Norican Simpson 42104 Operating Instructions Manual

Electronic universal sand strength machine
Table of Contents

Advertisement

Quick Links

Operating Instructions
Electronic Universal Sand Strength Machine
www.simpsongroup.com
Cold Tensile Strength Accessory
Splitting Strength Accessory
Green Deformation Accessory
Hot Shell Tensile Accessory
High Compression Strength Accessory
Core Transverse Strength Accessory
Shell Transverse Strength Accessory
Cold Shell Tensile Strength Accessory
Disk Transverse Accessory
MOR Bar Test Fixture Accessory
Model 42104
Accessories:
Model
42104C
42104D
42104E
24104F
42104H
42104K
42104L
42104N
42104P
211219

Advertisement

Table of Contents
loading

Summary of Contents for Norican Simpson 42104

  • Page 1 Operating Instructions Electronic Universal Sand Strength Machine Model 42104 Accessories: Model Cold Tensile Strength Accessory 42104C Splitting Strength Accessory 42104D Green Deformation Accessory 42104E Hot Shell Tensile Accessory 24104F High Compression Strength Accessory 42104H Core Transverse Strength Accessory 42104K Shell Transverse Strength Accessory 42104L Cold Shell Tensile Strength Accessory 42104N...
  • Page 2 Type: Electronic Universal Sand Strength Machine Model: 42104 Part No.: 0042104-ASM 0042104-M-ASM Serial Number: Name and address of manufacturer: Simpson Technologies 751 Shoreline Drive Aurora, IL 60504 For other Simpson Technologies offices around the world and for our contact information please visit us on the internet at www.simpsongroup.com on the Contacts page.
  • Page 3: Table Of Contents

    Table of Contents Table of Contents Introduction ..................1 Application and Designated Use .............. 1 Organizational Measures ................. 1 Safety ....................2 Safety Signs and Labels ................2 2.1.1 Safety Alert Symbols ..............3 2.1.2 Safety Symbol Labels ..............3 Lockout and Tagout System Procedure ...........
  • Page 4 Table of Contents 3.4.9 Disk Transverse Accessory (Model 42104P) ......14 3.4.10 MOR Bar Test Fixture Accessory ..........14 Unpacking and Installation ............... 15 Unpacking ....................15 Components ..................16 Installation ..................... 17 Electrical and Pneumatic Power Connection ......... 17 Connecting Power and Set-Up...............
  • Page 5 Table of Contents Green Deformation Accessory .............. 34 5.4.1 Description ................34 5.4.2 Operation .................. 35 High Compression Strength Accessory ..........36 5.5.1 Description ................37 5.5.2 Installation ................38 5.5.3 Operation .................. 38 5.5.4 Purging Air from High Compression Strength Accessory ..39 Core Transverse Accessory ..............
  • Page 6 Table of Contents 5.10 MOR Bar Test Fixture Accessory ............51 5.10.1 Description ................51 5.10.2 Installation ................52 5.10.3 Operation .................. 52 5.11 Error Messages ..................53 Maintenance and Calibration ............54 Maintenance ..................54 6.1.1 Daily Maintenance ..............54 6.1.2 Weekly Maintenance ..............
  • Page 7: Introduction 1

    Introduction 1 Introduction Congratulations, you have just purchased an extremely reliable sand testing instrument that is backed by the professional technical support and years of proven sand technology experience of Simpson Technologies . This laboratory equipment is constructed of quality materials and is the result of unsurpassed craftsmanship.
  • Page 8: Safety

    2 Safety Safety Before operating and/or performing maintenance or repair on Simpson Technologies designed and/or manufactured equipment, it is required that all personnel have read and understood the entire Operating Instructions manual. If any questions exist, you must contact your supervisor or Simpson Technologies before taking further action.
  • Page 9: Safety Alert Symbols

