MSL2S/3S/5S User Manual Slit lamp Bio-Microscope To ensure proper use of this instrument as well,as to avoid injury while operating instrument,understanding this manual completely before use is highly recommended...
Page 2
PREFACE Congratulations for purchasing our MIKO slit lamp. The slit lamp will have another capability of photography, if you purchase the appliance with the photography accessories. At the same time, in order to meet the needs of the customs, we provide sorts of accessories, which can be used widely.
INTRODUCTION MSL-2S/3S/5S Slit Lamp is a professional grade examination device that consists of a moveable light column and stereoscopic microscope with which the phthalmologist can examine the eye. Significant user convenience has been incorporated in both the mechanical and optical system of the equipment. Extra clarity and contrast is provided through the Optical System with illumination of 12V-50W halogen.
UNPACKING YOUR SLIT LAMP Remove the Slit-Lamp components from the packing boxes and before any of the packing material is discarded I check the components against the following equipment list. First Layer : Slit lamp wooden base with power source and accessories box Middle layer : chin rest stand Bottom layer...
INSTALLATION & ASSEMBLY INITIAL SETUP.. Fix the Manual stand or motorized stand Manual Table Motorized table Before you begin the assembly process: Be sure the floor is level. Be sure your stand is correctly positioned in the desired location. Wooden base: Place the table top on Motorized table/manual table /refraction unit or level base Chin-Rest: Fix the chin-rest bottom on the bottom wooden base plate.
Page 10
INSTALLATION (Continuous) Movement Base: Place the slit-lamp movement base on the tabletop. He perforated roller should be evenly fixed with the toothed rack on the table Insert the rack guards under the toothed rack Loosen the locking screw and insert the Microscope arm fitted with Microscope on the pillar of the base...
Page 11
INSTALLATION (Continuous) Loosen the locking screw and insert the Illumination column in the pillar of the base Connect the power cord and LED fixation lamp to labeled output voltage Insert the illumination lamp plug...
Page 12
INSTALLATION (Continuous) Switch On the light and control the intensity at low Check the movement and power supply...
OPERATING PROCEDURE Switch on the power using the On/Off switch located Below the wooden table on the front of the Power Supply Assembly box. Intensity of light can be adjusted by rotating the intensity level knob. Note: The maximum position is for intermittent use only. Continuous use will shorten lamp life Insert the Focusing Rod in the pivot post of the instrument body to make rough PD and focus adjustments.
Page 14
OPERATING (Continuous) To position a patient, adjust the chinrest height by turning the Chinrest Elevation Handle on the post of the Chin Rest Assembly until the patient’s canthus is in line with the Canthus Mark on the chin rest post. Microscope elevation is adjusted by rotating the joystick and observing the slit image through the Microscope Assembly until the slit is centered on the patient's cornea.
Page 15
OPERATING (Continuous) The angle between the illumination system and the microscope can be varied between 0° and 90° to either the left or right. The illumination angle is indicated on the Illumination Angle Scale on the slit lamp arm. Magnification is altered by rotating the Magnification Dial on the Microscope Assembly.
Page 16
OPERATING (Continuous) Slit Rotation By grasping the Slit Rotation Knob, the lamp housing can be rotated. This, in turn, rotates the slit from vertical to horizontal in either direction. The slit positions are click- stopped in 45° increments and stopped at 0° and 180° and is indicated by the scale. Illumination Inclination The Illumination Assembly can be inclined in the horizontal plane in 5°...
CLEANING External Cleaning Clean the external surfaces of this instrument using a clean, soft cloth moistened with a mild detergent solution (1 cc of liquid dish soap to one liter of clean, filtered water (filtered below 5 microns)). Forehead / Chinrest Preparation For hygienic reasons, wipe the forehead rest and chin rest with an alcohol wipe after each patient.
CHANGING THE BULB WARNING: NEVER REMOVE A BULB THAT HAS RECENTLY BEEN IN USE AS IT WILL BE VERY HOT. WAIT UNTIL IT HAS COOLED AND USE GLOVES OR A THICK CLOTH WHEN HANDLING ANY HALOGEN BULB. WARNING: NEVER TOUCH A HALOGEN BULB WITH BARE HANDS AS FINGERPRINTS WILL SHORTEN THE BULB LIFE.
The following chart outlines some common issues with the MSL-2S/3S/5S Slit Lamp and some steps you can take to correct the issue. If problems persist, please contact MIKO as listed in the Introduction section of this manual. Chart of Common Errors...
(An authorized dealer shall not be considered an original purchaser.) Under this warranty, MIKO sole obligation is to repair or replace the defective part or product at MIKO discretion.