P a g e 1335 Eagandale Ct, Eagan, MN 55121 Imperium 10010W Thermal System THM-SYS-10W2-01 User Manual December 6 , 2022 Rev A 12/6/2022 Original...
P a g e Table of Contents General Safety Instructions ........................3 Dry Air Supply Requirements ........................ 5 Performance Specifications ........................5 Installation ..............................6 Location ..............................6 Environmental ............................6 Setting Up the System ..........................7 Starting Up the System ..........................8 Operation ..............................
The system also contains Polyolester (POE) oil, which is used by the compressor. Contact Ironwood Electronics for the MSDS for either of these internal fluids.
Page 4
Directly connect the system to a dedicated electrical outlet in accordance with local codes. Do not use an extension cord or power strip to connect the system to an electrical outlet. If the supplied power cable is too short, consult with Ironwood Electronics for the appropriately rated replacement.
Consult your air dryer specification to determine its rating or consult the operator of the lab space. The compressed air should also be able to reach at least 90 PSI and 0.5 cfm for proper operation of the force control systems. Performance Specifications THM-SYS-10W2-01 Temperature range -55°C to 150°C Cooling Power 10 W @ -40°C...
P a g e Installation Location Note the following when choosing a location for the chamber: 1. Heat is produced when the chamber operates; enough ventilation must be provided to prevent excess heat build-up. At least 12 inches of space around the system is recommended for maximum performance.
If the value exceeds 8 reach out to Ironwood Electronics support. 5. Open the coolant fill port and verify the fill tank is between 50 – 80% filled. Do not overfill as the fill port cap needs space to seal on.
P a g e modes and a pop up menu on the bottom used to move between the different menus. The standard overlay is made up of the following components: 1. Start- Touching the start button will begin either a static or profile mode test depending on which mode is enabled.
0 to 80 Kgf. Refer to the socket drawing for the required force for the socket. If a force is not given contact Ironwood Electronics to get the correct value. NOTE: adjusting the socket force value while the force button is lit will not update the force system. The...
P a g e | 11 Performing a test Static Mode The static mode screen is made up of the temperature set point and ramp rate buttons. To adjust the temperature set point, touch the associated blue button and set the desired temperature. To adjust the ramp rate, touch the associated blue button and set the desired ramp rate.
P a g e | 12 Profile Mode Remote Mode To access the remote port, you will need a USB-A to USB-B cable. Insert the USB-B connector to the “Remote” port on the front of the system and connect the USB-A connector to a serial machine. The current supported commands are as follows: “Start the system”...
P a g e | 13 Graph Mode What to Do if the System Doesn’t Operate Maintenance Cleaning and Decontamination Cleaning and decontamination are crucial to ensure proper system performance. Before operation or application of TIM, make certain the thermal head, device, and socket are clear of any dirt or oils. It is recommended to wipe down contact surfaces on the thermal head with an alcohol wipe between uses.
Replacement Coolant may be purchased directly from Ironwood. DO NOT use coolants that are not supplied by Ironwood. It is recommended to replace the coolant after 1 year of use. You will need a 10mm hex wrench to access the coolant reservoir on the top of the system. Once open, siphon as much coolant as possible from the reservoir with a syringe.
P a g e | 15 Software Updates 1. Download software updates from https://www.ironwoodelectronics.com/products/imperium- thermal-system-10010w/ 2. Once downloaded, move the file to empty flash drive. This should be the only file on the flash drive. 3. With the system off, insert the flash drive into the “software” port on the front of the system as seen below: 4.
This Limited Warranty shall not apply to any product or component thereof which has been repaired or altered outside of Ironwood’s factory in any manner so as, in Ironwood’s sole judgment, to affect its serviceability, or to any product that has been subject to alteration, accident, misuse, abuse, neglect or normal wear.