Lympha Press PCD-51 731A User Manual

Lympha Press PCD-51 731A User Manual

Sequential compression therapy system

Advertisement

PCD-51™
Model 731A
Sequential Compression Therapy System
User Manual
L98005G3-EU-A

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PCD-51 731A and is the answer not in the manual?

Questions and answers

Karen Neila
February 16, 2025

where can I buy the Ktec adapter for the PCD-51 lymphe press

1 comments:
Karen Neila
February 16, 2025

@Karen Neila

John Warm
July 7, 2025

Can pcd-51 model731a be used in Europe without a converter

Summary of Contents for Lympha Press PCD-51 731A

  • Page 1 PCD-51™ Model 731A Sequential Compression Therapy System User Manual L98005G3-EU-A...
  • Page 3: Table Of Contents

    User Manual: PCD-51™ 731A Table of Contents Important - Before You Start ..............iii Indications ..................... iv Contraindications................... iv Symbols used in this Manual ..............v Safety Warnings and Precautions ............v Labels and Symbols on the Console, Garments, and/or Packaging ..viii 1.
  • Page 4 User Manual: PCD-51™ 731A Appendix I - Specifications ..............14 Appendix II - EMC Manufacturer's Declarations ........15 Copyright © 2020 by: Mego Afek AC Ltd. All rights reserved No part of this manual may be reproduced or copied in any form by any means including graphic, electronic or mechanical media, photocopying, typing, or other information retrieval systems, without the written permission of Mego Afek AC Ltd.
  • Page 5: Important - Before You Start

    User Manual: PCD-51™ 731A Important - Before You Start 1. Read this entire manual before setting up and operating the PCD-51™ system. Keep the manual for future reference. 2. Keep all packing materials in case you need to transport the console. The packing materials have been designed to protect the system from damage during shipment.
  • Page 6: Indications

    User Manual: PCD-51™ 731A Indications • Primary lymphedema • Secondary lymphedema • Venous insufficiency • Venous stasis ulcers • Dysfunction of the muscle pump PCD-51™ is intended for use by patients at home, and by medical practitioners in clinics. PCD-51™ is intended for use by patients above age of 12 years old and above weight of 30 kg.
  • Page 7: Symbols Used In This Manual

    User Manual: PCD-51™ 731A Symbols used in this Manual Symbol Description Note: Describes how the system functions or provides a note on how best to use it. Caution: Refers to the safe operation of the system and provides cautions where the possibility of loss of data or damage to the equipment exists.
  • Page 8 User Manual: PCD-51™ 731A Warning: Do not modify any part of the PCD-51™ system. Warning: Do not service or maintain any part of the PCD-51™ system during use by a patient. Warning: Connect to the PCD-51™ console only parts of the PCD-51™ system.
  • Page 9 User Manual: PCD-51™ 731A Caution: Use the PCD-51™ system up to 3000m above mean sea level. Caution: To prevent any damage to equipment, use only accessories, detachable parts and materials described in this user manual. Caution: While in use, place the console on horizontal firm surface only. Do not place the console on bed, blanket, mattress, pillow, or soft furniture.
  • Page 10: Labels And Symbols On The Console, Garments, And/Or Packaging

    User Manual: PCD-51™ 731A Labels and Symbols on the Console, Garments, and/or Packaging Symbol Description Location Degree of protection against ingress of IP21 On console base water Read instructions before use On console base Read instructions before use (for China only) On console base Read instructions before use On garment label Level of protection type BF applied part...
  • Page 11 User Manual: PCD-51™ 731A Symbol Description Location CE label showing compliance with European Directive 2007/47/EC amending Directive On console base 93/42/EEC, concerning medical devices Serial number of the console On console base Fragile, handle with care On package Keep dry On package This side up On package...
  • Page 12: Introduction - The Pcd-51™ System

    User Manual: PCD-51™ 731A 1. Introduction - The PCD-51™ System The PCD-51™ system is a sequential pneumatic compression therapy system for treatment and management of lymphedema, venous insufficiency, venous stasis ulcers and dysfunction of the "muscle pump". The PCD-51™ system consists of: •...
  • Page 13: System Components

    User Manual: PCD-51™ 731A 2. System Components The system consists of the console, compression garment(s), and hoses that deliver the air from the console to the garment. 2.1. Console and Accessories Table 1 - Console and Accessories Component Illustration The Console supplies air to the compression garments. There are two air outlets on the front of the unit.
  • Page 14: Compression Garments

    User Manual: PCD-51™ 731A 2.2. Compression Garments Your distributor will select the appropriate size and type of garment for you, according to your licensed medical practitioner's instructions. Some garments have pressure limitations, which are indicated in the garment instructions for use packed with the garment. Caution: If your licensed medical practitioner prescribes pressures higher than the limits for your garment, consult your distributor.
  • Page 15: Setting Up The System

