Kistler 9286B Series Instruction Manual

Portable multicomponent force plate for biomechanics
Table of Contents

Advertisement

Quick Links

Instruction
Manual
Portable
Multicomponent
Force Plate
for Biomechanics
Type 9286B...
ä
9286B_002-464e-09.08

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 9286B Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Kistler 9286B Series

  • Page 1 Instruction Manual Portable Multicomponent Force Plate for Biomechanics Type 9286B… ä 9286B_002-464e-09.08...
  • Page 2 Instruction Manual Portable Multicom- Portable ponent Force Plate Multicomponent for Biomechanics Force Plate Type 9286B... for Biomechanics Type 9286B… ä 9286B_002-464e-09.08...
  • Page 4 Foreword Foreword We thank you for choosing a Kistler quality product distin- guished by technical innovation, precision and long life. Information in this document is subject to change without notice. Kistler reserves the right to change or improve its products and make changes in the content without obliga- tion to notify any person or organization of such changes or improvements.
  • Page 5: Table Of Contents

    Portable Multicomponent Force Plate for Biomechanics Type 9286B... Content Introduction ........................... 4 Important Information........................5 For your Safety........................5 ä-Conformity........................6 How to Treat the Instrument....................6 With External Amplifier (Type 9286B) .................. 6 Hints for Using the Operating Instructions ................7 What Happens After Modifications? ..................
  • Page 6 8.3.1 General ..........................33 8.3.2 Mechanical and Environment....................34 8.3.3 Electrical..........................34 8.3.4 Built-In 8 Channel Charge Amplifier..................35 Frequency Response ......................36 Glossary............................37 Coordinate System......................37 Kistler Coordinate System ....................37 9.2.1 ISB Coordinate System ......................38 Parameters Calculation......................38 Warranty............................39 Declaration of Conformity ......................40 Total Pages 41 9286B_002-464e-09.08...
  • Page 7: Introduction

    Please take the time to thoroughly read this instruction manual. It will help you with the installation, maintenance, and use of this product. Kistler offers a wide range of products for use in measuring technology: Piezoelectric sensors for measuring force, torque, strain,...
  • Page 8: Important Information

    Important Information Important Information Please keep to the following rules without fail. This will en- sure your personal safety when working, and assure long, trouble-free performance by the instrument. For your Safety This instrument has been tested thoroughly and it left the works in a perfectly safe condition.
  • Page 9: Ä-Conformity

    Portable Multicomponent Force Plate for Biomechanics Type 9286B... ä-Conformity The Force Plate Type 9286B… is conforming to ä direc- tives according to the Declaration of Conformity enclosed with this operating instructions. How to Treat the Instrument The force plate may be used only under the specified environmental and operating conditions Protect the signal output against dirt and do not touch it with your fingers.
  • Page 10: Hints For Using The Operating Instructions

    Please keep these Operating Instructions in a safe place where they can be consulted any time. If the instructions get lost, just turn to your Kistler distribu- tor and they will be replaced without delay. All information and directives in these instructions may be modified at any time without prior notification.
  • Page 11: General Description Of The Instrument

    Portable Multicomponent Force Plate for Biomechanics Type 9286B... General Description of the Instrument What does a Multicomponent Force Plate Do? The multicomponent force plate provides dynamic and quasistatic measurement of the 3 orthogonal components of a force (F ) acting from any direction on the top plate.
  • Page 12: Two Different Versions

    General Description of the Instrument Two Different Versions The force plate Type 9286B... is available in two versions: Type 9286B with external 8 channel amplifier (e.g. Type 9865…) Type 9286BA with built-in amplifier Application Examples The force plate is particularly suitable to measure slow and medium actions where center of pressure (COP) accuracy is important or if portability is requested Biomechanical investigations, e.g.
  • Page 13 Portable Multicomponent Force Plate for Biomechanics Type 9286B... In these four force sensors the force introduced is resolved into three components. Fig. 2: 3-component force sensor The electrical charges yielded by the force plate are strictly proportional to the measured quantities. They are con- verted by charge amplifiers into analog voltages and can then be processed in any way.
  • Page 14: Design Of The Force Plate

    General Description of the Instrument Design of the Force Plate The force plate comprises an aluminium top plate under which the four 3-component force sensors are mounted under high prestress. This prestress is a prerequisite for transmissing shear forces by friction. Fig.
  • Page 15: Assembly, Installation And Putting Into Operation

    Please refer to the instruction Manual 002-034 "Force Pla- te Installation and Maintenance" for detailed procedures and information. For nonobservance of the mounting instructions Kistler will decline all responsibility. Page 12 9286B_002-464e-09.08...
  • Page 16: Choice Of Location

