● Do not use the CPR sensor on patients under 8 years old or below 25kg weight. ● The CPR sensor is used for single patient at a time. ● The CPR sensor can be use together with Mindray defibrillator and Mindray monitor supporting the CPR function.
● Replacing the battery if the CPR sensor is configured with is not allowed. It must be replaced by your service personnel. ● Properly performed CPR may result in fracturing of the patient’s ribs, external chest wall bruising or abrasion. ●...
● When cleaning or disinfecting the CPR sensor, avoid the CPR sensor connector. Product Specifications WARNING ● The CPR sensor may not meet the performance specifications if stored or used outside the specified temperature and humidity ranges. Degree of protection against Type CF defibrillation proof electrical shock Type of protection against...
Page 5
Sensor size < 150mm × 63mm × 20m Compression area < 92mm × 53mm × 19mm Physical Specifications Adhesive mount area ≤ 98mm × 45mm Sensor weight ≤ 220g (including battery, without CPR sensor cable) Compression depth Measurement range 0.0 to 8.0 cm Effective range 1.5 to 8.0 cm Accuracy...