PRODUCT OPERATION MANUAL
TE100FC/TE300FC/TE600FC
//Thank you for choosing the peristaltic pump, please read this manual carefully
before use to ensure correct operation, use effect and product maintenance.//
V1.0

Advertisement

Table of Contents
loading

Summary of Contents for TE TE100FC

  • Page 1 PRODUCT OPERATION MANUAL TE100FC/TE300FC/TE600FC //Thank you for choosing the peristaltic pump, please read this manual carefully before use to ensure correct operation, use effect and product maintenance.// V1.0...
  • Page 2 Important information Please read this operation manual carefully before using the product! The contents of this operation manual are for reference only. The company reserves the right to change the product (design or specification) without prior notice. If you need the latest version of this manual, please contact the company or an authorized distributor.
  • Page 3: Table Of Contents

    QUICK INSTALLATION OPERATION Pump head installation diagram Pump tube installation diagram Power supply wiring diagram Operation panel and display DB-15 The external control interface sketch External control input wiring diagram External control output wiring diagram Foot switch operating instructions Replace the fuse Replace the pump head coupling ABLE OF CONTENTS PART1...
  • Page 4 3.3Product structure ·Drive ·Pump head/tube selection and reference flow 3.4 Technical parameter PART4 Product installation 4.1 Pump head/pump tube installation 4.2Installation suggestions and precautions PART5 Product operation 5.1 Line connection ·Power supply wiring diagram ·External control wiring diagram 5.2 Power-on ·Power-on inspection ·The default factory settings for the first boot 5.3 Operation panel and display...
  • Page 5 ·External control input -External control input interface definition -External control input line color function definition ·External control output -External control output interface definition -External control output line color function definition ·Foot switch operating instructions Accessories PART6 Troubleshooting and maintenance 6.1 Troubleshooting 6.2 Product maintenance ·Replace the fuse ·Replace the pump head coupling...
  • Page 6: Part1

    >>> INTRODUCTION PART ONE PART1 Matters needing attention >> Matters needing attention Please read the operating instruction manual carefully before operating this equipment. Safety: The staff responsible for the installation or maintenance of this equipment should have the experience and ability to carry out related work. This product is not applicable to the ATEX explosion-proof directive and cannot be used in flammable and explosive environments.
  • Page 7: Part2 Unboxing

    >>> INTRODUCTION PART TWO PART2 Unboxing >> Unboxing 2.1 Unpacking inspection Confirm that the pump is packaged in good condition. Please check the packing list, when unpacking, check the product model and the number of accessories, and check whether the parts are damaged during transportation. If you have any questions, please contact us immediately.
  • Page 8: Product Features

    >>> INTRODUCTION PART THREE 3.2 Product features *Touch screen operation: 3.5-inch color LCD display in Chinese and English, intelligent operation; *Data storage: 8 groups of filling data can be stored for users; *Suckback function: can effectively prevent liquid from dripping and ensure transmission accuracy;...
  • Page 9: Pump Head/Tube Selection And Reference Flow

    PART THREE ·Pump head/tube selection and reference flow 【Digital speed type peristaltic pump-this series can match the pump head 】 Flow unit: ( ml/min) Tubemode( l ID*Wallthickness) mm Model Max Speed (0.8*1.6)(1.6*1.6)(2.4*1.6)(3.1*1.6)(4.8*1.6)(6.4*1.6)(7.9*1.6) 100rpm TE100FC 300rpm TE300FC 1140 YZ1515x 600rpm TE600FC 1020...
  • Page 10: Technical Parameter

    >>> INTRODUCTION PART THREE 3.4 Technical parameter Drive type TE100FC TE300FC TE600FC Max speed 100rpm(Reversible) 300rpm(Reversible) 600rpm(Reversible) Speed resolution 0.1rpm 0.1rpm 0.1rpm Max flow 480ml/min 1440ml/min 2280ml/min Display mode LCD(EN and CN) Speed control mode Touch screen + keypad AC220V±10% (Standard)or AC110V±10%(optional)...
  • Page 11: Pump Head Installation Diagram

    >>> INTRODUCTION PART FOUR PART4 Product installation >> Product installation 4.1 Pump head/pump tube installation Before performing any loading, unloading or maintenance activities, be sure to disconnect the pump from the main power supply. Pump head installation diagram: YZ1515x(YZ2515x) TX315(TX325) Turn clockwise to buckle on the connecting plate DG series pump head...
  • Page 12: Pump Tube Installation Diagram

