The base language for this document is ENGLISH. Any translations must be from the base language document. Printed copies are not controlled documents. Although current at the time of publication, Skytron's policy of continuous development makes this manual subject to change without notice. If current manuals are required, contact your local Skytron representative or contact SKYTRON ®...
Page 3 TABLE OF CONTENTS TITLE PAGE SECTION 1. GENERAL AND SAFETY INFORMATION ..........5 1-1. Training Required Before Use ...................5 1-2. Use Statement ......................5 1-3. Service Statement .....................5 1-4. Service Contact Information ..................5 1-5. Tools and Supplies Required ..................5 1-6. Basic Installation Sequence ..................5 1-7.
Page 4
Page 4 TABLE OF CONTENTS TITLE PAGE SECTION 5. MAINTENANCE AND SERVICE ...............20 5-1. Information for Service Personnel ................20 5-2-1. Preventive Maintenance ..................20 5-2-2. Service.......................20 SECTION 6. REPLACEMENT PARTS ................21 6-1. Sink Components .....................21 6-2. Plumbing Components .....................22 6-3. Eye Wash Components ....................23 SECTION 7.
1-6. Basic Installation Sequence combination of this product with other products, 1. Prepare the wall as necessary to structurally whether supplied by Skytron or by or by other support, plumb faucets, drains and valves manufacturers, unless such a combination has been 2.
Page 6 1-8. Safety Precautions This hot water master tempering valve The following is a summary of WARNINGS and cannot be used for tempering water CAUTIONS indicated throughout this manual. These temperature at fixtures. Severe bodily injury precautions are found where applicable. Carefully (i.e., scalding or chilling) and/or death read this section before operating the equipment.
Page 7
When cleaning the sink DO NOT use abrasive maintenance personnel should troubleshoot pads, scrapers, steel wool, or a wire brush. the SKYTRON Stainless Surgical Scrub When cleaning the sink DO NOT use water with Sink. Trouble shooting by unauthorized a pH higher than 7.0, hydrochloric acid, steam or...
If there is any damage, refer ■ “Optional Triple Sink Carrier” on page 18 to page 26 for damaged claim procedure. For seismic applications please contact your Skytron 2. Check water supply and waste terminal representative for seismic documentation.
Page 10 3-3. Attach Sink to Support Brackets 3-4-1. Sensor Initialization The sensors are pre-set and equipped with a logic 1. Install sink assembly on brackets making sure board. The sensors determines the range during the sink assembly is captured by the wall “Z” initialization period (The time after initial power until bracket(s) (Figure 2).
Page 11 3-5-1. Flow Timer Controller Specifications CAUTION Service Compatible Liquids Wetted Materials Housing: Brass See Switch Ratings before connecting power. Piston: polysulfone Flow settings for this switch is normally Spring: 316SS calibrated using water @ +70°F on O-Ring: Fluoroelastomer increasing flow.
Page 12 3-6-1. Connect Drain and Water Lines Periodic inspection and yearly maintenance by a licensed contractor is required. Corrosive water conditions, temperatures WARNING over 200°F, unauthorized adjustments or Local building or plumbing codes may repair could render the valve ineffective require modifications to the information provided.
SECTION 4. OPERATION AND CARE Page 19 4-1. Operation CAUTION ■ Press knee kick panel to activate water flow. When cleaning the sink DO NOT use water with a pH higher than 7.0, hydrochloric acid, steam or ■ Adjust water temperature with the mixing valve. high pressure water, bleach or any compounds ■...
Service this product according to the instructions provided in this manual. service manual Use original Skytron parts only The safety, reliability, and performance of the equipment can only be ensured if Skytron original replacement parts and accessories are used. Modifications and unauthorized Unauthorized repairs, temporary repairs, or modifications are not repairs permissible and will void the product’s warranty.
SECTION 7. TROUBLESHOOTING Page 24 WARNING Only facility-authorized SKYTRON trained, maintenance personnel should troubleshoot the SKYTRON Stainless Surgical Scrub Sink. Trouble shooting by unauthorized personnel could result in personal injury or equipment damage. Low water pressure: Confirm that the facility water pressure meets If not, the facility will need to supply sufficient the 20 - 50 PSIG requirements.
Page 25
SECTION 7. TROUBLESHOOTING CON’T. Page 25 Eye wash Water temperature is too hot or too cold. Inspect blending valve and adjust if necessary. troubleshooting: Will not turn on or off. Verify that sensor eye has a red light visible and flashes when it detects a target. Check that 120 volts are being supplied to If not, the facility will need to correct power at the transformer.
An example of an applicable honor the claim, as long as SKYTRON receives the statement would be; “Received in good order except full cooperation of the consignee in filing the claim. as noted” or “Crate damaged, possibility of concealed damage.”...
9 Image for F-1-010-5-M measurement changed from 35-25/64” to 32-3/8” 08-25-23 Revised manual layout to current Skytron standards. Added (Includes Operation, Installation, Maintenance and Parts) to Cover Added SECTION 1 on page 5 GENERAL AND SAFETY INFORMATION.
Page 28
5085 Corporate Exchange Blvd. S.E. Grand Rapids, MI 49512 • 1.616.656.2900 • FAX 1.616.656.2906...
Need help?
Do you have a question about the SS2121-MK and is the answer not in the manual?
Questions and answers