SereneLife SLOTOSPE016 - Ophthalmoscope Manual

SereneLife SLOTOSPE016 - Ophthalmoscope Manual

Advertisement

SereneLife SLOTOSPE016 - Ophthalmoscope Manual

SYMBOLS

Type B Applied Part
warning Attention, consult accompanying documents

SAFETY INFORMATION

To assure the correct use of the product, b.c safety measures should always be followed including the warnings and cautions listed in this instruction manual.

WARNINGS & CAUTIONS:

  • Do not use the device where it may be exposed to flammable gas.
  • Protect this device from water, moisture, Impact, dropping, direct sunlight, heat and fire.
  • Do not touch the lens directly by hand or something hard.
  • Please use the bulb type specified in this instruction brochure.
  • Keep the device out of the reach of unsupervised infants and children.
  • Do not disassemble or attempt to repair the device or components.

GENERAL

DESCRIPTION

HS-OP10 ophthalmoscope (hereinafter: device) is a classic medical device for ophthalmology. This device adopts a professional lens system, which makes doctors conveniently get a clear visual field for examining the interior structures of eyes, such as dioptricmedia etc.

TECHNICAL SPECIFICATIONS

  • Enhanced XHL Xenon Halogen Technology.
  • 8106 bulb used.
  • 5 different apertures
  • Range of lenses: -20 D to +20 D
+ 1 2 3 4 6 8 10 15 20
- 1 2 3 4 6 8 10 15 20

HOW TO USE THIS DEVICE

  • Turn on (or off) the device by pushing downwards (or upwards) as fig.1
  • Scroll the diopter wheel in two directions as fig.2 to select a suitable diopter, which helps to get a clearer observation vision for user
  • Scroll the diaphragm wheel as fig.3 for getting a needed aperture
  • Diagnose by aiming the light to the patient's eye as fig.4.

Attention:
it is an intermittent operated device, 1 minute ON, 5 minutes OFF

MAINTENANCE

Changing the bulb

  1. Turn off the power and unscrew the head from the handle of device
  2. Pull the old bulb downwards, and replace with new one (fig.5)
  3. Screw the head back on to the handle.

Attention:

  • Please use the specialized bulb (Model: 8106) for this device.
  • After the bulb burns out, the temperature will remain high for approximately 5 minutes. Please wait for a moment to avoid injury.

Changing the batteries

  1. Unscrew the bottom cover from the handle
  2. Remove the old AA batteries, and replace with new ones. Ensure the anode of the batteries points to the head of the device (fig.6)
  3. Screw on the bottom cover of the handle.

Attention:

  • AA alkaline batteries are recommended.
  • Replace all batteries simultaneously and do not use alkali manganese batteries with lithium as bulb life will be reduced.
  • Do not leave worn batteries in the handle.
  • Please take out batteries from the device if not in use for a prolonged period.

Cleaning and Disinfection

Clean the surface of device with 75, alcohol solution at least once a week, avoiding the lenses in the head. And corrosive or polishing detergent is prohibited.

Attention:
Do not soak, autoclave or let water dip into the device!

PERIODIC SAFETY CHECKS

Preventive inspection and maintenance to be performed including the frequency of such maintenance. The following safety check shold be performed at least every 12 months by a qualified person who has adequate training, knowledge, and practical experience to perform these tests.

  • Inspect the equipment and accessories for mechanical and functional damage.
  • Inspect the safety relevant labels for legibility.
  • Verify that the device functions properly as described in the instructions for use.

SPECIFICATION

Illumination Direct Illumination
Luminance 18. 5 Lumen
Bulb Specified Model: &106
2.5V, 0.7A, Φ = 4.8mm
Batteries 2 AA alkaline batteries
Operation Temperature 50°F~ 95°F
Humidity 30% ~ 75%
Air pressure 700 hPa - 1060 hPa
Storage & Transport Temperature -68°F ~ 158°F
Humidity 10% ~ 90%
(without condensation)
Air pressure 500 hPa - 1060 hPa
Standards MDD 93/42/EEC Class I
IEC 60601-1:1995, IEC 60601-1-2:2007

ELECTRICAL CLASSIFICATION

Equipment Classification with respect to protection from electric shock: Internally powered equipment.
Degree of protection from electric shock: TYPE B.
Equipment not suitable for use in the presence of flammable mixture: Yes.
Mode of operation: intermittent operation.

The manufacturer reserves the right to make technical changes without notice in the interest of progress. Prior notices will not be given in case of any amendments within this manual.

VISIT US ONLINE:
Have a question?
Need service or repair?
Want to leave a comment?

PuleUSA.com/ContactUs

Documents / Resources

Download manual

Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.

Download SereneLife SLOTOSPE016 - Ophthalmoscope Manual

Advertisement

Need help?

Need help?

Do you have a question about the SLOTOSPE016 and is the answer not in the manual?

Questions and answers

Table of Contents