  • Page 1 ™ SCXI Chassis User Manual SCXI-1000, SCXI-1000DC, SCXI-1001, and SCXI-2000 Chassis SCXI Chassis User Manual November 1996 Edition Part Number 320423D-01 © Copyright 1991, 1996 National Instruments Corporation. All Rights Reserved.
  • Page 2 Mexico 5 520 2635, Netherlands 0348 433466, Norway 32 84 84 00, Singapore 2265886, Spain 91 640 0085, Sweden 08 730 49 70, Switzerland 056 200 51 51, Taiwan 02 377 1200, U.K. 01635 523545 National Instruments Corporate Headquarters 6504 Bridge Point Parkway Austin, TX 78730-5039 Tel: (512) 794-0100...
  • Page 3 Customer's right to recover damages caused by fault or negligence on the part of National Instruments shall be limited to the amount theretofore paid by the customer. National Instruments will not be liable for damages resulting from loss of data, profits, use of products, or incidental or consequential damages, even if advised of the possibility thereof.
  • Page 4 SCXI-1000/1001 ................2-12 SCXI-1000DC.................. 2-13 SCXI-2000 ..................2-15 Baud Rate Selection on the SCXI-2000 ......2-16 Voltage Selection and Fuse Replacement for SCXI-1000/1001/2000 ... 2-17 Voltage Selection Procedure............2-19 Power Entry Module Fuse Replacement Procedure ......2-19 © National Instruments Corporation...
  • Page 5 Table of Contents Backplane Fuse Replacement and Check Procedure for SCXI-1000, SCXI-1001, and SCXI-2000 ..........2-19 Fuse Replacement for the SCXI-1000DC............2-21 Power Entry Fuse and +5 VDC Fuse Replacement Procedure ..2-21 Backplane Fuse Replacement and Check Procedure....... 2-22 Chassis Installation.......................
  • Page 6: Table Of Contents

    SCXI-1000/1001/2000 Voltage Selection and Fuse Ratings by Region....................2-18 Table 2-7. SCXI-1000/1001/2000 Fuse Part Numbers ........... 2-18 Table 2-8. SCXI-1000/1001/2000 Backplane Fuse Part Numbers ......2-20 Table 2-9. SCXI-1000DC Rear Panel Fuse Part Numbers ........2-21 Table 2-10. SCXI-1000DC Backplane Fuse Part Numbers........2-22 Table 2-11.
  • Page 7 Appendix A, Specifications, lists the specifications for the SCXI chassis. • Appendix B, Customer Communication, contains forms you can use to request help from National Instruments or to comment on our products and manuals. • The Glossary contains an alphabetical list and description of terms used in this manual, including abbreviations, acronyms, metric prefixes, mnemonics, and symbols.
  • Page 8 SCXIbus <signal name> line (or signal). The SCXIbus descriptor may be omitted when the meaning is clear. SCXI chassis SCXI chassis or simply chassis refers to the SCXI-1000, SCXI-1000DC, SCXI-1001, and SCXI-2000. When information pertains to only one chassis, that chassis is named explicitly; for example, the SCXI-1001 has 12 module slots.
  • Page 9 Consult these guides when you are making your connections. • SCXI chassis manuals—Read these manuals for maintenance information on the chassis and installation instructions. © National Instruments Corporation SCXI Chassis User Manual...
  • Page 10 SCXI-1000, SCXI-1000DC, and SCXI-1001. It does not contain remote programming information for the SCXI-2000 unless you are using it in SCXI-1000 mode. This manual is available from National Instruments by request. If you are using NI-DAQ, LabVIEW, or LabWindows/CVI, you should not need the register-level programmer manual.
  • Page 11 SCXI chassis kit. About the SCXI Chassis The SCXI-1000, SCXI-1000DC, and SCXI-2000 are four-slot SCXI chassis and the SCXI-1001 is a 12-slot SCXI chassis. All the chassis house National Instruments SCXI modules. The SCXI-1000, SCXI-1001, and SCXI-2000 chassis are powered with standard AC power.
  • Page 12 NI-DAQ for PC compatibles Your computer Software Programming Choices There are several options to choose from when programming your National Instruments plug-in DAQ and SCXI hardware. You can use LabVIEW, LabWindows/CVI, NI-DAQ, or register-level programming. LabVIEW and LabWindows/CVI Application Software LabVIEW and LabWindows/CVI are innovative program development software packages for data acquisition and control applications.
  • Page 13: Figure 1-1. The Relationship Between The Programming Environment