    Safety 2 2.1.1 Safety Alert Symbols This is the safety alert symbol. It is used to alert you to potential personal injury hazards. OBEY all safety messages that follow this symbol to avoid possible injury or death. DANGER! Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.
  • Page 10 2 Safety ELECTRICAL SHOCK / ELECTROCUTION (STC #217958) This label is located on the back panel above the electrical cord connecting inlet. With the back or bottom panels removed, the electrical power supply and electrical terminals are exposed. A hazardous voltage is present, can cause electric shock or burn and will result in serious injury.
  • Page 11 Safety 2 HAND CRUSH / FORCE FROM LEFT (STC #214047) This label is located on the right side of the cabinet by the Clamp Arm. When installing the Transverse Accessories with the power ON, it is possible to accidently press the Start button causing the Clamp Arm to travel and possibly pin your hand/fingers between the Transverse Accessory and the Stationary Clamp Holder, which may crush or cut body parts.
  • Page 12: Lockout And Tagout System Procedure

    2 Safety Lockout and Tagout System Procedure Whenever performing any type of maintenance or repair, whether in the form of cleaning, inspection, adjustment, mechanical or electrical maintenance, the equipment must be rendered into Zero Mechanical State (ZMS). Prior to any maintenance (routine or otherwise) or repair of equipment, a safety procedure should be established and maintained.
  • Page 13: Glossary

    Safety 2 2.2.2 Glossary: Authorized Person(s) - Personnel who have been designated by his/her department to perform maintenance or service on a piece(s) of equipment, machinery or system, and are qualified to perform the work through proper training on the Lockout/Tagout procedures for the equipment, machinery or system.
  • Page 14: Short Description & Specifications

    3 Short Description & Specifications Short Description & Specifications Application The Electronic Universal Sand Strength Machine, Model 42104, is used to determine the strength properties of clay and/or chemically bonded foundry sand specimens. A horizontally moving arm applies pressure on a sand specimen until failure.
  • Page 15 Short Description & Specifications 3 When testing green compression and shear strengths, the Electronic Universal Sand Strength Machine measures the force needed to reach the point of maximum strength. Maximum strength is defined at the point just before sand specimen destruction. Maximum strength is shown at point 1 on an example of a typical stress strain curve generated from a clay bonded molding sand (Figure 3.2.1).
  • Page 16: Specifications, Dimensions And Weights (Approximate)

    3 Short Description & Specifications The ability to accurately measure green deformation at maximum strength is another advantage of stopping the application of load to the sand specimen at maximum strength. Ease of measuring green deformation is an advantage of not destroying the sand specimen.
  • Page 17: Accessories

    Short Description & Specifications 3 Accessories 3.4.1 Cold Tensile Strength Accessory (Model 42104C) This accessory is mounted on the Electronic Universal Sand Strength Machine (Model 42104). The test determines the cold tensile strength of standard dog bone specimens prepared from oil, cold box, hot box and air set sands (no-bake). Specifications Cold Tensile Strength Accessory...
  • Page 18: Green Deformation Accessory (Model 42104E)

    3 Short Description & Specifications 3.4.3 Green Deformation Accessory (Model 42104E) This accessory measures the deformation of the green sand specimen after a compression test on the Electronic Universal Sand Strength Machine, Model 42104. Deformation is the change in the length of the sand specimen before and after the compression test.
  • Page 19: High Compression Strength Accessory (Model 42104H)

    Short Description & Specifications 3 3.4.5 High Compression Strength Accessory (Model 42104H) This accessory is used with the Electronic Universal Sand Strength Machine, Model 42104. It increases the force in a compression test up to a strength of 2100N/cm (3100 psi). Specifications High Compression Strength Accessory...
  • Page 20: Cold Shell Tensile Strength Accessory (Model 42104N)

    3 Short Description & Specifications 3.4.8 Cold Shell Tensile Strength Accessory (Model 42104N) This accessory to the Electronic Universal Sand Strength Machine, Model 42104, determines the cold shell tensile strength of an AFS standard ¼" (7.83 mm metric standard) thick dog bone specimen. It mounts mechanically to the arm and load cell of the strength machine.
  • Page 21: Unpacking And Installation

    Unpacking and Installation 4 Unpacking and Installation Unpacking Your new Laboratory Equipment has been closely inspected before being shipped to your plant. However, damage can occur in route, so it is wise to inspect all equipment on arrival. Notify both the carrier and Simpson Technologies of any damage at once.
  • Page 22: Components