    Assembly, Installation and Putting Into Operation Choice of Location The portable multicomponent force plate Type 9286B… needs no foundation and can therefore be set up any- where. Nevertheless the location must satisfy certain re- quirements listed below, if exact and reproducible meas- urements are to be obtained.
  • Page 17: Placing The Force Plate

    Portable Multicomponent Force Plate for Biomechanics Type 9286B... Placing the Force Plate Thoroughly clean the floor on which the force plate is to stand, and allow it to dry Carefully place the force plate on the floor and bring it into position Check whether the force plate rocks by pressing strongly and alternately on diagonally opposite corners...
  • Page 18: Floor Recessing

    Assembly, Installation and Putting Into Operation 4.3.1 Floor Recessing If the force plate is to be used for gait analysis it must be recessed into the floor and provided with the same floor covering as its surroundings. This will prevent the test per- son concentrating too much on the plate so that his gait is altered.
  • Page 19: Basic Circuitry And Cabling Of The Measuring System

    The following rules should be observed when cabling the measuring rig: Between force plate and charge amplifier the special cables supplied by Kistler must be used Ordinary cables may be used to link the charge amplifi- ers with the display or evaluation instruments Make sure that all work with electrical connections is done carefully and with cleanliness.
  • Page 20: Connections With External Amplifier (Type 9286B)

    Assembly, Installation and Putting Into Operation Connections with External Amplifier (Type 9286B) The following figure shows the elements required for a complete measuring system. Fig. 7: Measuring system with BioWare ® and two force plates Type 9286B Instead of the BioWare software system also other process- ing and indicating instruments can be connected to the charge amplifier.
  • Page 21: Connections With Built-In Amplifier (Type 9286Ba)

    Portable Multicomponent Force Plate for Biomechanics Type 9286B... It is very important that the correct type of cable be used only. If the contacts or insulating material are contaminated, they can be cleaned with cleaning spray (Type 1003) or clean white gasoline (rectified benzene). The insulation resistance 10 TΩ...
  • Page 22: Output Connector Of Force Plate Type 9286Ba (With Built-In Amplifier)

    Assembly, Installation and Putting Into Operation 4.6.1 Output Connector of Force Plate Type 9286BA (with Built-In Amplifier) Exct. GND Operate Control GND N.C. (reserved) x1+2 10 Fx3+4 11 F y1+4 12 F y2+3 13 A' (Range z select) 14 Signal GN 15 N.C.
  • Page 23: Cable Type 1758

    1758... (straight Fischer connector) is available from the force plate Fischer 19 pin connector to a D-Sub 37 pin neg. connector. It provides the following output connector and is compatible with the Kistler BioWare hardware. Signal GND Signal GND Signal GND...
  • Page 24: Digital Control

    Assembly, Installation and Putting Into Operation 4.6.3 Digital Control All Digital Inputs are isolated with optocouplers: Fig. 13: Digital Control input circuit Logical Signal Dig. Input <4,4 V 5 ... 45 V Table 1: Digital Control input signals Range Select Group I F (Shear) Range Select Group II F (Vertical)
  • Page 25: Operation

    Portable Multicomponent Force Plate for Biomechanics Type 9286B... Operation Range Selection Below are range recommendations for different applica- tions. It is important that for every application the smallest possible range is chosen. This will take full advantage of the maximum output voltage thus keeping noise and other errors the smallest possible.
  • Page 26: Balance (Posturography)

    Operation 5.2.1 Balance (Posturography) It is imperative the charge amplifier is switched to OPER- ATE before the subject Stepps on the force plate. Other- wise Center of Pressure COP can not be calculated cor- rectly. Time Constant Some charge amplifiers have a time constant (TC) function. A time constant is a high pass filter (AC mode) which al- lows dynamic measurement without the static (DC) com- ponent.
  • Page 27: Temperature Influences

    Portable Multicomponent Force Plate for Biomechanics Type 9286B... Temperature Influences The temperature coefficient of the sensitivity is negligibly small (sensitivity change with temperature, the entire force plate being at the same temperature). Temperature gradients on the other hand give rise to an error signal in the form of drift.
  • Page 28: Maintenance

    Maintenance 6. Maintenance Please refer to the instruction manual 002-034 "Force Pla- te Installation and Maintenance" for detailed procedures and information. The multicomponent force plate Type 9286B... is a de- pendable instrument requiring virtually no maintenance. To ensure long and reliable operation of the force plate, please read the next two sections.
  • Page 29: Drift Check

    Zero offset Drift Correct "amplification" ratio Charge [pC]/Voltage [V] This task needs special equipment and Kistler trained service personnel. Zero offset and drift tests can be done using the "voltme- ter" function of BioWare or any other software or voltme- ter directly at the analog output pins.
  • Page 30: Calibration Of The Force Plate