    >>> INTRODUCTION PART FOUR Pump tube installation diagram: YZ1515x(YZ2515x) TX315(TX325) DG series pump head...
  • Page 13: Installation Suggestions And Precautions

    >>> INTRODUCTION PART FOUR 4.2 Installation suggestions and precautions Installation suggestions >> Application accessories such as foot switch, countersunk head, check valve, filling nozzle, connector, etc. can be selected according to actual conditions. >> For the size and selection of the tube, please refer to 3.3 Product Structure·Pump head/tube selection and reference flow Related Content.
  • Page 14: Power Supply Wiring Diagram

    >>> INTRODUCTION PART FIVE ·When transferring viscous liquids, it is necessary to maintain a low speed operation to improve the filling efficiency. It is recommended to connect a flexible tube no less than 1 meter between the inlet and the outlet to reduce the pulse and reduce the pulse loss. ·Try to put the pump at the same level or a lower level of the liquid to be transferred to improve the transfer efficiency of the pump.
  • Page 15: Operation Panel And Display

    >>> INTRODUCTION PART FIVE 5.2 Power-on ·Power-on inspection ①Check whether the pump pipe has been installed correctly, and whether the tube inlet pipe and outlet pipe have been correctly connected. ②Check whether it is connected to a matching power supply. ③Check whether the peristaltic pump has been installed according to "4.2 Installation suggestions and precautions".
  • Page 16 >>> INTRODUCTION PART FIVE (2) Graphic menu area data lock voice remote ①Set:Function operation main menu, peristaltic pump can be set accordingly. ②Data : Save new data quickly and apply saved data. ③Lock: Lock screen, unable to touch screen operation Unlock, touch screen operation.
  • Page 17: Menu Function Operation

    >>> INTRODUCTION PART FIVE Note: After each start, until it stops automatically, it is a complete filling cycle; manual stop will end immediately. ②Transmission mode · Pump Head Model · Speed ( rpm) · Timing ( sec) · Pump Tubing Midel · Flow(μL/m) (mL/m)(L/m)· Flow volum( e μL/mL/L) Note: Each start and stop is a complete timing and Flow volume display.
  • Page 18: System Setting Instructions

    >>> INTRODUCTION PART FIVE ①Main interface → click the filling time value, display the keypad → input the value (as shown in the figure) ②C-Clear Exit-Exit without saving Confirm-Save the value and exit ③The input range of filling time: 0.1s-9999m, this value is the pump limit ④Note: Please set the liter/ml on the main interface, the pump does not support the setting when it is running...
  • Page 19 >>> INTRODUCTION PART FIVE Head Select the current use of the pump head models. →General→Click Head→You can roll the rolling bar to view all the Main interface→Click pump head information→Click to select the pump head model→Click to Enter storage or click Cancel and not save it.
  • Page 20 Suck Back Select Suck-back parameters →system→Click Suck Back→Small keyboard input suck Main interface→Click < < TE100FC: ( 10-100rpm) back angle (10-720deg)/Small keyboard input suck back speed TE300FC: ( 10-300rpm) TE600FC: ( 10-300rpm) →Click Enter, save and exit;Click Cancel, exit without saving.
  • Page 21: Calibration Guide

    >>> INTRODUCTION PART FIVE Language Choose Chinese/English →General→Click Language→Chinese/EnglishClick Main interface→Click Enter, save and exit;Click Cancel, exit without saving. ( Note: language customization is available) *After completing the required system settings, return to the previous menu or exit to the main interface to perform other operations. ·Calibration Guide This series peristaltic pumps have the function of flow calibration, which can achieve high-precision dispensing.
  • Page 22: Mode Selection Instructions

    >>> INTRODUCTION PART FIVE 60.0s Countdown interface Calibration Data interface ·Mode selection instructions This series peristaltic pump has two working modes:Filling mode、Transmission mode →Click General→Select Filling/Transmission(Or *Main interface→Click click Exit to return to the previous menu) Filling mode main interface Transmission mode main interfac ·Data operation instructions This series peristaltic pump can realize user data storage function: save 8 groups of filling data.
  • Page 23: The External Control Interface Sketch

    >>> INTRODUCTION PART FIVE 1 is a set of saved filling data: YZ1515x 7.0mL 12.0s 6.0s - Head: The currently set pump head model - Tube: The currently set tube model model - Vol: Filling Volume display - Copy: Filling Number display - Time: Filling Time display - Interval: Pause time display ①...
  • Page 24: External Control Input Wiring Diagram