    4 Unpacking and Installation Components Your new Electronic Universal Sand Strength Machine is shipped with the following accessories and installation components. Please take a moment to identify that the following items were included: • Electronic Universal Sand Strength Machine. Operating Instructions Manual. •...
  • Page 23: Installation

    Unpacking and Installation 4 Installation The installation of the apparatus is the responsibility of the Client to include procuring and preparing the material required for this purpose. It is recommended that the strength machine be situated close to the sand rammer or pneumatic sand squeezer.
  • Page 24: Connecting Power And Set-Up

    4 Unpacking and Installation A pressure regulator/filter and length of pneumatic hose required to connect the Electronic Universal Sand Strength Machine to the regulator/filter has been included with the Electronic Universal Sand Strength Machine. The compressed air should be free of dirt, debris and condensate.
  • Page 25: Airborne Noise Emission

    Unpacking and Installation 4 3. Assemble the provided pneumatic regulator/filter/lubricator according to the original equipment manufacturer's instructions supplied with the regulator/filter/lubricator. 4. Connect the assembled pneumatic regulator/filter/lubricator to the incoming compressed air line. 5. Remove the plastic shipping cap from the vent (Figure 7.1, Item 6) on the back of the strength machine.
  • Page 26: Operating Instructions

    5 Operating Instructions Operating Instructions For more information on how to use and care for your Simpson Analytics equipment accessories visit our Simpson Technologies channel on YouTube and search our library of videos. Subscribe to our channel to keep updated on new releases. Green or Dry Compression and Shear Strength Testing (Clay Bonded Molding Sands) 5.1.1...
  • Page 27 Operating Instructions 5 2. Install the adjustable clamp holder (Figure 5.1.2, Item 1) onto the moveable clamp arm of the strength machine (Figure 5.1.2, Item 5). Figure 5.1.2 3. Choose either the two compression clamps or the two shear clamps. Align the pins located on the back side of the clamps with the holes located on moveable clamp arm (Figures 5.1.2, Item 5) and stationary clamp holder (Figure 5.1.2, Item 6).
  • Page 28 5 Operating Instructions Figure 5.1.3 Figure 5.1.4 Refer to Figure 5.1.3. Showing detailed photos of Compression clamps properly installed on the strength machine. Refer to Figure 5.1.4 showing detailed photos of the shear clamps properly installed on the strength machine. Item Description Adjustable Clamp Holder...
  • Page 29 Operating Instructions 5 Operation 1. Turn on air supply and verify correct settings on the air regulator and oil level lubricator. Turn the power switch to the on position (Figure 7.1, Item 1). A "HELLO" message will briefly appear on the digital display. This message will be replaced by numerals.
  • Page 30 5 Operating Instructions Always transport prepared clay bonded molding sand samples from the shop floor to sand laboratory in a sealed container to help reduce compression strength measurement variation that may occur due to water evaporation from the sand sample. Temperature of the prepared molding sand sample should be tested while transporting the sample container from the shop floor to the laboratory.
  • Page 31: Cold Tensile Strength Accessory

    Operating Instructions 5 7. Remove the used sand specimen from the strength machine and record the compression strength value from the digital display. Do not run multiple compression or shear strength tests on the same sand specimen. The sand sample has been destroyed after the test. 8.
  • Page 32: Installation

    5 Operating Instructions 5.2.2 Installation 1. Remove the large knurled connecting screw (Figure 5.2.1, Item 1) from the base of the cold tensile accessory (Figure 5.2.1, Item 4). Remove both the tensile bridge (Figure 5.2.1, Item 2) and left tooling jaw from the base of the tensile accessory.
  • Page 33: Operation

