regulators are recommended for the calibration of the TIGER XT. If flow regulators are to be used
then a 0.3l/min is recommended, a flow adaptor should be used to avoid over-pressurising the
instrument (sold separately). Please ensure you are familiar with the entire calibration procedure
before attempting to calibrate your TIGER XT.
The calibration of your TIGER XT must be carried out in a clean air environment.
Ensure all parts of the Calibration Kit are available and ready for use.
Never calibrate the Zero with the span gas connected.
Custom Calibration Procedure
From the main running screen, select the 'CAL' icon.
1.
Highlight the PID option and press 'OK'.
2.
Select the 'Custom Calibration' icon.
3.
Zero calibration: Attach the Carbon Filter Assembly (if using) and select 'OK' to commence
4.
Zero calibration sequence. The display will countdown 30 seconds. Once complete, select
'OK' to proceed. Disconnect the Carbon Filter Assembly and replace the cap ends.
The useful life of the Carbon Filter Assembly will be shortened if it is open to
atmosphere for prolonged periods.