Summary of Contents for FMC Technologies Smith Meter Ultra Series
Page 1
Liquid Ultrasonic Flowmeters Smith Meter ® Ultra Series C Remote Display ™ Installation / Operation / Maintenance Manual Bulletin MNLS007 ║ Issue/Rev 0.0 (7/14)
Page 2
Important All information and technical specifications in this documentation have been carefully checked and compiled by the author. However, we cannot completely exclude the possibility of errors. FMC Technologies is always grateful to be informed of any errors. Smith Meter is a registered trademark of FMC Technologies.
The Ultra Series Ultrasonic Flowmeter Remote Display Unit is a touch screen display designed to be used as an HMI for FMC Technologies equipment with a web based operator interface. The display is enclosed in an explosion proof housing and is designed to be located outdoors and in hazardous environments.
Ultra Series C Remote Display I/O/M Manual Installation 2 – Installation 2.1. Mechanical Installation 2.1.1. General The display is designed to be mounted to a plate through the holes tapped in to the rear of the housing using size M6 x 1.8 screws. The display unit must be supplied with an Ethernet RJ45 connection and 24 VDC power.
Select wiring and cable glands suitable for 90°C operation. • Contact manufacturer at address listed for information on the dimensions of the flameproof joints. FMC Technologies Measurement Solutions, Inc. 1602 Wagner Avenue Erie, Pennsylvania 16510 2.1.3. For North American Installations: Conduit connections must be in accordance to: USA –...
Ultra Series C Remote Display I/O/M Manual Installation 2.1.4. All Installations – Notes CAUTION: To prevent ignition of hazardous atmospheres, disconnect from supply cir- cuit before opening any enclosure. Keep tightly closed when circuits are in operation. WARNING: Contains internal battery-powered circuit. To prevent ignition of hazardous atmospheres, do not open enclosure unless area is known to be non-hazardous.
Page 10
Ultra Series C Remote Display I/O/M Manual Installation Page 10 • MNLS007 ║ Issue/Rev. 0.0 (7/14)
Ultra Series C Remote Display I/O/M Manual Installation 3 – Display Control Board 3.1. Wiring Examples / Information 3.1.1. Power Supply The electronics are designed to be powered by 24 VDC, +20% / -15%, 7W. The shield of the cable should be connected only at the electronic enclosure, via the cable gland.
Ultra Series C Remote Display I/O/M Manual Display Control Board 3.1.3. Direct Connection to the Ultrasonic Meter The display unit can also be connected directly to the Ultrasonic meter using a CAT5 Ethernet cable (or better). Note that the Remote Display will still need to be configured with the Serial Number or IP Address of the ultrasonic meter to function.
If a host serial number is configured, the display will search the network for an FMC Technologies device with the specified serial number, obtain an IP address, and attempt to connect with that host. If it fails with the device serial number it will attempt to connect to the secondary URL.
Page 14
Ultra Series C Remote Display I/O/M Manual Powering Up the Remote Display Page 14 • MNLS007 ║ Issue/Rev. 0.0 (7/14)
Ultra Series C Remote Display I/O/M Manual Display User Interface 5 – User Display Interface 5.1. Main Menu The main menu will be displayed if the display does not connect to a device after pow- ering up. The main menu can also be accessed by interrupting the startup process by tapping the configuration icon during startup.
Ultra Series C Remote Display I/O/M Manual Display User Interface 5.2.1. Configuration – General Date/Time – Adjust the date and time of the display unit. Once the display has been sealed the date and time adjustment will be locked out. Set Passwords –...
Ultra Series C Remote Display I/O/M Manual Display User Interface IP address to the Remote Display connected to the local network. This is the default network setting from the factory. Fixed – If the network administrator prefers to fix the IP address then this configuration can be used.
Ultra Series C Remote Display I/O/M Manual Display User Interface 5.2.4. Modes The Modes setting is used to configure the electronic seal and to configure the back- lighting sleep mode for the display. Electronic Seal State This is used to set the electronic seal onto the display unit. Sealing locks out the ad- justment of parameters that would be required to be fixed by a weights and measures official.
Page 19
Ultra Series C Remote Display I/O/M Manual Display User Interface Backlight Timeout This sets the amount of time before the LCD backlight switched off for a sleep mode. The following backlight timer duration options are available from the configuration screen. •...
Ultra Series C Remote Display I/O/M Manual Display User Interface 5.3. Maintenance Settings Allows for the unit to be reset to factory defaults by clearing all new settings. 5.3.1. Touch Screen Calibration Opens a touch position calibration screen. This will calibrate the screen input by displaying a series of touch screen targets and matching the input values with the known location of the targets.
Ultra Series C Remote Display I/O/M Manual Display User Interface 1. Order a USB adapter cable from FMC Technologies; this cable connects to CN3 and presents a standard type A USB connector. 2. Prepare a USB flash drive by creating an empty file named “password-reset.txt” on the root directory (folder).
Ultra Series C Remote Display I/O/M Manual Display User Interface 5.5. Reconnect The Reconnect button initiates an attempt to connect to configured devices using the same sequence used during power up. If the connection attempt is unsuccessful the Remote Display will return to the main menu.
Ultra Series C Remote Display I/O/M Manual Maintenance 6 – Maintenance 6.1. Display HMI Electronics Note: The Display must be powered off while the electronics enclosure is open in a hazardous atmosphere. The Display must also be powered off it any electronic boards are to be replaced.
Returned Goods Policy A Return Material Authorization (RMA) number must be obtained prior to returning any equipment to FMC Technologies Measurement Solutions, Inc. for any reason. An RMA number can be obtained by contacting Customer Service via the information on page 2 of this manual.
Ultra Series C Remote Display I/O/M Manual Related Publications 7 – Related Publications The following literature can be obtained from FMC Technologies Measurement Solutions Literature Fulfillment at measurement.fulfillment@fmcti.com or online at www.fmctechnologies.com/mesurementsolutions. When requesting literature from Literature Fulfillment, please reference the appropriate bulletin number and title.