shelolab SRI3 Installation And Operational Manual

Low temperature incubator

Advertisement

Quick Links

LOW TEMPERATURE
INCUBATOR
SRI3 SRI3-2
Previously designated as:
LI5 LI5-2
INSTALLATION AND OPERATIONAL MANUAL
Sheldon Manufacturing Inc. P.O. Box 627 Cornelius, Oregon 97113
EMAIL:
tech@Shellab.com
INTERNET:
http://www.Shellab.com/~Shellab
1-800-322-4897
(503) 640-3000
FAX (503) 640-1366
REV. 01/11
4861574

Advertisement

loading
Need help?

Need help?

Do you have a question about the SRI3 and is the answer not in the manual?

Questions and answers

Summary of Contents for shelolab SRI3

  • Page 1 LOW TEMPERATURE INCUBATOR SRI3 SRI3-2 Previously designated as: LI5 LI5-2 INSTALLATION AND OPERATIONAL MANUAL Sheldon Manufacturing Inc. P.O. Box 627 Cornelius, Oregon 97113 EMAIL: tech@Shellab.com INTERNET: http://www.Shellab.com/~Shellab 1-800-322-4897 (503) 640-3000 FAX (503) 640-1366 REV. 01/11 4861574...
  • Page 2: Table Of Contents

    TABLE OF CONTENTS SECTION 1.0 RECEIVING AND INSPECTION SECTION 2.0 INSTALLATION SECTION 3.0 CONTROL OVERVIEW SECTION 4.0 OPERATION SECTION 5.0 MAINTENANCE SECTION 6.0 TROUBLESHOOTING SECTION 7.0 PARTS LIST UNIT SPECIFICATIONS SCHEMATICS This unit is a general purpose Biochemical Oxygen Demand (BOD) incubator for professional, industrial or educational use where the preparation or testing of materials is done at approximately atmospheric pressure and no flammable, volatile or combustible materials are being heated.
  • Page 3: Receiving And Inspection

    Section RECEIVING AND INSPECTION Your satisfaction and safety require a complete understanding of this unit. Read the instructions thoroughly and be sure all operators are given adequate training before attempting to put the unit in service. NOTE: This equipment must be used only for its intended application;...
  • Page 4: Installation

    Section INSTALLATION Local city, county or other ordinances may govern the use of this equipment. If you have any questions about local requirements, please contact the appropriate local agency. Installation may be performed by the end user. Under normal circumstances this unit is intended for use indoors, at room temperatures between 5...
  • Page 5: Control Overview

    Section CONTROL PANEL OVERVIEW Power Switch: The main power I/O (on/off) switch controls all power to the unit and must be in the I/ON position before any systems are operational. Main Temperature Control: This control is marked SET TEMPERATURE and consists of the digital display and UP/DOWN arrow pads for inputting set point temperatures and calibration.
  • Page 6: Operation

    Section OPERATION Check power supply against unit serial plate. They must match. Plug service cord into the grounded electrical outlet. Push the power switch to the I/On position, and turn the Overtemperature Thermostat to its maximum position, clockwise using a coin or flat edged tool. Set Main Temperature Controller: Enter desired set point temperature.
  • Page 7 Load the chamber with sample bottles. See Figure 2 for bottle placement. Figure 2...
  • Page 8: Maintenance

    Section MAINTENANCE Note: Prior to performing maintenance and/or service on this incubator, disconnect the unit from the power supply. Cleaning: Cleaning and decontamination are recommended on a regular basis. To prepare the unit for cleaning, disconnect the power cord and remove all interior parts, such as shelves, trays, etc.
  • Page 9: Troubleshooting

    Section TROUBLESHOOTING FOR PERSONAL SAFETY, ALWAYS DISCONNECT THE POWER BEFORE SERVICING. In the event the incubator does not operate properly, check the following before calling for service. Always make a visual inspection of the incubator and control console when troubleshooting to find loose or disconnected wires which may be the source of trouble.
  • Page 10 3/ Assure there is ample room around the unit as described in Installation. 4/ Compare ambient specifications to Unit specifications. If compressor isn’t running: 1/ Confirm that compressor cooling fan motor is operable a- Dirty coil or poor circulation b- Coil next to heat source c- Ambient temperature too high Ice build up in chamber 1/ Search for leak in door gasket.
  • Page 11: Parts List

    Section PARTS LIST Description 115V 220V Circuit Breaker 1100505 1100505 Circulating Fan, chamber 4880545 4880545 Convenience Outlet 6100525 6100526 Cooling Fan, compressor 4880564 4880564 Factory Set Over- 1750715 1750715 Temperature Control 2350508 2350508 Heating Element 44.6 ohms 150.1 ohms I/O (On/Off) Power Switch 7850570 7850570 Main Temperature Controller...
  • Page 12: Unit Specifications

    UNIT SPECIFICATIONS Weight Shipping SRI3 (LI5) 130 lbs. 108 lbs. SIR3-2 (LI5-2) Dimensions Exterior WxDxH Interior WxDxH SRI3 (LI5) 24 x 21 x 33.75 in. 16 x 12 x 21.5 in. SIR3-2 (LI5-2) 60.96 x 53.34 x 85.73 cm. 40.64 x 30.48 x 54.61 cm.
  • Page 13 WIRE DIAGRAM 110V (9851235) 120VAC POWER CORD AND PLUG 1100505 CIRCUIT BREAKER 15A GREEN LIGHTED POWER SWITCH 7850570 OUTLETS 6100525 4880545 INTERNAL FAN SENSOR TEMPERATURE NEUTRAL CONTROL 1750967 GROUND 4650553 SAFETY ACTIVATED LOAD 4650554 HEATING ACTIVATED 45 1750715 155°F/68°C 1750862 9570863 ELEMENT ASSY 2350508 ELEMENT COIL THE COMPRESSOR IS PART...
  • Page 14 220V (9851236) POWER CORD 1800541 2800502 EMI FILTER FUSE FUSE GREEN LIGHTED POWER SWITCH 7850570 OUTLETS 6100531 TEMPERATURE NEUTRAL CONTROL 1750968 GROUND LOAD SAFETY ACTIVATED 4650553 HEATING ACTIVATED 4650554  1750715 155°F/68°C ½ 1750813 ½ 1750813 9570864 ELEMENT ASSY 2350508 ELEMENT COIL 8350515 TRANSFORMER 240/120VAC 750VA THE COMPRESSOR IS PART...