Do you have a question about the ESCM-900 and is the answer not in the manual?
Questions and answers
Summary of Contents for Edaphic Scientific ESCM-900
Page 1
User Manual and Installation Guide ESCM- 900 Indus trial CO2 Detector Distributed & Supported By: Edaphic Scientific Pty Ltd www.edaphic.com.au info@edaphic.com.au Ph: 1300 430 928...
Page 2
(877) 678-4259 the manner specified, the protection provided by the equipment may be diminished. This equipment should be installed/serviced by qualified personnel only. 2 | P a g e Edaphic Scientific | www.edaphic.com.au | info@edaphic.com.au | Ph: 1300 430 928...
Introduction to the ESCM-900 The ESCM-900 Industrial Carbon Dioxide Storage Safety Alam is a (12-24) VDC powered monitor that is designed to reliably and accurately sense the CO2 levels present in confined spaces, harsh environments, or industrial settings. The safety alarm uses nondispersive infrared (NDIR) sensing technology to measure CO2 levels in areas of potential risk.
Alarm Icon: Appears after AL1 is triggered and will remain on display while the monitor is in alarming state Ventilation Icon: CO2 Levels exceed AL2 and Relay2 has been triggered. 6 | P a g e Edaphic Scientific | www.edaphic.com.au | info@edaphic.com.au | Ph: 1300 430 928...
32°F to 122°F (0°C to 50°C) Humidity Range 0 ~ 95% RH non-condensing Storage Conditions: Storage Temperature -4°F to 140°F (-20°C to 60 °C) 7 | P a g e Edaphic Scientific | www.edaphic.com.au | info@edaphic.com.au | Ph: 1300 430 928...
The ESCM-900 is designed to withstand harsh environments that would compromise other comparable safety monitors. Varying temperature, humidity, and dust/debris levels will not negatively impact the performance of the ESCM-900. For optimal use, a few considerations should be reviewed while selecting a mounting location: Avoid a location that risks high pressure washdown directed at the monitor.
Operation Start-up Verify that the ESCM-900 is properly wired, and 12-24VDC power is being supplied to the correct pin positions in the mating connector (see page 10). Check all wiring connections before powering. After power-up, the monitor provides accurate CO2 measurements after a 5 second warm up.
2. Press Enter to select alarm 3. AL3 level will now flash. 3. Use Up/Down arrow to select new alarm level. 4. Press Enter to confirm. 12 | P a g e Edaphic Scientific | www.edaphic.com.au | info@edaphic.com.au | Ph: 1300 430 928 G u i d e...
2. Press Mode, and choose Baro Mode 3. Press Up/Down to set Baro Mode On/Off. 4. Press Enter to Save. 13 | P a g e Edaphic Scientific | www.edaphic.com.au | info@edaphic.com.au | Ph: 1300 430 928 G u i d e...
7. If “PASS” press Enter to save. 8. If “FAIL” repeat the process. 9. If “FAIL” more than twice, contact CO2Meter technical support. 14 | P a g e Edaphic Scientific | www.edaphic.com.au | info@edaphic.com.au | Ph: 1300 430 928 G u i d e...
2. Press Enter Up/Down to choose either “Yes” or “No”. 3. Press Enter again to save the setting or press Mode to quit the setting. 15 | P a g e Edaphic Scientific | www.edaphic.com.au | info@edaphic.com.au | Ph: 1300 430 928 G u i d e...
Distributed & Supported By: Edaphic Scientific Pty Ltd www.edaphic.com.au info@edaphic.com.au Ph: 1300 430 928 16 | P a g e Edaphic Scientific | www.edaphic.com.au | info@edaphic.com.au | Ph: 1300 430 928 G u i d e...
Need help?
Do you have a question about the ESCM-900 and is the answer not in the manual?
Questions and answers