GENERAL INFORMATION
_
Introduction
This Operating and Service Manual contains
information about initial inspection, performance
tests, adjustments, operation, troubleshooting and
repair of the HP 84811A Peak Power Sensor.
On the title page of this manual is a "Microfiche"
part number. This number can be used to order a 4
x 6-inch microfilm transparency of the manual.
Specifications
Instrument specifications are listed in Table 1.
These specifications are the performance stand-
ards or limits against which the instrument is
tested.
Table 1. Specifications"
Power range: 0 to +20 dBm (1 mW to 100 mW).
Frequency range: 100 MHz to 18 GHz.
SWR: 100 MHz to 12 GHz «1.5,
12 GHz to 18 GHz «2.0.
.
Maximum Peak Power: +24 dBm (250 mW) for 5 min.
Connactor type: N(male).
Calibration: every 2 GHz from 2 to 10 GHz,
every 1 GHz from 11 to 18 GHz.
Operating temperature: 0 to +55°C.
Calibration accuracy: (+10 to +40°C).
+0.7 dB 0.1 to 12 GHz,
=1.0 dB 12 to 18 GHz.
General:
Dimensions: 30 mm H x 38 mm Wx 140 mm L
(1.2x 1.5 x 5.5 inches).
Weight: 0.5 kg, (1 Ib.) net.
Sensor cabla length: 1.2 metres (4.1 ft).
"Specifications only apply in combination with specifications of
HP 8900C/D peak power meters.
instruments Covered by Manual
This instrument has a two-part serial number. The
first four digits and the letter are the serial number
prefix. The last five digits form a sequential suffix
which is unique to each instrument. The contents
of this manual apply directly to instruments hav-
ing the serial number prefix listed under SERIAL
NUMBERS on the title page.
An instrument manufactured after the printing of
this manual may have a serial number prefix that
is not listed on the title page. This unlisted serial
number prefix indicates that the manual for this
HP 84811A
instrument is supplied with a Manual Changes sup-
plement that documents the differences.
In addition to change information, the supplement
may contain information for correcting errors in the
manual. The supplement for this manual is keyed
to this manual's print dateand part number, both of
which appear on the title page.
For information concerning a serial number prefix
notlisted on thetitle pageorin the Manual Changes
supplement, contact your nearest Hewlett-Packard
office.
Description
The HP
84811A
is a Peak Power Sensor. With
a compatible peak power meter it measures power
levels in a range from 0 to +20 dBm (1 mW to 100
mW), and at frequencies from 100 MHz to 18 GHz.
Compatible meters are the HP 8900C and HP 8900D
Peak Power Meters.
CAUTION
The HP 84811A Peak Power Sensor has a
maximum peak power rating of +24dBm
(250 mW). Maintaining an input power
devel of 920 mW far langer than five (5)
minutes or exceeding the maximum peak
power rating may cause severe compo-
nent damage.
NOTE
The diode (UI) is easily replaceable.
Hewlett-Packard recommends
that re-
placement diodes be kept as spares if the
HP 84811A is used in applications where
inadvertent high power overload might
occur.
For instructions on disassembly and ያድ
assembly of the HP 84811A, refer to the
service section of this manual.
Recommended Test Equipment
Table 2 lists the test equipment recommended to
check, adjust, and troubleshoot the Peak Power
Sensor. If substitute equipmentis used, it must meet
or exceed the critical specifications.