Mopec FP025 User Manual

Mopec FP025 User Manual

Mini formalin pumping station

Advertisement

Quick Links

Mopec FP025 Mini Formalin Pumping Station
FP025 Manual Rev.1
Installation Service User Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the FP025 and is the answer not in the manual?

Questions and answers

Summary of Contents for Mopec FP025

  • Page 1 Mopec FP025 Mini Formalin Pumping Station FP025 Manual Rev.1 Installation Service User Manual...
  • Page 2 DISCLAIMER Copyright © 2023 by MOPEC All rights reserved. No part of this publication may be reproduced, distributed, or transmitted in any form or by any means, including photocopying, recording, or other electronic or mechanical methods, without the prior written permission of the publisher.
  • Page 3: Table Of Contents

    Product elements ................................ 8 Warranty Statement ..............................9 Installation ....................................10 How to unpackage your FP025 Formalin Pumping Stations..................10 Decommissioning and returning the Unit ........................10 How to Store the Product ............................10 Quick Start Guide ..................................11 Startup ..................................
  • Page 4 Spare Parts ..................................... 16 Installation Service User Manual FP025 Manual Rev.1...
  • Page 5: Preface

    1 PREFACE 1.1 Description of use The Mopec FP025 Mini Formalin pump is the ultimate portable pumping unit, built for conveniently transferring formalin from Mopec prefilled formalin 5-gallon cubes or larger prefilled formalin drums to a dispensing valve. The system is portable and powered by a standard 110v outlet for convenience and ease of use.
  • Page 6: Obtaining Instructions

    Documentation, user instructions and technical information can be ordered by calling MOPEC at (800)-362-8491. 1.5.3 Documentation Feedback If you are reading MOPEC product documentation on the internet, any comments can be submitted on the support website. Comments can also be sent to customerservice@mopec.com We appreciate your comments.
  • Page 7: Description Of The Product

    2.1 Purpose of the Product The Mopec FP025 Mini Formalin pumping station’s purpose is to provide a way for laboratories to transfer and dispense formalin from prefilled formalin 5-gallon cubes or larger prefilled Formalin drums to a dispensing valve. The dispensing valve makes it easy and convenient to fill small cups or containers for use during the grossing process.
  • Page 8: Product Elements

    2.5 Product elements The Mopec FP025 Mini Formalin transfer pumping station is constructed as one assembly and has all the user features built into the exterior of the unit. A. Power Switch The Power Switch turns the unit on and off. When in the ON position it supplies power to the pump to allow it to build pressure in the system.
  • Page 9: Warranty Statement

    (if installation is provided by MOPEC), the defective or nonconforming Product or Part thereof will be repaired or (at MOPEC's option) replaced free of charge, FCA MOPEC's dock. All warranty claims must be in writing and received by MOPEC within the warranty period. The warranty is not transferable (other than to customers of MOPEC’s authorized Distributors), and will not apply unless the Equipment has been properly installed, maintained and operated in accordance with all instructions;...
  • Page 10: Installation

    3.1 How to unpackage your FP025 Formalin Pumping Stations. 3.1.1 Unboxing Contents: ¥ Remove the packaging and look the unit over for hidden damage. If found, take photos, and contact Mopec. ¥ Take photographs of any damage and contact MOPEC or your private freight carrier if applicable. ¥...
  • Page 11: Quick Start Guide

    4 Quick Start Guide 4.1 Startup The following section explains how to operate the FP025 Formalin Mini Pumping station and details of the operations. 4.1.1 Connecting the unit to the supplying 2.5/5 Gallon formalin cubitainer ¥ Ensure the main power switch is in the OFF position.
  • Page 12: Maintenance

    5 MAINTENANCE 5.1 How to Maintain the Product Minimal maintenance is required for the FP025. Simply keeping the unit clean and ensuring there are no leaks in the hoses or frayed wires will ensure optimal performance of the product. 5.2 Stainless Steel Maintenance & Cleaning 5.2.1...
  • Page 13: Mopec Service

    Reference the ID tag which is located on the inside front wall of the trailer on the electrical box. of the trailer. Product Model Number: Example: FP025-1 Product Serial Number: Example FP025-1-140203-001 Call 1-800-362-8491 and follow the prompts. Or email us at customerservice@mopec.com Installation Service User Manual FP025 Manual Rev.1...
  • Page 14: Electrical Schematic

    5.4 Electrical Schematic Installation Service User Manual FP025 Manual Rev.1...
  • Page 15: Troubleshooting And Repair

    Dispensing Valve opened too Open valve to the right of dispensing tap little Leak Replace damaged hose Damaged hose Replace pump/Fittings Unit has leak in the system Damaged Pump/Fitting Replace Dispense Spigot Leaking Dispense Spigot Installation Service User Manual FP025 Manual Rev.1...
  • Page 16: Appendices

    Electrical Part Description Mopec # OEM # Use(s) Electric Pump PP0529 DEPCO # 8005-733-155 Formalin Pump On/Off Switch PE0940 Power Switch IEC Power Cord PE0923 NEMA 5-15P TO IEC320 C13 Hospital Grade Cord Installation Service User Manual FP025 Manual Rev.1...

Table of Contents