Download Print this page

Fema Electronica ASGARD CCT-01 Manual

Converter-transmitter

Advertisement

Quick Links

P.I. Santiga Altimira 14 (Talleres 14, Nave 2)
Tel: (+34) 93.729.6004 Fax : (+34) 93.729 .6003
1.- COMMON TECHNICAL SPECIFICATIONS
Output 0/20mA ó 4/20mA R
Output 0/10 Vdc
Response Time
Galvanic Isolation
Isolation Test
Accuracy
Ripple
Pass Band
Storage Temperature
Working Temperature
Temperature Coef.
Power Supply
Consumption
Weight
Wire Crossection
Housing and Cover
Terminals
Housing and Cover
Terminals
Mounting
2.- ADJUST AND CALIBRATION PROCEDURE
1.- Open the housing to access the instrument internal circuits
2.- Select apropriate jumpers on boards ME, MP and MS
3.- Connect signal generator to signal input terminals
Connect multimeter to signal output terminals
4.- Power up the instrument as indicated on the label
5.- Generate the signal level low
Operate potentiometer P1 on ME until indication shows desired signal output low level
6.- Generate the signal level high
Operate potentiometer P2 on ME until indication shows desired signal output low level
7.- Repeat steps 6 to 9 in order to correct deviations and check adjust
8.- On model CCT-32, check excitation output voltage for transducers, 24 Vdc, maximum 25 mA
< 600 Ohms max. 22 mA ±3%
L
R
> 1000 Ohms max. 11 V ±3%
L
≤ 250 mSec
2 KV
50 Hz/1 min (between all circuits)
eff.
4 KV
50 Hz/1min
eff.
Class <0,2 for CCT-32 and CCT-01
Class <0,3 for CCT-04 and CCT-08
≤ 0,5 %
1,5 Hz (-3 dB)
-30 to +80 °C (-22 to +176 ºF)
-10 to +60 °C (+14 to +140 ºF)
≤ 0,015 %/ °C
See label on instrument
<1,5 VA
270 gr.
2
4 mm
maximum
IP-40
IP-20
Polycarbonate, Light Grey, RAL 7032, UL 94 V-1
Polycarbonate, Dark Grey, UL 94 V-2
Standard DIN rail (DIN 46277, DIN EN 50022)
(35 x 7,5mm) (1,38 x 0,3") .
CONVERTER-TRANSMITER
CCT- 01 .- Signals in DC Voltage (Vdc)
CCT- 04 .- Signals in AC Current (Aac)
CCT- 08 .- Signals in AC Voltage
CCT- 32 .- Signals in Process (mAdc)
E 08210-BARBERÀ DEL VALLÈS (SPAIN)
fema@fema.es
MECHANICAL DIMENSIONS in mm (inches)
1
3
4
5
6
H i
INPUT / ENTRADA
OUTPUTS / SALIDAS
mA
Vdc
+
-
+
-
Lo
7
9 10 11 12
45
(1,771")
13/96 rev.0911/03
.- Serie CCT
www.fema.es
36
(1,417")
110
(4,33")
75
(2,952")

Advertisement

loading

Summary of Contents for Fema Electronica ASGARD CCT-01

  • Page 1 13/96 rev.0911/03 CONVERTER-TRANSMITER .- Serie CCT CCT- 01 .- Signals in DC Voltage (Vdc) CCT- 04 .- Signals in AC Current (Aac) CCT- 08 .- Signals in AC Voltage CCT- 32 .- Signals in Process (mAdc) P.I. Santiga Altimira 14 (Talleres 14, Nave 2) E 08210-BARBERÀ...
  • Page 2 3.- ACCESS TO INTERNAL CIRCUITS 1.-With a flat screwdriver, force the front cover ands walls towards A and B, until fixations ‘3’ are free. 2.-Take the instrument from points ‘2’ ‚ and extract it pulling towards C, until the internal circuits appear. 3.-Internal circuits are connected with the help of pins and allow access to potentiometers ans jumpers for range adjustment.
  • Page 3 7.- MODELS 01 and 08 .- RANGES, JUMPERS, CONNECTIONS The CCT converter can be configured in order to adequate input and output signal ranges to the final application. We show herunder technical information on the most common ranges. <1V >1V RANGE 0/100 mV 0/1 V...
  • Page 4 11.- PRELIMINARY NOTES EXCITATION VOLTAGE.- The model CCT-32 supply the Excitation voltage for sensors, through the terminals 3 & 5. Do not connect these terminals to other INSTALLATION .- PRECAUTIONS The installation and use of this unit must be done by qualified external power supply, permanent damages may result to the unit.

This manual is also suitable for:

Asgard cct seriesAsgard cct-04Asgard cct-08Asgard cct-32