EIUK PTB 600 Instruction Manual

Portable temperature calibrator

Advertisement

Quick Links

PTB 600
Portable Temperature Calibrator
Instruction Manual
PTB600-MAN-V10.4
Eurotron
Instruments (UK) Ltd
Unit 13 Riley Close, Royal Oak Industrial Estate
NN11 8QT- Daventry, United Kingdom
T+ 0044 (0) 1327 871044
F: 0044 (0) 1327 301 255
www.eurotron-uk.com
info@eurotron-uk.com
1
PTB 600 Instruction Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PTB 600 and is the answer not in the manual?

Questions and answers

Summary of Contents for EIUK PTB 600

  • Page 1 Portable Temperature Calibrator Instruction Manual Eurotron Instruments (UK) Ltd Unit 13 Riley Close, Royal Oak Industrial Estate NN11 8QT- Daventry, United Kingdom T+ 0044 (0) 1327 871044 F: 0044 (0) 1327 301 255 www.eurotron-uk.com info@eurotron-uk.com PTB600-MAN-V10.4 PTB 600 Instruction Manual...
  • Page 2: Table Of Contents

    232/C ....................26 OMMUNICATION ROTOCOL 10.4 - L : 230V ...................... 29 IST OF SPARE PARTS MODEL 10.5- D ................29 ECLARATION OF CONFORMITY AND CHECK REPORT   10.6 - D ......................29 RAWING AND WIRING DIAGRAM PTB600-MAN-V10.4 PTB 600 Instruction Manual...
  • Page 3: Warning

    SYMBOLS BEING USED IN THIS MANUAL OR ON THE INSTRUMENT CAUTION: HOT SURFACE OR PART CAUTION: REFER TO ACCOMPANING DOCUMENTS CAUTION: RISK OF ELECTRICAL SHOCK Note: In this manual: where not specified, the numbers in parentheses make reference to the annexed drawing. PTB600-MAN-V10.4 PTB 600 Instruction Manual...
  • Page 4: Introduction

    * Using of the equipment * Re-calibration procedure * Preventive maintenance * Typical faults and ways of their remedies Users must observe all the usual safety rules out in this manual for own security and to avoid equipment failure. PTB600-MAN-V10.4 PTB 600 Instruction Manual...
  • Page 5: Scope Of Supply

    : 2 lines 20ch x line (3.2x5.5) back lighting. * Resolution : 0.01°C/0.1°C. * processor : 80C552 * A/D converter : Σ-∆ 24 bits * E2PROM memory for recording parameters. * RS232 Single serial output. PTB600-MAN-V10.4 PTB 600 Instruction Manual...
  • Page 6: Service (Function)

     completely check the oven by PC  automatically or manually calibrate many probes  cyclically check temperature sensor long life or stress condition  register and print results obtained according to ISO 9000 standards. PTB600-MAN-V10.4 PTB 600 Instruction Manual...
  • Page 7: Quantity

    PTB600‐INS‐04  Insert DN 35x190 mm custom( max 3 holes from 3,5 to 20mm)  PTB600‐INS‐05  Insert DN35x190 mm custom ( max 2 holes from 3,5 to 20mm)  PTB600‐INS‐IR  Black body Insert DN 35x190 c/w Reference Sensor  PTB600‐TC‐OPT‐01  SRS‐PRT‐45‐190B reference sensor and special UKAS calibration  PTB‐CASE‐02  Aluminium Transport Case with space for accessories(PTB/PLB)  PTB‐CASE‐01  Vinyl Case with Pockets for accessories (PTB/PLB)   Certification: all the instruments are supplied with final testing, stability and accuracy certification traceable to International standards. UKAS calibration available on request PTB600-MAN-V10.4 PTB 600 Instruction Manual...
  • Page 8: General Recommendations

    (fig. 3). Always verify the range of the probes to be calibrated before using; the maximum temperature of the probes should be higher then the internal temperature otherwise the probe could break Fig.1 Fig.2 Fig.3 PTB600-MAN-V10.4 PTB 600 Instruction Manual...
  • Page 9: Safety Instructions

    The equipment adopt the following devices to protect operation from hazard:  Max. temperature safety thermostat (10) to disconnect the heating system.  Protection grid to avoid any contact with the internal oven.  Protection fuses (3)  Ground conductor. PTB600-MAN-V10.4 PTB 600 Instruction Manual...
  • Page 10: Preparation Of Operation

    The calibrator runs on a voltage of 230 Vac (115V by request), 50/60Hz. A 2.5mt. cable is supplied with the calibrator fitted with 2 conductors plus earth. Make sure that the plant is earthen correctly before switching the instrument on. PTB600-MAN-V10.4 PTB 600 Instruction Manual...
  • Page 11 COMMAND LIST POS. DESCRIPTION SUPPLY SOCKET MAIN SWITCH PROTECTION FUSES SWITCH TEST BUSHES RS-232 SOCKET THERMOREGULATOR +DISPLAY HEATING LED COOLING LED SWITCH TEST LED EXTERNAL PROBES SOCKETS (TC version only) PTB600-MAN-V10.4 PTB 600 Instruction Manual...
  • Page 12: Operation Procedure

    Bear in mind, however, that constant use at extreme temperatures reduces the life of the resistance itself. Limit the number of hours at which the resistance is used at maximum temperatures to the time required by the calibrator in order to prolong the life of the resistance. PTB600-MAN-V10.4 PTB 600 Instruction Manual...
  • Page 13: Equalising Block