    Chapter 1 Introduction NI-DAQ Driver Software The NI-DAQ driver software is included at no charge with all National Instruments DAQ hardware and the SCXI-2000 chassis. NI-DAQ has an extensive library of functions that you can call from your application programming environment. These functions include routines for analog...
  • Page 14 Chapter 1 Introduction Register-Level Programming The final option for programming any National Instruments DAQ hardware is to write register-level software. Writing register-level programming software can be very time-consuming and inefficient and is not recommended. Even if you are an experienced register-level programmer, consider using NI-DAQ, LabVIEW, or LabWindows/CVI to program your National Instruments DAQ hardware.
  • Page 15: Table 1-2. Scxi Cable Accessories

    DIO-32F 1 m shielded cable assembly SCXI-1349 Shielded cable assembly for MIO E Series SCXI-1350 Multichassis adapter SCXI-1351 One-slot cable extender SCXI-1353 Shielded cable assembly for 100-pin E Series SCXI-1354 Ribbon cable assembly for DAQCard-1200 © National Instruments Corporation SCXI Chassis User Manual...
  • Page 16: Table 1-3. Scxi Chassis Accessories

    Unpacking Remove the chassis from the package and inspect the chassis for loose components or any other sign of damage. Notify National Instruments if the chassis appears damaged in any way. Do not install a module into a damaged SCXI chassis.
  • Page 17: Table 2-1. Scxi Chassis Front View Items

    Brings power, control lines, and analog bus connections to modules Front-threaded strips Secure modules in the chassis and attaches front panels DIP switches Determine the chassis address (SCXI-1000, SCXI-1001, and SCXI-2000) and baud rate (SCXI-2000 only) 9-pin DSUB RS-232 connector (SCXI-2000 only)
  • Page 18: Figure 2-1. Scxi-1000 Front View Diagram

    Configuration and Installation 1 DIP Switches 3 Power Switch 5 Module Guides 7 Backplane 2 Reset Button 4 Indicator Light 6 Threaded Strips 8 Slot 0/Power Supply Figure 2-1. SCXI-1000 Front View Diagram SCXI Chassis User Manual © National Instruments Corporation...
  • Page 19: Figure 2-2. Scxi-1000Dc Front View Diagram

    1 Jumpers (behind front panel) 3 Power Switch 5 Module Guides 7 Backplane 2 Reset Button 4 Indicator Light 6 Threaded Strips 8 Slot 0/Power Supply Figure 2-2. SCXI-1000DC Front View Diagram © National Instruments Corporation SCXI Chassis User Manual...
  • Page 20: Figure 2-3. Scxi-1001 Front View Diagram

    Configuration and Installation 1 DIP Switches 3 Power Switch 5 Module Guides 7 Backplane 2 Reset Button 4 Indicator Light 6 Threaded Strips 8 Slot 0/Power Supply Figure 2-3. SCXI-1001 Front View Diagram SCXI Chassis User Manual © National Instruments Corporation...
  • Page 21: Figure 2-4. Scxi-2000 Front View

    3 Reset Button 6 Status LEDs Figure 2-4. SCXI-2000 Front View Tables 2-2, 2-3, and 2-4 describe the rear view items shown in Figures 2-5, 2-6, 2-7, and 2-8 on the following pages. © National Instruments Corporation SCXI Chassis User Manual...
  • Page 22: Figure 2-5. Scxi-1000 Rear View Diagram

    Chapter 2 Configuration and Installation Table 2-2. SCXI-1000/1001 Chassis Rear View Items Item Description Power entry module IEC receptacle for power input, voltage selection board, and fuse Fuse Protects both you and the chassis in case of a fault in the SCXI chassis...
  • Page 23: Figure 2-6. Scxi-1001 Rear View Diagram

    Protect the power supply from shorts on modules Rear-connector space For module space, connector mounting brackets, or adapter boards Rear-threaded strips Secure cable connections, mounting brackets, or filler panels to the chassis © National Instruments Corporation SCXI Chassis User Manual...
  • Page 24: Figure 2-7. Scxi-1000Dc Rear View Diagram

    7 +5 VDC Internal Power Supply Fuse 2 Fan and Filter 5 Power Entry Connector 8 Backplane Fuses (behind fan) 3 Rear-Threaded Strips 6 Power Input Fuse Figure 2-7. SCXI-1000DC Rear View Diagram SCXI Chassis User Manual © National Instruments Corporation...
  • Page 25: Table 2-4. Scxi-2000 Chassis Rear View Items

