iM3 CR 8 VET Replacement Instructions

Image plate scanner

Advertisement

Quick Links

Inlet Guide Replacement Kit
Part Number: IC0118
Introduction. This document provides the instructions necessary to replace the Inlet Guide of a CR 8 VET system. Make
sure to read this entire document before proceeding with any service.
The following information is supplementary to the device installation and operating instructions.
As a general rule, the installation and operating instructions for the device must also be observed. These instructions
include important information such as safety instructions and information on the setup, electrical connections, disinfection
process, cleaning process, etc.
Any repairs above and beyond routine maintenance must only be carried out by suitably qualified personnel.
Scope of Delivery. The following items are included in the scope of delivery:
– Inlet Guide (Qty 1)
– M3 X 8 Socket Head Cap Screws (Qty 10)
Required Tools (not supplied).
– 2.5 mm Allen wrench
– 5 mm flat-head screwdriver
NOTICE
Electrostatic-sensitive devices on the
assemblies
Always switch off the operating voltage before
i
carrying out any assembly or installation work on
the device.
Before and during the installation or assembly
i
work on the device, always ensure that the person
carrying out the work is grounded.
Before working on the appliance or in case of
danger, disconnect it from the mains (e. g. pull the
mains plug).
Task Guidelines. Personnel performing the replacement tasks should use standard industry guidelines for working
on equipment. These include the following:
Always use a clean well-lit work area with ample space required for the size of the job.
i
Observe all warnings and precautions for safety as shown by the labels placed on the equipment. Keep all attaching
i
hardware and fastening screws together with the associated removed assembly.
If necessary use separate storage containers or envelopes for each hardware group.
i
Prior to removing any part or assembly, note location and orientation of assemblies being removed.
i
Tag wires and associated mating connectors before disconnecting.
i
Use care when disconnecting mating connectors so as not to damage the connector keys and connection to
i
the associated printed circuit board, wire or cable.
Be aware of the damage impact of electrostatic discharge (ESD) on electronic devices and use ESD precautions
i
when handling printed circuit boards and wiring.
Do not lift or move the unit by the top Touchscreen Assembly.
i
Only llift the unit by the sides of the Transport Assembly and/or the bottom base.
i
Replacement Instructions
Page 1 of 4
Image Plate Scanner

Advertisement

loading

Summary of Contents for iM3 CR 8 VET

  • Page 1 Image Plate Scanner Replacement Instructions Introduction. This document provides the instructions necessary to replace the Inlet Guide of a CR 8 VET system. Make sure to read this entire document before proceeding with any service. The following information is supplementary to the device installation and operating instructions.
  • Page 2 Removal Procedures Power Removal. Prior to performing the procedures contained in this document, turn off power by disconnecting the line cord from the wall outlet. CAUTION Risk of damage to unit as a result of improper handling. Do not move the unit during operation. Do not lift or move the unit by the top Touchscreen Assembly.
  • Page 3 Re-Assembly Procedures Installing the Inlet Guide. Refer to Figure 2 and install the Inlet Guide as follows: CAUTION Risk Wiring Damage Make sure to feed all of the wires from the Touchscreen Assembly through the center hole of the replacement Inlet Guide. Make sure all Main board wires are properly connected.
  • Page 4 Recommissioning Procedures Check Scanner Operation. Check the scanner operation with the replacement Inlet Guide by performing the following steps: 1. Reconnect the power supply to the wall outlet. 2. Turn the unit power ON. 3. Initialize the unit for scanning normal images using the user’s imaging software. 4.