SE-MAN-003 Sample Conditioning Systems Technical Manual PERMA PURE LLC Revision: 000 Page 3 of 16 Specifications Property 4S-9A 4S-9PA 4S-9AA 4S-9PAPA 4S-9- 10410 Sample gas flow range Water content Maximum cooling rate Maximum inlet sample temp. Maximum inlet pressure Maximum pressure drop Ambient temperature range Outlet sample gas dew point Voltage Power supply...
Page 4
Sample Conditioning PERMA PURE LLC Systems User’s Manual Doc. #580: Revision: 000 Page 4 of 16 Copy M-Class Safety info...
SE-MAN-003 Sample Conditioning Systems Technical Manual PERMA PURE LLC Revision: 000 Page 5 of 16 Installation Mounting 1. Rack 2. Wall Wiring Plumbing Utilities Start-up Instructions LED Summary (see Cooler Manual) Start-up The gas sample pump is controlled and will operate only under certain conditions. The sample pump will only be allowed to run if the temperature is 10°C or less AND the water slip sensor indicators are green.
Page 6
Sample Conditioning PERMA PURE LLC Systems User’s Manual Doc. #580: Revision: 000 Page 6 of 16 Troubleshooting and Maintenance Guide Symptom Diagnostic No LED(s) and no fan. AC power input. Ensure that AC power is connected. No LED(s) and fan on. AC input fuse on control board.
Page 7
SE-MAN-003 Sample Conditioning Systems Technical Manual PERMA PURE LLC Revision: 000 Page 7 of 16 relay/water slip option installed). Ambient air temperature too high Ambient must not exceed 43°C (110°F) Operation of peristaltic pump(s) or Verify that pump or eductor is eductor draining condensate from impingers;...
SE-MAN-003 Sample Conditioning Systems Technical Manual PERMA PURE LLC Revision: 000 Page 9 of 16 Appendix A: Installation of Peristaltic Pump NOTE: Perma Pure peristaltic pumps are rated for a maximum pressure of 20 psi. If supplied loading key is not available, use a flat head screwdriver. Single pump head loading Separate the end bells (the pump head halves).
Page 10
Sample Conditioning PERMA PURE LLC Systems User’s Manual Doc. #580: Revision: 000 Page 10 of 16...
Sample Conditioning PERMA PURE LLC Systems User’s Manual Doc. #580: Revision: 000 Page 12 of 16 Appendix C: Perma Pure Sample Pump Maintenance Instructions Item Description VALVEBODY BEARING CWEIGHT CAPACITOR ECCENTRIC FAN, BLUE FAN, WHITE HEAD STD, SS PLATE DIAPH. CONNECTING ROD TOL.
SE-MAN-003 Sample Conditioning Systems Technical Manual PERMA PURE LLC Revision: 000 Page 13 of 16 Appendix D: Dual Stream Operation The dual stream option was not purchased for this unit. COPYRIGHT 2008, ALL RIGHTS RESERVED...
Sample Conditioning PERMA PURE LLC Systems User’s Manual Doc. #580: Revision: 000 Page 14 of 16 Appendix E: Conversion Tables CHART OF VOLUME PERCENT WATER CONCENTRATIONS AT SATURATION FOR VARIOUS TEMPERATURES AT STANDARD PRESSURE (ATMOSPHERIC PRESSURE) DEGREES C DEGREES F VOLUME % DEGREES C DEGREES F...
Page 15
SE-MAN-003 Sample Conditioning Systems Technical Manual PERMA PURE LLC Revision: 000 Page 15 of 16 MOISTURE CONVERSION TABLE DEW POINT DEW POINT VAPOR PRESSURE RELATIVEHUMIDITY ppm by WEIGHT ppm by VOLUME °F °C Mm Hg @70°F in air -110 -166 .0000010 .00132 .0000053 0.00082 -108 -162 .0000018...
Sample Conditioning PERMA PURE LLC Systems User’s Manual Doc. #580: Revision: 000 Page 16 of 16 Appendix F: Warranty and Disclaimer Perma Pure LLC Perma Pure (Seller) warrants that product supplied hereunder shall, at the time of delivery to Buyer, conform to the published specifications of Seller and be free from defects in material and workmanship under normal use and service.