    Do not put a damaged instrument into operation! Leave all repairs (e.g. replacement of damaged plug and socket connections or defective sensors) to Kistler trained service personnel 6.5.1 With External Amplifier (Type 9286B) Keep the connector meticulously clean. If necessary...
  • Page 31: Cleaning And Sterilization

    Portable Multicomponent Force Plate for Biomechanics Type 9286B... Cleaning and Sterilization In principle, because of the usual high temperatures sterili- zation of the force plate is not permissible. Instead of it we recommend the application of a mild, water-soluble clean- sing or disinfecting agent (clinically tested) with a pH-value 6 ...
  • Page 32: Trouble Shooting

    Defective Force Plate Proceed as follows if your force plate is defective: Contact your Kistler distributor and advise him that you are sending in the defective instrument Send the defective force plate in its original packing to your Kistler distributor...
  • Page 33: Technical Data

    Portable Multicomponent Force Plate for Biomechanics Type 9286B... Technical Data Dimensions and Connection Fig. 14: Dimensions and connection of the multicompo- nent force plate Type 9286B… Page 30 9286B_002-464e-09.08...
  • Page 34: Specifications For Type 9286B

    Technical Data Specifications for Type 9286B 8.2.1 General Range –2,5 ... 2,5 0 ... 10 Overload –3 … 3 0 … 12 Calibrated range –2,5 ... 2,5 0 ... 10 Calibrated partial range 0 ... 0,25 0 ... 1 Threshold <0,05 ≈–7,8 Sensitivity (nominal)
  • Page 35: Mechanical And Environment

    Portable Multicomponent Force Plate for Biomechanics Type 9286B... 8.2.2 Mechanical and Environment Distance of force plate surface from x-, y-plane –22 Distance of sensor axis from y-axis Distance of sensor axis from x-axis Weight (without mounting frame) Material: Aluminium DIN 3.3535 Top plate Degree of protection (DIN 40050) IP63...
  • Page 36: Specifications For Type 9286Ba

    Technical Data Specifications for Type 9286BA 8.3.1 General Nominal Range 1 (per channel) N/5V ≈125 N/5V ≈250 Nominal Range 1 (summed) ≈250 ≈1 000 Ratio Range 1 : 2 : 3 : 4 1 : 5 : 10 : 20 Range switching accuracy <±0,5 Mechanical overload...
  • Page 37: Mechanical And Environment

    Portable Multicomponent Force Plate for Biomechanics Type 9286B... 8.3.2 Mechanical and Environment –22 Distance of force plate surface from x-, y-plane Distance of sensor axis from y-axis Distance of sensor axis from x-axis Weight (without mounting frame) Material: Aluminium DIN 3.3535 Top plate Degree of protection (DIN 40050) IP52...
  • Page 38: Built-In 8 Channel Charge Amplifier

    Technical Data 8.3.4 Built-In 8 Channel Charge Amplifier 9286B_002-464e-09.08 Page 35...
  • Page 39: Frequency Response

    Portable Multicomponent Force Plate for Biomechanics Type 9286B... Frequency Response Force plate loaded with 700 N Page 36 9286B_002-464e-09.08...
  • Page 40: Glossary

    There are two different coordinate systems in use. In these operating instructions, the Kistler coordinate sys- tem is used exclusively. Kistler Coordinate System The following coordinate system is used for all Kistler force plates. Fig. 15: Kistler coordinate system 9286B_002-464e-09.08...
  • Page 41: Isb Coordinate System

    Portable Multicomponent Force Plate for Biomechanics Type 9286B... 9.2.1 ISB Coordinate System The ISB (International Society of Biomechanics) proposes the following coordinate system. Fig. 16: ISB coordinate system In the ISB-system not the active (acting) but counteracting (reacting) forces are described. The ISB coordinate system is used above all in connection with comprehensive move- ment analyses.
  • Page 42: Warranty

    Warranty 10. Warranty Regarding the warranty reference is made to the agree- ment between the respective contracting parties. 9286B_002-464e-09.08 Page 39...
  • Page 43: Declaration Of Conformity

    Portable Multicomponent Force Plate for Biomechanics Type 9286B... 11. Declaration of Conformity EG-Konformitätserklärung EC Declaration of conformity / Déclaration de conformité CE Hersteller Kistler Instrumente AG Manufacturer / Fabricant CH-8408 Winterthur erklärt, dass das Produkt declares that the product / déclare que le produit...
  • Page 44 Declaration of Conformity With External Amplifier Force Plate Cable Amplifier Data Acquisition Type 9281E Type 1686A... Type 9865E... Type 5691A1 with cable Type 1769A Type 9285 Type 1685B... Type 9286B Type 1685B... Type 2812A… Type 9287C Type 1686A... With Built-In Amplifier Force Plate Cable Amplifier...