    Operating Instructions 5 3. Place the left tooling jaw and tensile bridge on the top surface of tensile accessory base. Gently slide the bridge pin (Figure 5.2.1) into the hole located in the center of the stationary clamp holder (Figure 5.2.2, Item 6). Refer to Figures 5.2.4 and 5.2.6, which show detailed photos of the Cold Tensile Strength Accessory, Model 42104C, properly installed on the Electronic Universal Sand Strength Machine, Model 42104.
  • Page 34 5 Operating Instructions Figure 5.2.5 5. Press the "Test Selection" button (Figure 7.4, Item 13E) until the arrow pointing to "Core Tensile" illuminates. 6. Zero the digital display by pressing the "Zero" button (Figure 7.4, Item 13C). The digital display will read zero. 7.
  • Page 35: Splitting Strength Accessory

    Operating Instructions 5 Figure 5.2.6 Splitting Strength Accessory Part No. 0042104D 5.3.1 Description The Splitting Strength Accessory, Model 42104D, to the Electronic Universal Sand Strength Machine, Model 42104, consists of two specially designed clamps mounted to the strength machine to determine the splitting strength of a clay bonded molding sand in green or dry conditions.
  • Page 36: Preparing The Molding Sand Specimens For A Splitting Strength Test

    5 Operating Instructions It has been found that there is a close correlation between splitting strength and tensile strength in clay bonded molding sand. Since the tensile strength test is difficult and delicate, it is preferable to determine tensile strength as a function of splitting strength.
  • Page 37: Operation

    Operating Instructions 5 Figure 5.3.2 The splitting strength clamp located on the moveable clamp arm is designed to fit into the adjustable clamp holder. The screw on the adjustable clamp holder can be rotated clockwise and counterclockwise to change the location of the splitting strength clamp mounted onto the moveable clamp arm relative to the splitting strength clamp mounted onto the stationary clamp holder.
  • Page 38 5 Operating Instructions 3. Gently place a prepared 2" x 2" (50mm x 50mm) sand specimen in a vertical position on the flat supporting plates located on the bottom of the splitting strength clamps. The sand specimen shall be positioned so that the specimen will be pressed on the cylindrical surfaces between the splitting strength clamps.
  • Page 39 Operating Instructions 5 At the completion of the splitting strength test, the sand specimen is not physically broken. By continuously sensing the strength measurements, the strength machine automatically stops the load application at the exact point of maximum strength. Please refer to Section 3.2 “Description” for more detailed information regarding this advanced feature of the Electronic Universal Sand Strength Machine.
  • Page 40: Green Deformation Accessory

    5 Operating Instructions Green Deformation Accessory Part No. 0042104E 5.4.1 Description The Green Deformation Accessory allows an exact and simple measurement of the deformation of green sand samples subject to compression testing. The green deformation measurement is based on the unique feature of the Electronic Universal Sand Strength Machine, Model 42104, of not destroying the sand specimen upon completion of a green compression strength test.
  • Page 41: Operation

    Operating Instructions 5 5.4.2 Operation 1. Prepare a standard (AFS or metric) sand specimen using a standard sand rammer or pneumatic squeezer. (Simpson Technologies Models 42100, 42117 or 42160). 2. Gently lift the stem of the dial indicator on the Green Deformation Accessory (Figure 5.4.1) and place the sand specimen between the base and foot of the dial indicator (see Figure 5.4.2).
  • Page 42: High Compression Strength Accessory

    5 Operating Instructions 4. Gently lift the stem of the dial indicator and remove the sand specimen from the green deformation accessory after zeroing the dial indicator scale. 5. Carefully, load the measured sand specimen to the Electronic Universal Sand Strength Machine, Model 42104, and run a standard green compression strength following the instructions in Section 5.1 of this manual.
  • Page 43: Description

    Operating Instructions 5 5.5.1 Description The High Compression Strength Accessory, Model 42104H, is a force multiplier that when used with the Electronic Universal Sand Strength Machine, Model 42104, can perform a compression strength on a standard 2 in. x 2 in. AFS sand specimen (50mm x 50mm metric sand specimen) with a compressive strength range of 250 - 3100 PSI (170 - 2100 N/cm The accessory features a frame supporting a hydraulic force multiplier with...
  • Page 44: Installation