    Rear-threaded strips Secure cable connections, mounting brackets, or filler panels to the chassis Screw terminal Connects screw terminal block to make RS-485 connections connector Parallel port connector For cabling to SCXI-1200 module © National Instruments Corporation SCXI Chassis User Manual...
  • Page 26: Figure 2-8. Scxi-2000 Rear View Diagram

    2 Fans and Filters 6 Voltage Selection Wheel 9 Parallel Port Connector 3 Rear-Threaded Strips 7 Fuse (concealed) 10 Backplane Fuses (behind fan) 4 Rear-Connector Space Figure 2-8. SCXI-2000 Rear View Diagram SCXI Chassis User Manual 2-10 © National Instruments Corporation...
  • Page 27 To clean the chassis, brush off light dust with a soft, non-metallic brush. Remove other contaminants with deionized water and a stiff non-metallic brush. Make sure the chassis is completely dry and free from contaminants before returning it to service. © National Instruments Corporation 2-11 SCXI Chassis User Manual...
  • Page 28 0 and the switch setting for chassis address 19. Note: Older SCXI-1001 and SCXI-1000 chassis do not have switches; they use jumpers like those on the SCXI-1000DC. Very old SCXI-1000 chassis do not have switches or jumpers and respond to any address. (To determine if your chassis has no jumpers, look at the label on the back of your chassis.
  • Page 29: Figure 2-9. Address Setting Examples

    If the SCXI-1000DC part number has no letter or it has a B or C, it has no jumpers.) You cannot use jumperless chassis in a multichassis system. © National Instruments Corporation 2-13 SCXI Chassis User Manual...
  • Page 30: Figure 2-10. Scxi-1000Dc Chassis Address Jumper Settings

    4. Set all three jumpers for the chassis address you want. 5. Replace the front panel of Slot 0. 6. Replace the four cheesehead screws. Do not overtighten. 7. Replace the two panhead screws. Do not overtighten. SCXI Chassis User Manual 2-14 © National Instruments Corporation...
  • Page 31 DAQ device to control multiple chassis), you must specify the address of the chassis as if it were an SCXI-1000. Set the address on the chassis with switches 1 through 5 on the front panel of the SCXI-2000 to select an address between 0 and 31 (switches 6 through 8 are not used).
  • Page 32: Figure 2-11. Baud Rate And Address Setting Examples

    If you are not controlling your chassis via its serial port, set the four baud rate switches to the SCXI-1000 setting shown in Table 2-5. The chassis will operate like an SCXI-1000 chassis and the serial ports and parallel port will not function.
  • Page 33: Table 2-5. Scxi-2000 Baud Rate Settings

    Reserved Functions as SCXI-1000 Voltage Selection and Fuse Replacement for SCXI-1000/1001/2000 If you have ordered your chassis with the appropriate part number (the -0x extension of the kit part number corresponds to your geographical region), the voltage selection wheel and fuse will be correct for operation in your geographical region.
  • Page 34: Table 2-6. Scxi-1000/1001/2000 Voltage Selection And Fuse Ratings

    Caution: For continued protection against fire, replace fuses only with fuses of the same type and rating. Table 2-6. SCXI-1000/1001/2000 Voltage Selection and Fuse Ratings by Region Region Line Voltage Fuse Rating...
  • Page 35 SCXI-1001, and SCXI-2000 In addition to the power entry module fuse, the analog supply lines on the backplane are fused at 1.5 A on the SCXI-1000 and SCXI-2000 chassis and at 4 A on the SCXI-1001 chassis. The manufacturer part ©...
  • Page 36: Table

    On the SCXI-1000 and SCXI-2000, the backplane fuses are located behind the fan. On the SCXI-1001, the backplane fuses are located behind the rightmost fan, closest to the power entry module, as viewed from the rear of the chassis.
  • Page 37: Table 2-9. Scxi-1000Dc Rear Panel Fuse Part Numbers

    (F1) is a 6.3 A, 5 x 20 mm time lag fuse. The internal +5 VDC supply is fused by a 3.15 A, 5 x 20 mm time lag fuse (F2). Table 2-9 shows the manufacturer part numbers National Instruments uses for these fuses.
  • Page 38: Table 2-10. Scxi-1000Dc Backplane Fuse Part Numbers