    User Manual: PCD-51™ 731A 3. Setting up the System Before using the PCD-51™ system for the first time, arrange the treatment area so that the patient can reach the console at any time during treatment. Caution: When the system has been stored in extreme temperature conditions of -20°C (-4°F) or 70°C (158°F) between uses, wait for two (2) hours before using the system.
  • Page 16: Preparing For Treatment

    User Manual: PCD-51™ 731A 4. Insert the DC power adapter cable into the DC adapter socket on the console, and then insert the DC adapter into an appropriate 100-240 Volt wall outlet. Note: The console is "multi-voltage" and can be used for travel. An appropriate socket adapter must be used in countries with incompatible wall outlets.
  • Page 17: Attaching The Hose Bundle To The Console

    Also note the two air outlets on the front of the console. 2. Insert the hose bundle connector into one of the two air outlets at the front of the console, making sure the logo Lympha Press ® faces UP. The hose bundle connector "clicks" into place. If you do not hear a click, move it gently to make sure it is secure.
  • Page 18: Operating Pcd-51

    User Manual: PCD-51™ 731A 4. Operating PCD-51™ Following is a description of the console operating panel. Read this section before operating the console for the first time. Lock Knob Status Indicator Pressure Knob Time Knob Start/Stop Button Figure 1 - The Console Operating Panel The Console Controls Table 2 - Console Controls Control...
  • Page 19: Setting Treatment Pressure And Time

    User Manual: PCD-51™ 731A Control Function Start/Stop Button: Used to start or stop treatment. Power Switch: Used to turn On the console (located at the bottom-right backside of the console) DC Adapter Socket: Used for connecting the DC power adapter (located under the Power Switch) 5.
  • Page 20: Starting The Treatment Session

    User Manual: PCD-51™ 731A 6. Starting the Treatment Session Now you are ready to operate the console and begin your treatment. Note: Hold the console handle with one hand while turning On / Off the Power Switch with your other hand. 1.
  • Page 21: Maintenance & Storage

    User Manual: PCD-51™ 731A 8. Maintenance & Storage 8.1. Cleaning the Console 1. Ensure the DC power adapter is unplugged before cleaning the console. 2. Wipe clean using a slightly damp cloth. 3. Do not allow any moisture or liquids to enter the console! 8.2.
  • Page 22: Storage

    User Manual: PCD-51™ 731A 8.3. Storage 1. Keep the console, all components, and garments away from direct sunlight. 2. Store in a dry, shaded place (see caution below). 3. Do not twist or fold the hose bundle. 4. Store the DC power adapter lightly wrapped and secured. 5.
  • Page 23: Troubleshooting

    User Manual: PCD-51™ 731A 9. Troubleshooting Table 3 - Troubleshooting Symptom Possible Cause Corrective Action No electricity Check the electrical wall outlet. DC power adapter Verify that the DC power adapter cable is connected to the DC adapter socket The console is not on the console, and the DC power adapter is connected to the working.
  • Page 24: Repairs And Technical Support

    User Manual: PCD-51™ 731A 9.1. Repairs and Technical Support Contact your local distributor for repairs and/or technical support. 9.2. Authorized Representative in Europe MedNet EC-REP GmbH Borkstrasse 10 48163 Muenster Germany Tel: +49 251 32266-61 Fax: +49 251 32266-22 Web: http://www.mednet-eurep.com 9.3.
  • Page 25 User Manual: PCD-51™ 731A Appendix I - Specifications Sequential Therapy Device Model: 731A, appliance coupler, transportable, rated: 100-240 VAC, 50-60 Hz Type of protection against electric shock Class II Degree of protection against electric shock Type BF Degree of protection against ingress of water IP21 Degree of safety of application in the Equipment not suitable for use in the presence of a flammable anesthetic mixture...
  • Page 26 User Manual: PCD-51™ 731A Appendix II - EMC Manufacturer's Declarations Electromagnetic Compatibility (EMC) statement for home healthcare environment The PCD-51™ system has been evaluated to international standard IEC 60601-1-2 "General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests".
  • Page 27 User Manual: PCD-51™ 731A Model 731A - Electromagnetic Emissions - Manufacturer Declaration The Model 731A is intended for use in the electromagnetic environment specified below. The customer or the user of the Model 731A should assure that it is used in such an environment. Emissions Test Compliance Electromagnetic Environment - Guidance...
  • Page 28 User Manual: PCD-51™ 731A Model 731A - Electromagnetic Immunity - Manufacturer's Declaration The Model 731A is intended for use in the electromagnetic environment specified below. The customer or the user of the Model 731A should assure that it is used in such an environment . Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment - Guidance Portable and mobile RF communications equipment should be used no closer...