    5 Operating Instructions 5.5.2 Installation Remove the accessory mounting nut (Figure 5.5.1, Item 3) from the supporting frame of the High Compression Strength Accessory. Gently slide the High Compression Strength Accessory onto the moveable clamp arm (Figure 7.3, Item 12) of the strength machine. Replace the mounting nut (Figure 5.5.1, Item 3) and hand tighten to secure the accessory onto the moveable clamp arm of the strength machine (see Figure 5.5.2).
  • Page 45: Purging Air From High Compression Strength Accessory

    Operating Instructions 5 4. Gently hold in place a prepared 2" x 2" (50mm x 50mm) sand specimen between the moving head (Figure 5.5.1, Item 5) and the stationary clamp (Figure 5.5.1, Item 6) of the High Compression Strength Accessory. Using the sample clamp adjusting nut slowly tighten the sand specimen between the moving head and stationary clamp to secure the sand specimen between the two clamps.
  • Page 46 5 Operating Instructions 2. With the High Compression Strength Accessory mounted on the strength machine in an upright position remove the vent cap (Figure 5.5.1, Item 2). 3. By hand, quickly move the slide rod (Figure 5.5.1, item 7) in and out while observing any air bubbles that may escape through the oil reservoir (Figure 5.5.1, Item 1).
  • Page 47: Core Transverse Accessory

    Operating Instructions 5 Core Transverse Accessory Part No. 0042104K / 0042104K-M 5.6.1 Description The Core Transverse Accessory, Model 42104K, is used for transverse strength testing of chemically bonded foundry sand samples. The clamps of this accessory are mounted onto the Electronic Universal Sand Strength Machine, Model 42104.
  • Page 48: Operation

    5 Operating Instructions Figure 5.6.1 5.6.3 Operation 1. Turn on air supply and verify correct settings on the air regulator and oil level lubricator. Turn the power switch to the on position (Figure 7.1, Item 1). A "HELLO" message will briefly appear on the digital display. This message will be replaced by numerals.
  • Page 49: Shell Transverse Strength Accessory

    Operating Instructions 5 5. Press the "Start" button (Figure 7.4, Item 13F). The strength machine will begin applying load to the sand specimen. It will continue to apply load until sand specimen failure. The moveable clamp arm will return to home position and the tensile strength value will be displayed in the digital display.
  • Page 50: Installation

    5 Operating Instructions 5.7.2 Installation Turn the power switch (Figure 1, Item 1) to the off position when fastening the shell transverse accessory onto the sand strength machine. Be aware of a potential pinch hazard when inserting the shell transverse accessory onto the strength machine and when loading a sand specimen onto the transverse accessory.
  • Page 51 Operating Instructions 5 4. Zero the digital display by pressing the "Zero" button (Figure 7.4, Item 13C). The digital display will read zero. 5. Press the "Start" button (Figure 7.4, Item 13F). The strength machine will begin applying load to the sand specimen. It will continue to apply load until sand specimen failure.
  • Page 52: Cold Shell Tensile Strength Accessory

    5 Operating Instructions Cold Shell Tensile Strength Accessory Part No. 0042104N / 0042104N-M 5.8.1 Description The Cold Shell Tensile Accessory, Model 42104N, is used to determine tensile strength of shell coated (Croning) mold and core sands when mounted on the Electronic Universal Sand Strength Machine, Model 42104.
  • Page 53 Operating Instructions 5 2. Gently slide the threaded boss of tensile accessory base onto the moveable clamp arm (Figure 7.3, Item 12) of the strength machine. Replace the knurled screw and hand tighten to attached the tensile accessory base to the moveable clamp arm (see Figure 5.8.1).
  • Page 54: Operation

    5 Operating Instructions 5.8.3 Operation 1. Turn on air supply and verify correct settings on the air regulator and oil level lubricator. Turn the power switch to the on position (Figure 7.1, Item 1). A "HELLO" message will briefly appear on the digital display. This message will be replaced by numerals.
  • Page 55: Disk Transverse Attachment

    Operating Instructions 5 7. Press the "Start" button (Figure 7.4, Item 13F). The strength machine will begin applying load to the sand specimen. It will continue to apply load until sand specimen failure. The moveable clamp arm will return to home position and the tensile strength value will be displayed in the digital display.
  • Page 56: Operation

