Instrument Parameter
Commands and Queries
371B User Manual
The Instrument Parameter commands and queries group is helpful for determin-
ing the status of the 371B when problems are encountered. The 371B can be
queried for its firmware version, all current settings, a list of all valid command
and query headers, and whether debug mode is enabled. A command is included
in this group to place the 371B into a known operating condition by initializing
most front-panel settings.
Foe active troubleshooting, the 371B can be placed in debug mode to display the
last 15 characters of a received string. A test routine is available to report back
ROM and RAM problems.
The commands and queries in this group include:
DEBug
DEBug?
HELp?
ID?
INIt
SET?
TESt?
An alphabetical listing of all command and query headers is given in the
"Command Index".
A cross reference for front-panel controls and corresponding GPIB commands
and queries is also given in Table 4–6.