    (reference to paragraph 4); therefore switch on the oven. Consulting chapter 9 – typical faults – for any problems on the thermostat. N.B.: the thermostat mounted on standard ovens has been calibrated in factory to intervene at 560 or 610°C ±5°C depending on the models. PTB600-MAN-V10.4 PTB 600 Instruction Manual...
  • Page 14: Start-Up Instructions

    Make reference to the temperature value of the external standard instrument inserted in the equalizer block and connected to an external instrument (Fig. 5). When possible: put the sensitive elements of the probes near and at the same dept (reference to fig.1-3). Fig.5 PTB600-MAN-V10.4 PTB 600 Instruction Manual...
  • Page 15  Before switch off the calibrator make sure that the temperature of the block is almost the same as ambient temperature. Cooling: To reduce the oven’s temperature, change the set point and wait for the cooling. PTB600-MAN-V10.4 PTB 600 Instruction Manual...
  • Page 16: Use Of The Functions

    EST SENSOR FAIL : wrong connection or configuration of the EXT probe REF SENSOR FAIL: wrong connection or configuration of the REF probe SENSORS FAIL: wrong connection or configuration of the REF and EXT probes PTB600-MAN-V10.4 PTB 600 Instruction Manual...
  • Page 17: Switch Test (Sw. On Sw. Off)

     To re-calibrate the instrument is necessary to have a standard temperature instrument, the software ‘CALIBRA’ and follow the instructions of the software or alternately follow the instructions of item 10.1. 8 - SEQUENCE OF MAINTENANCE  Not applicable PTB600-MAN-V10.4 PTB 600 Instruction Manual...
  • Page 18: Typical Faults

    - The control probe (9) is cut off - Replace the thermocouple. “Overrange” or RTD failure or is in short circuit. - The thermo regulator (7) is - Replace the thermo regulator faulty. PTB600-MAN-V10.4 PTB 600 Instruction Manual...
  • Page 19: Appendices

    It reduce the value showed on on the display 2. If you keep the display 2. If you keep pressed the key “UP” the pressed the key “DOWN” the speed of increase rises speed of reduce rises PTB600-MAN-V10.4 PTB 600 Instruction Manual...
  • Page 20 Display reading resolution; Press the key to select 0,1 or 0,01 and press E key to accept. SW. ON Switch on; displays the temperature at which the thermostat connected to the terminals "SWITCH TEST" is closed. PTB600-MAN-V10.4 PTB 600 Instruction Manual...
  • Page 21 REF SENSOR TYPE: J, R, S, N, K, E, Pt100, Pt100 3wires This parameter allows selecting the kind of sensor read on the display and connected to the four REF. Sensor terminals.(item 6.4.1) PTB600-MAN-V10.4 PTB 600 Instruction Manual...
  • Page 22 It indicates the swinging value of the temperature, which has been set to see on the Display the symbol of the oven ÷ steadiness. The symbol light on when the temperature is stable for over 6 minutes. PTB600-MAN-V10.4 PTB 600 Instruction Manual...
  • Page 23 * which appears on the display after about 5 seconds. 8. Select CAL: INT set Yes and confirm by pressing E key . Calibration begins. The procedure takes a few seconds, at the end of which there is a Beep. PTB600-MAN-V10.4 PTB 600 Instruction Manual...
  • Page 24 * which appears on the display after about 5 seconds. Select CAL: REF Set Yes and confirm pressing E Key. Calibration begins. The procedure takes a few seconds. At the end there is a Beep. PTB600-MAN-V10.4 PTB 600 Instruction Manual...
  • Page 25 10.2 - Microprocessor regulator: control description PTB600-MAN-V10.4 PTB 600 Instruction Manual...
  • Page 26: Communication Protocol Rs232/C

    * the variable 8 is available only for the models SOLAR-2I-X; the value of the variable corresponds to the table. ** the variable 25/26 is available only for the models SOLAR-2I-X; the value of the variable corresponds to the table. *** the value of the variable 10 corresponds to the table. PTB600-MAN-V10.4 PTB 600 Instruction Manual...
  • Page 27 The response does not include the measure unity, to read the unity read the variable 10: the command string is: $1RVAR10_<cr> the response string is: *1_0 for °C the response string is: *1_1 for °F PTB600-MAN-V10.4 PTB 600 Instruction Manual...
  • Page 28 26 to the number corresponding to the type of sensor which you want to read (2 for the thermocouple K) and then set the variable 8 to 3. the command strings are: $1WVAR26_2<cr> $1WVAR8_3<cr> Do likewise for the other variables. PTB600-MAN-V10.4 PTB 600 Instruction Manual...
  • Page 29: List Of Spare Parts: 230V Model

    The declaration of conformity CE is at the end of the English manual, the test report is included with the calibrator 10.6 - Drawing and wiring diagram The drawings are at the end of the English manual. PTB600-MAN-V10.4 PTB 600 Instruction Manual...
  • Page 30 Eurotron Instruments (UK) Ltd Massimiliano Moltrasio Managing Director RAPPORTO PROVA ESEGUITO E REDATTO DA: CENTRO MISURE RADIOELETTRONICHE P.M.M. IN DATA 07/03/2000 E SEGUENTI CENTRO SIT N°08/E ENTE EMITTENTE: QUALITÀ Date : 17/04/1996 Revision 13/04/2010 PTB600-MAN-V10.4 PTB 600 Instruction Manual...
  • Page 32 PTB600-MAN-V10.4 PTB 600 Instruction Manual...
  • Page 33 PTB600-MAN-V10.4 PTB 600 Instruction Manual...

Table of Contents