    5 Operating Instructions 2. Align the pins located on the back of the disk transverse support clamp and disk transverse fixed point clamp with the holes on the moveable clamp arm (Figure 7.3, Item 12) and the stationary clamp holder (Figure 7.3, Item 11). Gently slide the clamps onto the moveable clamp arm and the stationary clamp holder of the strength machine.
  • Page 57: Mor Bar Test Fixture Accessory

    Operating Instructions 5 2. Gently place a prepared disk transverse sand specimen onto the two support pins located on the bottom of the shell transverse accessory located on the moveable clamp arm. Position the sand specimen so that it makes contact with the edge of both outer blades of the disk transverse accessory.
  • Page 58: Installation

    5 Operating Instructions 5.10.2 Installation 1. Place the MOR Bar Test Fixture adapter ring onto the adjustable clamp holder supplied with accessory. The adapter ring is designed for and must be used with standard MOR specimens in order to properly space the specimen from the fixed portion of the test fixture mounted to the stationary clamp holder (Figure 7.3, Item 11).
  • Page 59: Error Messages

    Operating Instructions 5 5.11 Error Messages This information can be used to identify and troubleshoot any error messages that may appear in the digital display (Figure 7.4, Item 13B) of the Electronic Universal Sand Strength Machine, Model 42104. Any time during operation, if the machine senses a system functioning out of control, an error message will be shown on the digital display.
  • Page 60: Maintenance And Calibration

    6 Maintenance and Calibration Maintenance and Calibration For more information on how to use and care for your Simpson Analytics equipment and accessories visit our Simpson Technologies channel on YouTube and search our library of videos. Subscribe to our channel to keep updated on new releases.
  • Page 61: Monthly Maintenance

    Maintenance and Calibration 6 6.1.3 Monthly Maintenance • Lightly oil external part of the main stem with an SAE oil. 6.1.4 Biennial Maintenance (every two years) Change the hydraulic oil with a Shell Tellus 27 or equivalent. Proceed in the •...
  • Page 62: Calibration

    6 Maintenance and Calibration Make certain that the end of the hose located in the oil container is kept below the oil level to allow oil to be suctioned into the strength machine. 8. Start the machine and continue to cycle the machine until the oil level in the container does not descend.
  • Page 63: Calibrating The Strength Machine

    Maintenance and Calibration 6 6.2.2 Calibrating the Strength Machine The electronic load cell on the Model 42104 should be calibrated once every six months and the mechanical load cell should be calibrated every two years, provided it is not subjected to abuse. 1.
  • Page 64 6 Maintenance and Calibration 8. Replace the attachments with the adjustable support ball and fixed support ball that are included with the Mechanical Load Cell, Model 42125. The adjustable support is inserted onto the movable clamp arm (Figure 7.3, Item 12) by gently sliding into the mating hole (see Figure 6.2.1).
  • Page 65 Maintenance and Calibration 6 10. Prior to mounting the mechanical load cell into the strength machine, clean and lightly oil the two seating sockets located on the right and left of the mechanical load cell. Figure 6.2.3 11. Mount the load cell between the stationary clamp holder (Figure 7.3, Item 11) and the movable clamp arm (Figure 7.3, Item 12).(See Figure 6.2.4) Figure 6.2.4 12.
  • Page 66 6 Maintenance and Calibration Example of Control Standardization Table Check Points Calibration Set Point Point 1 Point 2 Point 3 Deflection Force Deflection Force Deflection Force .01 mm Newtons 0.1 mm Newtons 0.1 mm Newtons 227.0 3198 44.5 135.0 1909 NOTE: The values for each mechanical load cell will be different than in the table above.
  • Page 67: Regulating The Load Application Rate

    Maintenance and Calibration 6 6.2.3 Regulating the Load Application Rate The Electronic Universal Sand Strength Machine is capable of an infinitely variable load application rate. By adjusting the speed of the hydraulically regulated pneumatic cylinder the user can choose and set a load rate.
  • Page 68: Set-Up And Reset Functions

