Do you have a question about the SAVERCELL Series and is the answer not in the manual?
Questions and answers
Subscribe to Our Youtube Channel
Summary of Contents for medi SAVERCELL Series
Page 1
Users Guide & Testing Instructions SAVERCELL Series Ques�ons? 800-765-3237 or visit www.mediproducts.net...
Page 2
Reiterate that the system is not intended to support life or life- • suppor�ng equipment but rather for emergency power in non-life- threatening medical procedures. _____________________________________________________________________________________ Medicanix/Medi-Products | 281 Fields Lane, Suite 2B | Brewster, NY 10509 800.765.3237 • www.mediproducts.net...
Page 3
Clearly state that it is the sole responsibility of the end user or • equipment owner to test, maintain, and keep records of the batery backup power system. _____________________________________________________________________________________ Medicanix/Medi-Products | 281 Fields Lane, Suite 2B | Brewster, NY 10509 800.765.3237 • www.mediproducts.net...
Page 4
Over Voltage Protec�on • Overload Protec�on • Over Temperature Protec�on • Short Circuit Protec�on • Low Temperature Protec�on • Low Voltage Protec�on • Over Current Protec�on • _____________________________________________________________________________________ Medicanix/Medi-Products | 281 Fields Lane, Suite 2B | Brewster, NY 10509 800.765.3237 • www.mediproducts.net...
Page 5
Ensure the appliance that is being supported is on and opera�ng. • 6. Monitoring: Monitor the SaverCell’ s status on the display. • Ensure that the connected refrigerator or freezer is receiving backup power. • _____________________________________________________________________________________ Medicanix/Medi-Products | 281 Fields Lane, Suite 2B | Brewster, NY 10509 800.765.3237 • www.mediproducts.net...
Page 6
When you're not using the SaverCell, power off the connected refrigerator • or freezer. Press and hold the power buton on the SaverCell un�l it powers down. • _____________________________________________________________________________________ Medicanix/Medi-Products | 281 Fields Lane, Suite 2B | Brewster, NY 10509 800.765.3237 • www.mediproducts.net...
Page 7
Standard run�me for opera�ng room equipment is 2 hours. • Dura�on: The test should not last more than 2 to 3 minutes. • _____________________________________________________________________________________ Medicanix/Medi-Products | 281 Fields Lane, Suite 2B | Brewster, NY 10509 800.765.3237 • www.mediproducts.net...
Page 8
Maintain detailed records of these tests for the protec�on of vaccines and • valuable inventory. Resources: U�lize test logs provided in service manuals or available for download from • the Medi-Products website. _____________________________________________________________________________________ Medicanix/Medi-Products | 281 Fields Lane, Suite 2B | Brewster, NY 10509 800.765.3237 • www.mediproducts.net...
Page 9
Avoid touching the metal Extra Batery Port; clean it with a dry cloth if • needed. Connect the cable �ghtly to prevent overhea�ng, which could affect • performance or pose a fire risk. _____________________________________________________________________________________ Medicanix/Medi-Products | 281 Fields Lane, Suite 2B | Brewster, NY 10509 800.765.3237 • www.mediproducts.net...
10°F Icon flashes Batery overload Remove high-power protec�on short circuit electrical appliances protec�on during discharging Batery failure Contact SAVER cell customer service Icon stays ON _____________________________________________________________________________________ Medicanix/Medi-Products | 281 Fields Lane, Suite 2B | Brewster, NY 10509 800.765.3237 • www.mediproducts.net...
Page 11
Any malfunc�on or failure of this product while it is in the possession of the owner during the warranty period if the malfunc�on or failure is not caused by a defect in material and workmanship of MEDI-PRODUCTS or if the malfunc�on or failure is caused by unreasonable use, including the failure to verify the equipment’s u�lity and usefulness prior to...
Weekly Test Log: Date: Test Start Time: Test End Time: Status: Tested by: Monthly Test Log: Date: Test Start Time: Test End Time: Status: Tested by: _____________________________________________________________________________________ Medicanix/Medi-Products | 281 Fields Lane, Suite 2B | Brewster, NY 10509 800.765.3237 • www.mediproducts.net...