Date of Purchase: PREVENTATIVE MAINTENANCE: Just like your car - a Podiatry Drill will operate best when serviced regularly. In consultation with the Manufacturer, Briggate Medical Company recommend that this appliance be sent to us for inspection every: 18 MONTHS...
Briggate Medical Company’s service facility; 23-25 Lakewood Blvd, Braeside, Victoria, 3195. No freight charges will be accepted for the return of this appliance to Briggate Medical Company - regardless of warranty claim outcome. The Suda Pedo Sprint warranty does not cover the wear and tear on the clutch mechanism in the handpiece.
Page 3
INDEX Technical Data Safety Instructions Main Features & Controls Initial Set-up General Functions - Tool Rotation Speed (R.P.M) - Adjusting Tool Rotation - Vacuum Strength Level - Starting/Stopping the Device - Programming Memory Buttons Inserting/Removing Tools Tool Requirements Replacing the Fuse Replacing Filter Bags Replacing the Vacuum Motor Filter Handpiece Dismantling &...
Technical Data Measurements: 250mmW x 140mmH x 185mmD Equipment: Main Control Box, Micro-Motor Handpiece, Suction/Power Hose, Dust Bag, Mains Power Lead, Handpiece Holder, Cleaning Brush Total Weight: 3.1kg Handpiece Weight: 98 g Noise Level: 60 dB Voltage: 220-240V / 50-60 Hz Warranty: 24 Months (Return to Base) Fuse:...
• The Dust Bag Faceplate must be connected properly into the Dust Bag Chamber Location of Main Control Unit When choosing the location of your Suda Pedo Sprint 50SP, please follow the below guidelines: - place the device on a firm, level surface...
General Functions Starting/Stopping the Device 1. Press the On/Off Button on the control panel to start the device. vacuum rotation 2. Once the device is started the Handpiece can be paused via the on/off on/off Handpiece On/Off Button located near the connection of the Handpiece to the Suction Hose.
Only use quality tools which have a shaft diameter of 2.35mm (some inferior tools may have undersized shafts which may slip and damage the Handpiece). All tools supplied by Briggate Medical Company meet quality and size requirements as specified by the manufacturer.
Page 9
7. Ensure the Dust Bag Faceplate is connected properly to the Main Control Unit prior to operation. ONLY USE ORIGINAL SUDA DUST BAGS AS SUPPLIED BY BRIGGATE MEDICAL COMPANY NEVER RE-USE FILTER BAGS. DOING SO IS UNHYGIENIC AND MAY LEAD TO A BAG TEAR RESULTING IN DAMAGE TO THE VACUUM MOTOR.
Page 10
Handpiece Dismantling & Cleaning ALWAYS ENSURE UNIT IS SWITCHED OFF AND MAIN POWER LEAD IS REMOVED FROM WALL SOCKET PRIOR TO DISMANTLING THE HANDPIECE. This proceedure should be carried out at least every two weeks to ensure optimum performance and reduce the risk of damage to your Suda Pedo Sprint Podiatry Drill.
Page 11
Handpiece Dismantling & Cleaning (cont.) Suction Hose Connection 4. To allow the Micro-Motor to fit snugggly in the Suction Hose Connection, gently pull on the Suction Hose. This will help draw the Micro-Motor into the Fastening Pegs located within the Suction Hose Connection. Ensure to line-up the green marking near the On/Off Button to the...
Page 12
Hints & Troubleshooting Fault Cause Solution Remove and re-insert Mains Power Lead to Mains Power Lead not connected properly Device does not ensure proper connection function Plug another electrical device into the Wall No power supply from Mains Power Wall Socket Socket to determine if the Wall Socket has power Refer to page 6 for fuse replacement...
Page 13
Quality Drill Cosumables Refer to www.briggatemedical.com for our full range of drill consumables available from the following manufacturers: Burrs: Sanding Discs: Mandrels:...