    In addition to the power entry and the +5 V supply fuses, the analog supply lines on the backplane are fused at 1.5 A on the SCXI-1000DC chassis. The manufacturer part number National Instruments uses for this fuse is given in Table 2-10.
  • Page 39 Category II environment per IEC 644. This category requires local level supply mains-connected installation. SCXI-1000 and SCXI-1001 Chassis Installation Use the following procedure to install the SCXI-1000 and SCXI-1001 chassis: 1. If necessary, change the chassis address of the box by following the instructions in the Chassis Address Selection section earlier in this chapter.
  • Page 40 Caution: Connecting the power source with opposite polarity may cause permanent damage to the SCXI chassis. National Instruments is liable for any damage or injuries resulting from improper power connections. 6. Insert the power plug into the header J1.
  • Page 41 Some computers use a 25-pin connector and some NEC computers use a 14-pin connector. Note: You must use a null modem cable to connect your computer directly to your SCXI-2400. © National Instruments Corporation 2-25 SCXI Chassis User Manual...
  • Page 42: Table 2-11. Rs-232 Signal Connections

    RS-485 Cable You can use only an RS-232 or an RS-485 cable; you cannot use both at the same time. SCXI Chassis User Manual 2-26 © National Instruments Corporation...
  • Page 43: Figure 2-12. Rs-485 Connections For Two Scxi-2000 Chassis

    Figure 2-12 shows a diagram of a two-chassis network. Last First SCXI Chassis SCXI Chassis TX- RX- TX- RX- Computer RS-485 Port Figure 2-12. RS-485 Connections for Two SCXI-2000 Chassis © National Instruments Corporation 2-27 SCXI Chassis User Manual...
  • Page 44: Table 2-13. Scxi-2000 Front Panel Led Indicators

    If your switch settings are invalid, only the PROG and TEST LEDs will flash on and off together. • If the chassis is ready for normal operation and set for SCXI-1000 mode, the PROG, TEST, and ONLINE LEDs will flash in sequence.
  • Page 45: Table 2-14. Scxi-2000 Troubleshooting

    TEST, PROG, and Return the chassis for servicing. ONLINE flash simultaneously TEST, PROG, and If you are using SCXI-1000 mode, check the address switches, baud ONLINE flash in rate switches, and configuration in NI-DAQ Configuration Utility for sequence but nothing the chassis.
  • Page 46 Align the panel so that the serial number faces into the chassis and the screw holes are on the right. 2. Using a flathead screwdriver, screw the two screws through the rear panel into the threaded strip holes. SCXI Chassis User Manual 2-30 © National Instruments Corporation...
  • Page 47 6. Replace the fan filter by aligning the fan and filter with the fan holes, making sure that the label side of the fan is face down. Reinstall the four screws and make sure the assembly is secure. © National Instruments Corporation 2-31 SCXI Chassis User Manual...
  • Page 48 Max load 680 mA 2.04 A +5 V Tolerance limits include peaks +4.75 to +5.25 V +4.75 to +5.25 V Ripple (peak-to-peak) 50 mV 50 mV Max load 250 mA 600 mA © National Instruments Corporation SCXI Chassis User Manual...
  • Page 49 Max AC operating current at 120 VAC SCXI-1000........0.5 A SCXI-2000........0.5 A SCXI-1001........1.5 A SCXI-1000DC Input voltage ...........12 VDC nominal (9.5 to 16.0 VDC) Max DC operating current at 9.5 VDC ..........5.5 A SCXI Chassis User Manual © National Instruments Corporation...
  • Page 50 Appendix A Specifications Physical Weight SCXI-1000 ..........3.9 kg (8 lb 10 oz.) SCXI-1000DC........3.3 kg (7 lb 5 oz.) SCXI-1001 ..........6.8 kg (14 lb 14 oz.) SCXI-2000 ..........3.8 kg (8 lb 8 oz.) Refer to the following figures for the physical dimensions of the 4-slot chassis (SCXI-1000, SCXI-1000DC, and SCXI-2000) and the 12-slot chassis (SCXI-1001).
  • Page 51: Table A-1. Max Data Transfer Rates (Samples/S

