000A2787B (21" x 24")
000A2788B (15" x 24")
000A2789B (9" x 12")
000A4076A (6" x 8")
Thank you for purchasing a Hallowell EMC Heated Hard Pad. They come equipped with Auto-Shutoff Quick Connect fittings; when they are disconnected from
the pump, they close. No water will leak out making it unnecessary to use, or forget to use, pinch clips.
If you have purchased the Hallowell EMC Heated Hard Pad and find your pump is not equipped with fittings to adapt, here are some easy solutions. The first is to
obtain the appropriate adaptor kit from HEMC: Part #000A4637 for Gaymar T/Pumps and Hamilton Pumps, or Part #000A4629 for Adroit HTP-1500 Pump -
each come with installation instructions. The second solution would be to evaluate the equipment and decide if it may be time to upgrade. Please refer to our
website
www.hallowell.com
for full description and ordering details.
Heated Hard Pads w/ Leads and Connectors
Adroit
HTP-1500
P/N 000A4084
Simply adapt quick connect fittings on hard pad into opposite gender on pump.
The Hallowell EMC Hard Pad has been designed to be used with existing heat therapy systems such as the Gaymar T/Pump, Hamilton Pump and Adroit HTP-
1500 pump. Follow the instructions in your specific operator's manual in regards to startup, shutdown, storage, draining and cleaning instructions.
In addition to the Startup Instructions in you Pump's Operator Manual (refer to instructions in your specific operator's manual) it is necessary to ensure that
there are no air bubbles left in the Hard Pad. The presence of air bubbles will block the flow of heated water to that area downstream of the bubbles.
When filling the system, it is most convenient to prop the pad up on its edge so that the water flows in the bottom up and out the top. Add room temperature
distilled water to the pump as the level in the sight window of the pump drops and the table fills. When the table appears full, tilt it a bit first toward one end then
toward the other end. In each position tap the table gently to allow any remaining air bubbles to flow up and be drawn out the top hose and back to the pump.
In general, an eye should be kept on the system to ensure that no bubbles have developed and are inhibiting the even flow of water throughout the entire surface of
the Hard Pad. Should such bubbles develop simply tilt and tap the table as described above.
Plan ahead, you must allow 20 to 30 minutes for the larger size and 10 minutes for the small pad to come up to temperature. Covering the Hard Pad will help
reduce the warm up time. Animals may be placed directly on the table for maximum heat transfer then wrapped and covered with blankets or drapes as required.

Advertisement

loading

Summary of Contents for Hallowell EMC 000A2787B

  • Page 1 No water will leak out making it unnecessary to use, or forget to use, pinch clips. If you have purchased the Hallowell EMC Heated Hard Pad and find your pump is not equipped with fittings to adapt, here are some easy solutions. The first is to obtain the appropriate adaptor kit from HEMC: Part #000A4637 for Gaymar T/Pumps and Hamilton Pumps, or Part #000A4629 for Adroit HTP-1500 Pump - each come with installation instructions.
  • Page 2 This side of your pad will be upright. Hold This side of your pad will in this position until you see water fill Heated Hard Pad Filling Instructions be upright. Hold in this each channel. position until you see 1. Attach water circulating pump to the heated hard pad. Ensure the water fill each channel.