    >>> INTRODUCTION PART FIVE Drive external control interface (DB-15 description) 1、The using method of the external interface (A) Start/Stop wire and Ground wire connect or shut, control the start and stop of the pump. (B) Direction wire and Ground wire connect or shut, control the running direction of the pump.
  • Page 25: External Control Input Line Color Function Definition

    >>> INTRODUCTION PART FIVE 0-5V Direction(on-off) 485B 485A 0-10kHz 4-20mA Start/stop(on-off) 0-10V 1、 3、 6、 10、 13 are all E-c Ground External control input wiring diagram 【External control input line color function definition】 Serial Function Corresponding function of wire number 485 communication brown---485A blue---485B...
  • Page 26: External Control Output Wiring Diagram

    >>> INTRODUCTION PART FIVE ·External control output -External control output interface definition: 4-20mA Enable Start-stop Direction DEFINITION output output B output B output B Ground Ground Ground negative 4-20mA Enable Start-stop Direction 0-10kHz 0-10V 0-5V output output A output A output A output output...
  • Page 27: Foot Switch Operating Instructions

    >>> INTRODUCTION PART FIVE 【External control output line color function definition】 Serial Corresponding function of wire Function number start-stop output brown---Start-stop output A blue---Start-stop output B brown---4-20mA output positive 4-20mA output blue---4-20mA output negative brown---0-5V/0-10V/0-10kHz output 0-5V/0-10V/0-10kHz output blue---E-c Ground brown---Start-stop output A blue---Start-stop output B start-stop、...
  • Page 28: Accessories

    >>> INTRODUCTION PART FIVE *Accessories The following accessories are products selected by customers. If you want to know more or need to buy, please contact the company. Image Accessories Features Flat mouth Filling nozzle Closing-type Connecting tube, precise filling, splash proof Filling nozzle Foot switch Control peristaltic pump start and stop...
  • Page 29: Part6 Troubleshooting And Maintenance

    >>> INTRODUCTION PART SIX PART6 Troubleshooting and maintenance >> Troubleshooting and maintenance Note: There are no parts in the pump that can be repaired by the user. If you need repairs, please contact the dealer or the company! 6.1 Troubleshooting >>If a circuit protection device is installed, confirm that the circuit has not tripped >>Confirm that the power plug is inserted into a...
  • Page 30: Replace The Fuse

    >>> INTRODUCTION PART SIX 6.2 Product maintenance Warning: Before attempting any maintenance, be sure to cut off the power to the pump. ·Replace the fuse ①Place the power switch in the "off" position(“|”On,“○”Off) ②Disconnect the AC power input cord from the outlet. ③Take out the spare built-in fuse from the power socket of the pump.
  • Page 31: Warranty Information

    >>> INTRODUCTION PART SIX ④The pump head roller does not need to add lubricating oil, and improper operation may cause the tube to shift or corrode. ⑤Not used to deal with chemical substances incompatible with the pump head or tube. ⑥...
  • Page 32: Relevant Details

    >>> INTRODUCTION PART SIX ·Relevant details ① The company will not be responsible for any direct or indirect losses caused by external reasons such as operating errors or human negligence that are not the product's own quality problems. Direct loss: products, supporting machinery, working environment, surrounding buildings, etc.
  • Page 33: Disclaimer

    >>> INTRODUCTION PART SIX ③When applying for after-sales service, please provide a clear reason for the return, specific information about the contacted substances, and responsible for the cleaning of the product (especially products that have been exposed to toxic chemical substances or harmful substances to the human body) ④...
  • Page 34: Appendix 1:1 Comparison Table Of Tube Size

    >>> INTRODUCTION APPENDIX Appendix 1:1 comparison table of tube size >> 1:1 comparison table of tube size Micro flow tube 0.5x0.8 1 x 1 2 x 1 2.4x0.8 3.2x0.8 Tube Tube section (1:1) Wall thickness(mm) 0 . 8 Inside diameter(mm) 2 .
  • Page 35: Industrial Tube

    >>> INTRODUCTION APPENDIX Industrial tube Tube 7 3 # 8 2 # Tube section (1:1) Wall thickness(mm) Inside diameter(mm) 9 . 6 1 2 . 7 Continuous 0 .1 7 0 . 1 Pressure (Mpa) Interval 0 .2 7 0 . 1 8 6 # 9 0 # Tube...
  • Page 36 TE L AB ORATORY TE CH NOLOGY C O.,LTD Address: 5F., No. 510, Zhongzheng Rd., Xindian Dist., New Taipei City 231 , Taiwan Tel.: 02-7746-3341 / 02-2874-4741 E-mail: sales@te-lab-equipment.com.tw Website: https://www.te-lab-equipment.com.tw/...