    6 Maintenance and Calibration Set-up and Reset Functions Simpson Technologies does not recommend using the “SET-UP” and “RESET" modes when calibrating this machine. These functions and procedures are provided for Simpson Technologies technical servicemen. 1. Turn the power switch to the on position (Figure 7.1, Item1). 2.
  • Page 69 Maintenance and Calibration 6 SET-UP: 5. The "SET-UP" function sets the content of the strength machine's registers. To select "SET-UP" press the "Start" button (Figures 7.3 & 7.4, Item 13F) while the "SET-UP" message appears on the digital display. After pressing, the display will show the "HELLO"...
  • Page 70 6 Maintenance and Calibration RESET: 8. The "rESET" function sets the machine within the fabrication parameters. To reset the fabrication parameters, press the "Start" button (Figure 7.4, Item 13F) when the "rESET" message appears on the digital display. After pressing the "Start"...
  • Page 71: Apparatus Layout 7

    Apparatus Layout 7 Apparatus Layout Figure 7.1: Back View Item Description Power Switch Fuse Holder Power Cord Receptacle RS232 Port Pneumatic Input Vent Port Vent Port Vent Port man-stc-42104-V18-7 Electronic Universal Sand Strength Machine...
  • Page 72 7 Apparatus Layout Figure 7.2: Right Side View Item Description Hydraulic Brake Adjustment Motor Adjustment Electronic Universal Sand Strength Machine man-stc-42104-V18-7...
  • Page 73 Apparatus Layout 7 Figure 7.3: Front View Item Description Stationary Clamp Holder Moveable Clamp Arm Electric Panel Leveling Foot (4) man-stc-42104-V18-7 Electronic Universal Sand Strength Machine...
  • Page 74 7 Apparatus Layout Figure 7.4: Electronic Panel Item Symbol Description Electronic Panel Units of Measure Display Digital Display Zero Button Clear Display Button Test Selection Button Start Button Test Selection Indicator Standard Selector Switch Electronic Universal Sand Strength Machine man-stc-42104-V18-7...
  • Page 75 Apparatus Layout 7 Figure 7.5: 13G Test Selection Detail Item Description 13G-1 Compression Strength Symbol 13G-2 Shear or Split Strength Symbol 13G-3 Core Tensile Strength Symbol 13G-4 Shell-Hot Box Tensile Strength Symbol 13G-5 Core Transverse Strength Symbol 13G-6 Force Symbol man-stc-42104-V18-7 Electronic Universal Sand Strength Machine...
  • Page 76: Parts List / Ordering Parts / Returns

    8 Parts List / Ordering Parts / Returns Parts List / Ordering Parts / Returns Spare Parts List Simpson maintains a large inventory of common spare parts for all current Simpson Analytics products. The following table provides part numbers for common spare parts for this device.
  • Page 77: Returned Goods Policy

    Parts List / Ordering Parts / Returns 8 Returned Goods Policy Simpson Technologies strives to provide their customers with maximum follow up support and, in order to offer the most practical flexibility, the following conditions apply to returned goods. Adherence to these procedures will assure the most prompt and efficient service.
  • Page 78: Decommissioning

    10 Location of Safety Symbols Decommissioning Before doing any work, review the Safety Procedures in Section 2 and Lockout/Tagout all the power sources to the machine and peripheral equipment. Failure to follow safety procedures could result in serious injury. Use qualified personal and follow safety procedures, applicable local policies and regulations in decommissioning the Electronic Universal Sand Strength Machine and peripheral equipment.
  • Page 79 9 Decommissioning This Page is Intentionally Blank man-stc-42104-V18-7 Electronic Universal Sand Strength Machine...
  • Page 80 In North America Simpson Technologies 751 Shoreline Drive Aurora, IL 60504-6194 Tel: +1 (630) 978 0044 Fax: +1 (630) 978 0068 In Europe Simpson Technologies GmbH Roizheimer Strasse 180 53879 Euskirchen, Germany Tel: +49 (0) 2251 9460 12 Fax: +49 (0) 2251 9460 49 simpsongroup.com Copyright 2023.

This manual is also suitable for:

Simpson 0042104-asmSimpson 0042104-m-asm

Table of Contents