The Auxiliary Input accepts signals ranging from -10V to +10V and the conversion
value, identified by a scale-factor, can be stored in the PS in order to have the correct
reading from it.
Example: for a temperature sensor having a 20 mV/°C gain, the input scale-factor
needs to be configured considering that the equivalent temperature at a potential ADC
full-scale of 5 V would be the following:
Having the ADC a 16-bit resolution, the LSB value, equivalent to the scale-
factor that needs to be saved to the PS for a correct reading is:

K-type thermocouple connector

The option FB1K5OPT0001 includes the possibility to connect a K-type
thermocouple. The board mounts a MAX31856MUD+ IC that digitizes the input signal
coming from the thermocouple via SPI-interface.

3.5 Output connectors

The load needs to be connected to the output connector placed on the PS rear
panel. Two types of connectors are provided depending on the maximum output voltage
of the PS.

High-Current (HC) Models

For High-Current models (HC, see Table 1.1), busbars terminals are present on
the rear panel, see Figure 22(a), to offer a convenient and reliable way of connecting
the load. Figure 22(b) shows the output terminals with the protecting cover, while
Figure 23 shows a proper connection with the load cables (refer to section 2.5 for
proper screws and cables characteristics).
FAST-Bi-1K5 – User's Manual
5 ��
��
=
��������−����������
20 ����/°��
��
��������−����������
��
=
������
16
2
− 1
= 250 °��
≅ 0.0038 °��