    Byte format........One start bit, one stop bit, no parity, eight data bits Table A-1. Max Data Transfer Rates (Samples/s) Baud Rate Operating System Windows 95 Windows 3.1 57.6 kbps 1,500 1,500 38.4 kbps 1,000 1,000 SCXI Chassis User Manual © National Instruments Corporation...
  • Page 52 20,000 samples or less, the chassis can buffer data at 12,000 samples per second. Acquisitions that fit in the input FIFO of the SCXI-1200 can be run at the maximum specified rates of the SCXI-1200. © National Instruments Corporation SCXI Chassis User Manual...
  • Page 53 Electronic Services Bulletin Board Support National Instruments has BBS and FTP sites dedicated for 24-hour support with a collection of files and documents to answer most common customer questions. From these sites, you can also download the latest instrument drivers, updates, and example programs. For recorded instructions on how to use the bulletin board and FTP services and for BBS automated information, call (512) 795-6990.
  • Page 54 Fax and Telephone Support National Instruments has branch offices all over the world. Use the list below to find the technical support number for your country. If there is no National Instruments office in your country, contact the source from which you purchased your software to obtain support.
  • Page 55 National Instruments for technical support helps our applications engineers answer your questions more efficiently. If you are using any National Instruments hardware or software products related to this problem, include the configuration forms from their user manuals. Include additional pages if necessary.
  • Page 56 SCXI Chassis Hardware Configuration Form Record the settings and revisions of your hardware and software on the line to the right of each item. Also fill out the hardware and software configuration forms for all modules in the chassis, all relevant DAQ boards, and all other chassis in the application. By completing these forms accurately, our applications engineers will be able to answer your questions efficiently.
  • Page 57 Documentation Comment Form National Instruments encourages you to comment on the documentation supplied with our products. This information helps us provide quality products to meet your needs. Title: SCXI™ Chassis User Manual Edition Date: November 1996 Part Number: 320423D-01 Please comment on the completeness, clarity, and organization of the manual.
  • Page 58 +5 V +5 volt signal amperes alternating current analog to digital AIGND analog input ground AISENSE analog input sense signal Celsius clear to send signal © National Instruments Corporation SCXI Chassis User Manual...
  • Page 59 Slot 0 scan list feet gram gain ground high-speed CMOS TTL HDLC high-level data link control hexadecimal hertz inches International Electrotechnical Commission input/output light-emitting diode meters ONLINE online LED ounce PROG program LED SCXI Chassis User Manual © National Instruments Corporation...
  • Page 60 SCXI Signal Conditioning eXtensions for Instrumentation serial peripheral interface communication protocol TEST test LED transistor-transistor logic transmit LED transmit data signal USER user-defined signal volts volts, alternating current volts, direct current watts © National Instruments Corporation SCXI Chassis User Manual...
  • Page 61 2-11 cleaning instructions, 2-11 configuration, 2-12. See also installation. address selection address selection SCXI-1000 and SCXI-1001, jumper settings (figure) 2-12 SCXI-1000 and SCXI-1001, 2-12 figure, 2-13 figure, 2-13 SCXI-1000DC, 2-13 SCXI-1000DC, 2-13 figure, 2-14 figure, 2-14 SCXI-2000, 2-15...
  • Page 62 (table), A-1 to A-2 filler panel installation, 2-30 electronic technical support, B-1 to B-2 module installation, 2-30 environment specifications, A-6 SCXI-1000 and SCXI-1001, 2-23 equipment, optional, 1-4 to 1-6 SCXI-1000DC, 2-24 SCXI-2000, 2-24 indicator lights, 2-28 parallel port cable, 2-25...
  • Page 63 2-19 voltage selection and fuse ratings by installation, 2-23 region (table), 2-18 rear view diagrams, 2-6 rear view items (table), 2-6 source power requirements, A-2 voltage selection, 2-17 © National Instruments Corporation I -3 SCXI Chassis User Manual...
  • Page 64 SCXI 4-slot chassis (figure), A-4 1-1 to 1-2 SCXI-1001 (figure), A-5 optional equipment, 1-4 to 1-6 source power requirements overview, 1-1 SCXI-1000/1001/2000, A-2 software programming choices, 1-2 to 1-4 SCXI-1000DC, A-2 SCXI Chassis User Manual I -4 © National Instruments Corporation...
  • Page 65 B-1 to B-2 TEST LED, SCXI-2000, 2-28 transfer characteristics, A-6 TX LED, SCXI-2000, 2-28 voltage selection. See also fuse replacement. SCXI-1000, SCXI-1001, and SCXI-2000, 2-17 procedure, 2-17 voltage selection and fuse ratings by region (table), 2-18 weight specifications, A-3 ©...