Squirrel Dormir 8535 MRS Dormir 8535 MRS / Dormir 8535 Plus MRS Instructions for Use Read the instructions in this booklet thoroughly prior to use. IMPORTANT! Keep these instructions for future reference.
ALWAYS keep these instructions with the Dormir 8535 MRS for reference Do ensure that you fully read & understand these Instructions for Use • Do ensure the Dormir 8535 MRS is assembled and installed in accordance with • these Instructions for Use Intended Use Ensure the Dormir 8535 MRS is always used as intended, see 2 - •...
Product Models & SWL The Dormir 8535 MRS is a pressure redistributing (active) mattress replacement system. It is comprised primarily of two components – the Pump and the Mattress. These instructions cover the following models; Systems Safe Working Model Description...
Warning Symbols WARNINGS given in this manual identify possible hazards in procedures or conditions, which if not correctly followed, could result in death, injury or other serious adverse reactions. Cautions given in this manual identify possible hazards in procedures or conditions, which if not followed, could result in equipment damage or failure, and indicate moderate risk.
2. Intended Use The Squirrel Medical Active Mattress Systems are intended to be used as an aid to assist in the managed treatment and prevention of pressure ulcers when used in conjunction with a patient care plan and/or pressure ulcer prevention program and must not be considered or relied upon as an independent curative device.
The Squirrel Dormir 8535 MRS is a mattress replacement and must not be used as an overlay. The intended Safe Working Load (SWL) of a Dormir 8535 MRS is up to 150Kg. There should only every be a single occupant on a Dormir 8535 MRS.
Collateral standard. Electromagnetic compatibility. Requirements and tests. Product Description The Dormir 8535 MRS is comprised of two main parts, the mattress and the pump. Both parts are identifiable by a serial number and part number or name. The model of compatible Dormir 8535 pumps are:...
3.2.1 Identification of Pump Parts Dormir 8535 Pumps (PUMP8535 illustrated) 8535 Pump unit Hook Covers Control Panel Air Filter Tube Set Connector Socket Extra Deep Pump Hook Fuse Holders Serial Number Label Standard Pump Hooks (fitted) Power Cord Pump Model Identification PUMP8535 - Dormir 8535 Pump PUMP8535CLP - Dormir 8535 Plus Pump Sep 2023...
3.2.2 Identification of Mattress Parts Dormir Mattress Dormir Mattress Top Cover Transport Cap Tube Set “O” Ring CPR Valve Air Cells Tube Set Connector Plug Foam Mattress Base Page 10 of 41 Copyright Squirrel Medical® 2017 - E&OE...
4. Installation Unpacking Carefully open the boxes and unpack all components. While unpacking, be aware some of the larger components are heavy. Technical Specifications. If required, please take precautions before attempting to lift or move any items. To avoid danger of suffocation, keep packaging and plastic bags away from babies and children.
Page 12
Each time the Dormir 8535 MRS is set up, specific attention must be drawn to any potential dangers which can occur despite correct operation. Before putting the Dormir 8535 MRS into service for the first time, the Instructions for Use must be read conscientiously and in detail by the User / Care personnel.
Page 13
Make sure that the power cable and tube set are routed so it does not snag the bed or cause a trip hazard. NEVER route the tube set on the floor between the castors at the Head or Foot Ends of an adjustable height Bed. Doing so and raising a bed will pull the tubes apart and damage your unit.
5. Dormir 8535 MRS Operation Pre-Use Routine Each time after you have set up the Dormir 8535 MRS, check the following before proceeding with normal operation: Pre-Use Checklist Check that these Instructions for Use (IFU) are easily accessible. Check that the pump hooks are the correct size for the footboard and that ...
The Dormir 8535 Pump Controls 5.2.1 8535 Control Panel – PUMP8535 5.2.2 8535 Plus Pump Control Panel - PUMP8535CLP On / Off Button This button toggles the power off and on. Comfort Adjust Buttons These buttons set the pressure in the mattress. Comfort Level Indicator These lights indicate the current Comfort level setting.
The Comfort Level Indicator will indicate softer or firmer setting. NOTE: The Dormir 8535 MRS has been designed to aid in the managed care and prevention of pressure ulcers and must be used with sound clinical judgment...
The Dormir 8535 pumps are designed with a safe working patient weight range of 150Kg in prone position. There is no weight guide/setting needed. The Dormir 8535 MRS has been designed to aid in the managed care and prevention of pressure ulcers and must be used with sound clinical judgment. Alarms & Indicators...
Press down to seal before each use In an emergency, pull CPR Strap to open valves and quickly deflate mattress Page 18 of 41 Copyright Squirrel Medical® 2017 - E&OE...
