Operating Details; Detection Methods - UNI-T UT336A User Manual

Refrigerant leak detector
Hide thumbs Also See for UT336A:
Table of Contents

Advertisement

UT336A User Manual

6.2 Operating Details

a)
Power-on Self-detection: Long press POWER button
by level and then off (meanwhile with Di sound), followed by Yellow/Orange/Red light on
then off. Reset as per the current gas condition, with regular Di sound in device, backlight
of POWER button in green, finishing the
normal detection.
b) Detection: Hold the device to move probe to find the leak location in the pipeline and
interface of refrigerating system. When the refrigerant is detected, there will be with Di
sound increasing in frequency, and alarm lights in Yellow/Orange/Red are on level by level
as per the leak concentration.
c) Sensitivity Adjustment: The default sensitivity of device is level 3, and short press
Sensitivity Increase button
needed, six levels of green lights will also be adjusted as per the different sensitivity.
d) Reset: Short press RESET button
alarming when high leak concentration is detected, and the reset green light will be off after
2s on during the reset. Short press the RESET button to set zero if it is alarming before the
leak location is detected until no alarm (sound/light) shown to go on the detection process.
e) Mute: Short press MUTE button
f)
Power Off: Long press POWER button
level after on to shutoff the device.

6.3 Detection Methods

a) Visually inspect the
leak on the connector, valve, maintenance port, copper welding spot or pipeline.
b)
Move the probe to carefully check every possible area, along the pipeline to check and
avoid any potential leak missed. Mark it down and go on the detection when any leak
area is detected.
c) The speed of moving probe should be ≤
distance should be kept in 1-3mm. See followings:
d) ①Alarming of device identifies the approaching leak location, repeatedly detect the
surroundings to check if any repeat alarm occurred.
② When the leak location is ensured, move the probe from different directions of non-alarm
area to the alarm area to locate the leak source.
③ Move the device away from the leak area, then reset the device and gradually decrease
power-on self-detection procedure
or Sensitivity
to set the current gas concentration to zero, it will be
to turn on/off the buzzer as needed.
and six levels of green lights will be off level by
refrigerating system to check if any oil and dust on the pipeline, any
, and six green lights are on level
button
to adjust the sensitivity as
decrease
1cm/s in the detection site, and the probe
to go on the
7