Summary of Contents for Kobold SEN-Y48382

  • Page 1 Operating Instructions Pressure Sensor Model: SEN-Y48382/3B95...
  • Page 2: Table Of Contents

      EU Declaration of Conformance ..............11     UK Declaration of Conformity ..............12     Manufactured and sold by: Kobold Messring GmbH Nordring 22-24 D-65719 Hofheim Tel.: +49(0)6192-2990 Fax: +49(0)6192-23398 E-Mail: info.de@kobold.com Internet: www.kobold.com page 2 SEN-3Y K01/0324...
  • Page 3: Note

    Vessels referred to in Article 4(1)(a)(i), second indent 3. Regulation Use Any use of the Pressure Sensors Model: SEN-Y48382/3B95 which exceeds the manufacturer’s specification may invalidate its warranty. Therefore, any resulting damage is not the responsibility of the manufacturer. The user assumes all risk for such usage.
  • Page 4: Operating Principle

    Should the damage to a device be visible, we recommend a thorough inspection of the delivery packing. In case of damage, please inform your parcel service / forwarding agent immediately, since they are responsible for damages during transit. Scope of delivery:  Pressure Sensor model: SEN-Y48382/3B95 page 4 SEN-3Y K01/0324...
  • Page 5: Mechanical Connection

    6.2. Sealing the process connection Process connection of SEN-Y48382/3B95 is tapered thread (1/8” NPT). For sealing process connections with tapered threads, the sealing must be made in the threads using additional sealing material, e.g. PTFE tape (EN 837-2).
  • Page 6: Installing The Instrument

    SEN-Y48382/3B95 6.3. Installing the instrument When screwing the instrument in, the force required to do this must not be applied through the case or the cap ring, but only through the spanner flats provided for this purpose and using a suitable tool.
  • Page 7: Electrical Connection

    SEN-Y48382/3B95 7. Electrical Connection  Ensure that the power is disconnected during connection of the cable.  To connect the instrument, the cable outlet must be assembled in advance.  Only use cables with suitable characteristics for the particular operating conditions.
  • Page 8: Trouble Shooting

    SEN-Y48382/3B95 8. Trouble Shooting Trouble Possible reason What's to do No/wrong output signal no supply voltage Check your power supply and broken wire wiring. lf necessary replace defective parts Sensor has been wired Check the wiring according to improperly the sketches and adjust wiring if necessary.
  • Page 9: Technical Information

    SEN-Y48382/3B95 9. Technical Information Sensor element: Model: SEN-Y48382/3B95 High precision pressure sensor for industrial application Material housing/ Stainless steel 316L wetted parts: Non-linearity ≤ ±0.25 % of span, per BFSL per IEC 61298-2 Zero-point error: ≤ ±0.2 % of span, factory set Stability/year: ≤...
  • Page 10: Maintenance

    SEN-Y48382/3B95 10. Maintenance The pressure sensor described in this manual is maintenance free! It does not contain any components which may be repaired or exchanged locally. Repairs are only possible in our factory. 11. Disposal Note!  Avoid environmental damage caused by media-contaminated parts.
  • Page 11: Eu Declaration Of Conformance

    SEN-Y48382/3B95 12. EU Declaration of Conformance We, KOBOLD Messring GmbH, Nordring 22-24, 65719 Hofheim, Germany, declare under our sole responsibility that the product: Pressure Sensor Model: SEN-Y48382/3B95 to which this declaration relates is in conformity with the following EU directives...
  • Page 12: Declaration Of Conformity

    SEN-Y48382/3B95 13. UK Declaration of Conformity We, KOBOLD Messring GmbH, Nordring 22-24, 65719 Hofheim, Germany, declare under our sole responsibility that the product: Pressure Sensor Model: SEN-Y48382/3B95 to which this declaration relates is in conformity with the following UK directives stated below: S.I.