Mattress Transfer Mode This mode allows for some air to be sealed inside the mattress base allowing the mattress to be transferred to different locations (e.g. for cleaning and servicing) without deflating. Do not use mattress transfer mode with a patient in the bed. The transfer mode must not be used if a patient is using the mattress.
When storing the Pump and Mattress, following the guidance below will help to ensure the 10 - Technical Specifications, page 33. working life of the Dormir 8535 MRS. Also see Before the Dormir 8535 MRS is stored or transported, please ensure; The pump and the mattress have been cleaned and are fully dry ...
Never use any type of water or steam high pressure cleaner, jet or spray on the Dormir 8535 MRS, use detergent/disinfectant on a damp cloth only. Do not use phenol-based disinfectant solutions. Never attempt to clean a Dormir 8535 MRS while there is a patient occupying it. Excessive heat can damage the cover Sep 2023 Issue 4.5...
Cross contamination is a major hazard to patients. The Mattress Cover should therefore be laundered in between patients. Spare covers are available to ensure that a Mattress is never out of use. Page 22 of 41 Copyright Squirrel Medical® 2017 - E&OE...
Unzip and remove the PU top cover from the mattress. The Mattress Cover all around the perimeter of the Mattress and wash according to the instructions below. Always ensure that the Squirrel Dormir 8535 MRS is cleaned prior to transfer to another location i.e. for repair.
Staff should refer to the safe work practice sheet on the use of workplace equipment contained in the department safety statement. To disinfect the Pump, wear suitable protective clothing Page 24 of 41 Copyright Squirrel Medical® 2017 - E&OE...
Pre-Use Routine - See Section 5, there are weekly and monthly checks you can do to ensure your Dormir 8535 MRS is maintained and in a safe working condition. These can be found in the Service Manual.
Any faults, defects, damage or excessive wear must be dealt with immediately or the Dormir 8535 MRS should be taken out of service A full service of the Dormir 8535 MRS should be performed by Squirrel Medical or by suitably qualified and trained personnel at least once per year •...
Changing the Air Filter To change the Air Filter, you will need the following; Spare Air Filter To Change the Air Filter Disconnect the Pump from the mains supply Disconnect the Tube Set and position the Pump unit so you have easy access to the Air Filter Door Gently press in the small lever, and prise open the Air Filter Door as shown.
Screw the new fuse and cover back into place firmly, taking care not to overtighten. Repeat the procedure above with the second Fuse Holder Reconnect the Tube Set and Mains power, and test the system Warning! Never open the Dormir 8535 Pump Casing Page 28 of 41 Copyright Squirrel Medical® 2017 - E&OE...
Disposal of Electrical Equipment Contact Squirrel Medical for advice on disposal of the Dormir 8535 MRS equipment. Warranty The warranty of the Squirrel Medical Dormir 8535 MRS shall be as detailed in the terms and conditions of the sale. Product Returns...
Assurance: that the goods will be placed in suitable biohazard packaging (available upon request) prior to collection Squirrel Medical will log this e mail and arrange collection normally within 24 working hours. The mattress will be collected and returned to our service centre.
Has the mattress or Tube Set been punctured? If you want the Dormir 8535 MRS to be; Mains Power not connected ON - reconnect mains power ...
Page 32
MRS will require servicing. Take the Dormir 8535 MRS out of use until it has been serviced. It is important that the mattress and patient are checked frequently to ensure that there is sufficient air to safely support the patient.
Foam Base: MAXIMUM MIN = 30Kg PATIENT MAX = 150 Kg WEIGHT Dormir System GMDN Codes Part Number GMDN Code Dormir 8535 MRS, 90cm DOR090200BD 63241 Dormir 8535 Plus MRS, 90cm DORPLUS090200 63241 Dormir Mattress, 90cm DOR303 63641 Dormir 8535 Pump...
Page 34
To isolate unit from Mains Supply; switch off at wall socket and remove plug. In the event of a serious incident occurring in relation to the Squirrel Medical Dormir System: Report the event and details of the occurrence immediately to Squirrel Medical for further assessment (e-mail sales@ squirrelmedical.com /...
11. Appendix The Dormir 8535 pump is EMC compliant (emissions & immunity) under its representative model, the PL-8900 PLUS. 11.1 Guidance and manufacturer’s declaration – electromagnetic emissions Sep 2023 Issue 4.5 Page 35 of 41...
Do not continue to use a Pump if you find or suspect any damage to the Mains Cable Any faults, defects, damage or excessive wear must be dealt with immediately or the Dormir 8535 MRS should be taken out of service www.squirrelmedical.com sales@squirrelmedical.com 0845 003 7979 Squirrel Medical Ltd.