Maintenance Maintenance Replenishing the Internal Liquid (3.3 mol/L KCl) Maintain the level of internal liquid (3.3 mol/L KCl higher than the level of measured liquid. Replenish the consumed amount of internal liquid periodically. For the method of replenishing, see the Instruction Manual for the KCl Tank. About Calibration by Standard Liquid For a pH Meter The properties of an electrode change over time, thus causing errors between measured values.
  • Page 14: Cleaning An Electrode

    Maintenance Cleaning an Electrode Any dirt of an electrode may cause an decrease in response speed, indication drift, and instability. Check the tip of an electrode (metallic pole and responsive film) periodically. Clean all dirt and coatings off an electrode with water to maintain the electrode and the liquid junction section in a clean condition.
  • Page 15: About Dirt Of Orp Electrode (Metallic Pole)

    Maintenance About Dirt of ORP Electrode (metallic pole) If dirt is not wiped off as a result of following steps in the “Method of Various Dirt of an Electrode” as described above, causing any troubles in making measurements, place a polishing agent (cerium oxide of about 1 μm in particle size) on a polishing pad, and then drop deionized water slightly.
  • Page 16: Attaching And Detaching A Sensor

    Maintenance Attaching and Detaching a Sensor Attaching a Sensor See “Before Measurement” and “Incorporating a Sensor”. Detaching a Sensor Note An electrode is made of glassware. An electrode may be broken if strong force is applied. 1. Stop the supply of measured liquid from the sampling line. 2.
  • Page 17: Specifications

    Moreover, measurements cannot be made when a measured liquid is in a freezing or boiling state. If any problem with weatherability occurs under direct sunshine, use a holder made of PVC or a holder made of SUS316+PPS. For the sample properties that affect FKM (fluorine rubber) (strong alkali, etc.), please consult with HORIBA Advanced Techno.
  • Page 18: Replacement Parts

    Replacement Parts Replacement Parts Replacement of consumables Internal liquid Part Name Model Specifications Part No. Internal liquid for comparison electrode #300 3.33 Mol/L KCl solution 250 mL 3200043618 (For pH, ORP, F-) Powder for comparison electrode #350 KCl powder 500 g 3200043623 internal liquid (For pH, ORP, F-)
  • Page 19: Method Of Disposal

    Method of Disposal Method of Disposal For disposal of this instrument, depose of the instrument in accordance with the laws, ordinances and regulations stipulated in local regions. Parts of this product are made of following materials. Plastic Body Sensor Pan-head screw M5×8, 3S (SUS304) Names CF-501 CF-501P...
  • Page 21 Headquarters 31, Miyanonishi, Kisshoin Minami-ku, Kyoto, 601-8306 TEL:+81-75-321-7184 FAX:+81-75-321-7291 Osaka Office 4F Shin Osaka Ueno Toyo Building, 7-4-17, Nishinakajima Yodogawa-ku Osaka, 532-0011 TEL:+81 6 6390 8211 FAX:+81 6 6390 8222 Tokyo Office 4F Higashikanda Fukoku Seimei Building, 1-7-8, Higashikanda, Chiyoda-ku, Tokyo 101-0031...